Introduction : A wrong estimate of the dry weight in hemodialysis patients increases their morbidity and mortality. The aim of our study was to compare the results of the estimated dry weight using the clinical examination with the results of the lung ultrasound, impedance measurement, ultrasound of the inferior vena cava and B-type Natriuretic Peptide (BNP) assay. Methods : A cross-sectional study was conducted in an hemodialysis center at Fez. The estimated dry weight of the patients was evaluated before and 30-60 minutes after the hemodialysis session using the four above-mentioned methods. Results : The values, measured by the different techniques used, drop significantly after hemodialysis (p<0.001). The measurement studied before and after the hemodialysis, had shown a significant correlation between the results of the impedance measurement, lung ultrasound, the maximum and minimum of the Venous Collapsibility Index (VCI) index diameter as well as its collapsibility. However, no correlation was found between BNP levels and other methods. A significant difference between the results provided by the techniques studied and the results of the clinical evaluation was found in 84.4% of patients. In univariate analysis, only age was significantly shown as an associated factor with error estimation (p<0.04). Conclusion : This study is one of the few studies using lung ultrasound to assess the hydration status of hemodialysis patients. It also showed a good correlation with other methods. The existence of a gap between the subjective and objective target weight suggests the usefulness of a systematic periodic use of bioimpedance and ultrasound techniques even if clinical symptoms are absent.
Keywords: Dry weight; Hemodialysis; Bioelectrical impedance; Inferior vena cava diameter; Ultrasound lung; B-type natriuretic peptide; Dry weight; Hemodialysis; Bioelectrical impedance; Inferior vena cava diameter; Ultrasound lung; B-type natriuretic peptide
This is a cross-sectional study, conducted during March 2014 in the hemodialysis center of Al Ghassani provincial Hospital in Fez.
Using an arteriovenous fistula, the patients included in this study were aged over 18, and have begun chronic hemodialysis since at least 3 months.
following criteria:
• An acute event (infectious episode or a hospitalization) within the 3 months preceding the study regardless of the cause. • A lung disease with pulmonary fibrosis, or dyspnea stage IV of NYHA showing a heart failure that may affect the results of lung ultrasound regardless of the state of hydration. • Inability to wear bioimpedance spectroscopy (BIS) (prostheses, pacemaker ...).
All our patients receive a conventional intermittent dialysis for two to three sessions per week, using low permeability polysulfone membranes with a standard bicarbonate dialysate. The Ultrafiltration rate is prescribed according to the interdialytic weight gains compared to the target weight clinically determined by the treating nephrologist.
The survey was carried out by a nephrologist not belonging to the team of hemodialysis center where the study is conducted. A pre-operating sheet was used for the collection of information.
In addition to demographic data and those related to kidney disease, comorbidities and dialysis prescription, we collected all episodes of intradialytic hypotension, cramps or postdialytic asthenia, during the last three sessions of hemodialysis.
The dry weight assessment parameters were measured in two phases: one hour before the Hemodialysis (HD) session and between 30 to 60 minutes following the end of the same session.
Therefore, the parameters collected are:
• Weight, under the usual conditions of the center using an electronic scale.
• The supine blood pressure measurement is done after 10 minutes rest, using a validated electronic device.
Bioimpedance spectroscopy (BIS): We used the BCM (Body ® Composition Monitor, Fresenius Medical Care®, Germany), dedicated to the analysis of body composition and nutritional status of patients. The impedance was performed to the patients, who were lying down flat in bed, after having rested immediately 10 minutes before and 30 to 60 minutes after the hemodialysis session. The electrodes were placed on the wrist of the contralateral arm of the Arteriovenous Fistula (AVF) and on the homolateral ankle.
Lung Ultrasound: Lung ultrasound was performed by the same trained nephrologist, who did not have access to patients’ clinical data and also to the impedance results. The imaging was made by using an ultrasound (KOLTRON Magic Maestro) with
tial syndrome characterized by the presence of specific artefacts called “B-lines” or “comet tails “ which are proof of pneumonitis (congestion of the lungs) caused by fluid overload in he- 7,8 modialysispatients.7,8 Ultrasound examination was performed to the patients in supine position, with a longitudinal scan from the second to the fourth intercostal space of the left hemi-thorax and from the second to the fifth intercostal space of the right hemithorax at the medioclavicular and midaxillary lines of each side. B-lines were defined as a hyperechogenic linear artifact that are emerging from the pleural line, going to the bottom of the screen, and are being coherent with respiratory movements. The number of B-lines has been determined by the sum of the B-lines found in each examined site. Thus, the selected number reflects the extravascular accumulation of the liquid in the lung.
