Knee kinematics has been studied since the seventeen century with increasing enthu-siasm: recent studies showed that femoral rollback, femoral external rotation and tibia internal rotation are all required for normal daily living. Total Knee Arthroplasty (TKA) is a successful procedure in treating subjects with severe knee osteoarthritis: unfortunately, knee joint kine-matic