Ultrasound of the inferior vena cava (IVC): We used an ultrasound (KOLTRON Magic Maestro) with a 3.5 Mhz cardiac probe. The same trained operator explored the IVC within the sub-xiphoid window at 2.5 cm of the IVC-right atrial junction. The measure of the minimum diameter of IVC (DIVC min) was made during inspiration and during the expiration; the maximum diameter of the IVC (DIVC max) was measured.
Dosage of natriurectic peptide type B (BNP): We assayed the BNP immediately before and after the hemodialysis session using a patented enzyme immunoassay (TOSOH®).
• The subjective target weight: it was the prescribed weight taken from the patient logbook, estimated by the attending nephrologist, and based on clinical criteria such as: weight, blood pressure, presence of edema or vascular congestion, as well as the weight on cardiac index measured by the chest x-ray.
• The objective Target weight: it was the determined weight based on the various performed techniques (Impedance, BNP, lung ultrasound, ultrasound of the inferior vena cava). • Weight loss: is the difference between the weight before and after hemodialysis.
The reference values determined by the impedance to define euvolemia in the normal population are within the range of -1.1 L to 1.1 L.9 Referring to these values, we have classified our patients into three groups:
• Dehydrated patients if the fluid volume is less than the reference value-1.1 L • Patients in euvolemia if the volume of fluid is within the reference range±1.1 L • Patients in overload if the fluid volume is greater than the
The index of the IVC diameter (iIVCD) was measured by dividing the IVCD maximal and IVCD minimal on body surface (Dubois formula) to obtain respectively the iIVCD max, and the iIVCD min.
The collapsibility index of the IVC (CiIVC) was calculated by using the following formula:
[(IVCD max - IVCD min)/IVCD max × 100].
Referring to the criteria of ultrasound,10 we classified patients:
- Dehydrated if the iIVCD max <8 mm/m².
- Euvolemic if 8 mm /m²≤iIVCD max ≤11.5 mm/m².
- In overload if the iIVCD max >11.5 mm/m².
Pulmonary congestion (Lung congestion) by fluid overload was retained in patients with the following characteristics:11
1. Scan which shows several lines “B”: number >2 (criteria of the SFAR) ® 2. Positivity diffuse in more than one scans 3. Bilateral positivity
For the BNP assay method used in our study, values that are below 120 pg/ml are considered as normal; whereas, those above 400 pg/ml are considered as high.12 For the other biological parameters, we used the mean value of serum sodium, calcium, phosphate, CRP, hemoglobin, PTH and albuminemia in the last three months.
The data were entered into an Excel sheet and analyzed using SPSS software Version 20. In the descriptive analysis, quantitative variables were expressed as mean±standard deviation and qualitative variables as percentages. The comparison of means was made using Student test; whereas, the comparison of percentages was performed using Chi-square test. To measure the correlation between the results of the four evaluated objective methods, we have used the bivariate correlation method and estimated the r coefficient for each correlation. Subsequently, three groups of patients were defined according to the difference between the objective target weight determined by the different methods used in our study and the subjective target weight predetermined clinically by the health care team:
- Group A: The objective target weight is less than the subjective target weight with a deviation of more than one kilogram
jective target weight does not exceed one kilogram. This is the standard acceptable gap in our study.
- Group C: The objective target weight is greater than the subjective target weight with a deviation of more than one kilogram.
An univariate analysis was performed to show the correlates of the subjective estimation error of the target weight.
An informed consent for participating in the study was obtained from all patients. No invasive investigation was used. All additional costs associated with the study were funded by the research budget of the Nephrology Department of the University Hospital Hassan II of Fez.
We included 77 patients, aged 48.13±16 years with a sex ratio (M/F) of 1.1. The average length of hemodialysis was 10.1±0.5 years. Initial nephropathy was vascular, glomerular and diabetic in respectively 35%, 20% and 8.5% of cases. 44% of patients are hypertensive. Dyspnea was found in 35.1% of cases, mostly stage I (24.7%) and no patient had dyspnea stage III or IV. Anemia was found in 52.6% with a mean hemoglobin rate of 9.62±2.17 g/dl. The mean albumin rate was 39.73±7.5 g/l. Clinical, biological and demographic characteristics are presented in Table 1.
Hypertension (%) 44 Cardiac ejection fraction (%) 66.5 Dyspnea stage (NYHA %) 35.1 Stage I 24.7 Stage II 10.4 Stage III. IV 0 Intradialytic hypotension (%) 7.8% Weekly sessions (%) 3 times 2 times Residual diuresis (ml/j) 146±394 Rate UF/session (ml) 2500 IQR (2000-3100) Range (1000-4500) Hemoglobin rate (g/dl) 9.6 IQR (8.0-11) Range (4.8-14.1) Albumine rate (g/l) 40.9 IQR ( 36-44.9) Range (20.3-48)
Table 1. Baseline characteristics (mean±SD).
The values of fluid status assessment, which are measured by the impedance, the number of B-lines in lung ultrasound, the IVC index, and the BNP rate, decreased significantly (p<0.001) after hemodialysis compared to predialytic values (Table 2). The same results were shown when we consider the categories of each measurement method: in patients with fluid overload, the values of impedance, IVC, the number of B-lines
(Figure 1). On the other hand, in post dialysis, there were more patients who were classified hypovolemic by ultrasound of the IVC (57%) than by impedance (38%).
Bioimpedance (liter) +1.2±1.2 -0.36±1.3 <0.001 Lung ultrasound 4.4±3.4 1.3±1.6 <0.001 Number of B-lines Ultrasound of VCI iDVCI max (mm) 10.2±2.6 7.6±2.6 <0.001 iDVCI min (mm) 5.1±3.1 2.4±3 <0.001 CVCIi % 53.2±24.1 75.1±29.2 <0.001 BNP (pg/ml) 603±606.5 405.5±560.7 <0.001 VCI: Inferior vena cava; iDVCI: index of the inferior vena cava diameter; CVCI: Collapsibility index of the inferior vena cava; BNP: B natrialpeptid.
The correlation analysis between the results of different 50.6% techniques before and after hemodialysis is summarized in Table 49.4% 3. There was a very significant correlation between the impedance, lung ultrasound, the index of the maximum and minimum IVC and its collapsibility both before and after hemodialysis session. However, there was not any significant correlation between BNP rate and other studied methods.
We also studied the correlation between the change of the ultrasound results and the movement of water during hemodialysis session expressed by weight loss. Only the reduction of the number of the B-lines was significantly correlated with weight loss (β coefficient=0.36, p<0.005) (Table 4).
In our study, only 15.6% of patients did not have a gap between the subjective target weight and objective target weight, while 84.4% of patients had a significant difference between the
- Before hemodialysis session Over hydratation by bioimpedance 0.691
- Afterhemodialysis session Over hydratation by bioimpedance 0.756
* ** 0.001<p<0.05 p<0.001 VCI: Inferior vena cava; iDVCI: index of the inferior vena cava diameter; CVCI: Collapsibility index of the inferior vena cava; BNP: B natrial peptid.
cal evaluation. 70.1% of these patients had a negative gap (group A) and 14.3% of them had a positive gap (group C).
In the univariate analysis, several factors were studied in order to show correlates of these differences including: age, sex, comorbidities, length of hemodialysis, inflammation, anemia, nutritional status, left ventricular hypertrophy, systolic ejection fraction. Age was the only significant factor that was related to the differences since the patients with a negative gap (group A) had an average age of 47.2±12 years; whereas, the subjects with no gap (Group B) had an average age of 58±13 years and those with a positive gap (group C) had a mean age of 50.7±13.7 years (p<0.04).
This is a cross-sectional study conducted to compare firstly the different techniques of assessing water status before and after hemodialysis session, and secondly to assess hydration status of our patients and have a more objective view of their dry weight in order to establish a strategy to optimize their care.
All the techniques investigated in this study show that
** ** ** * 0.372 0.461 -0.244 0.08 ** ** ** B-Lines _ 0.610 0.607 -0.485 0.095 ** ** iDVCI min _ _ 0.788 -0.558 0.052 ** iDVCI max _ _ _ -0.916 0.013 CVCIi _ _ _ _ -0.005 BNP _ _ _ _ _
** ** ** ** 0.559 0.581 -0.507 0.128 ** ** ** B-Lines _ 0.679 0.749 -0.592 0.101 ** ** iDVCI min _ _ 0.837 -0.959 0.001 ** iDVCI max _ _ -0.737 0.097 CVCIi _ _ _ _ 0 BNP _ _ _ _ _
B-linesreduction (%) 0.36 0.005 DVCI max reduction (%) 0.12 NS DVCI min reduction (%) 0.04 NS CVCI reduction (%) -0.11 NS DVCI: the inferior vena cava diameter; CVCI: collapsibility of the inferior vena cava.
Table 4. Linear regression with weight loss.
there is a good correlation between these different methods except the BNP assay. This result fits in with the one found by F Basso, et al. in 2013.13 Although, the new techniques are promising, they have important practical and theoretical limits. For example, skin lesions, wrongly placed electrode, electrical interference, and obesity are the limits of theimpedance.6 In addition, the data of the impedance change 120 minutes after the end of hemodialysis.14 Similarly, Agarwal, et al. has shown that the measurement of IVCD is a method that well reflects the intravascular volume but has a low sensitivity for detecting a change in the fluid in post-dialysis time.15 Consistently, our results shown that inpost-dialysis, there are more patients classified hypovolemic by ultrasound measurements of the IVC than by conductivity measurement. This difference is due to the time required for transferring the fluid from the interstitial sector to the intravascular one “refilling’’. At the end of the hemodialysis session, a relative hypovolemia may exist. The conductivity measurements achieved just after the dialysis session, underestimate the degree of dehydration. While ultrasound of inferior vena cava (IVC) overestimates the degree of dehydration in these patients. The results obtained by these two techniques should be close enough if the measurements were made a few hours after hemodialysis session.16 Among our hemodialysis patients, the rate of BNP is
load, and significantly reduces in post-dialysis time. However, it is not correlated with the results of the other methods. This could partly be explained by the existence of other determinants than the volume overload that may influence BNP rate in these patients, particularly any myocardial aggression.17 On the other hand, this hormone is dialyzable with the membranes of high and low permeability.18 This allows us to infer, as it has already been shown, that the dosage of the BNP does not have an interest in the evaluation of water status in hemodialysis patients,19 since this assay does not distinguish between euvolemic and dehydrated patients.20
Lung ultrasound is a simple method, that is easy, inexpensive, without irradiation, andwhich can be used at the bedside of the patients.21 Nevertheless, it is an operator-dependent method. Previously, it was shown that the number of B-lines was correlated to the extravascular water.21 In our study, the correlation between weight loss and ultrasound data, showed that only the reduction of B-lines was significantly correlated; whereas, no correlation existed with the reduction of IVCD. This confirms that the two preceding techniques evaluate two different fluid compartments: the IVCD reflects the volume of intra-vascular water, and the number of B-lines reflects the volume of extravascular water. Thus, the IVCD is not sensitive in assessing rapid changes of fluids during hemodialysis,22 whereas, lung ultrasound can be performed immediately after hemodialysis.23
At the end of this study, we had shown that a significant difference existed between the objective target weight and the subjective target weight. This confirms that the actual determination of the dry weight is difficult and the clinical estimation remains a non-specific method3 and therefore insufficient. The underestimated dry weight in most of our young chronic hemodialysispatients had no clinical or interdialytic relevance. It misled the nephrologist in 70% of cases. This would be due to a better hemodynamic tolerance of Ultrafiltration or rather of dehydration in young patients who have a good cardiovascular condition. Studies have clearly shown an increased risk of death and cardiovascular events when intradialytic hypotension happens in hemodialysis patients in older population.24 Young hemodialysis patients, asymptomatic despite the hypovolemia, are often underestimated in different studies. Cardiovascular morbidity and mortality risk deserves to be studied in this subgroup of patients. It also raises the question of the value of a periodic systematic use of impedance and ultrasound techniques in all hemodialysis patients even in the absence of a clinical anomaly.
This study is among the few studies that have used lung ultrasound to assess the state of hydration of hemodialysis patients. It allows us to show the right correlation between the results of this review and those of the impedance and the ultrasound of the IVC. The existence of a gap between the subjective target weight and the objective target weight suggests the
according to a strategy adapted to the characteristics of each patient.
The authors declare that they have no conflicts of interest.
The abstract of this study has been selected by the scientific Program Committee for moderated poster presentation at the World Congress of Nephrology 2015.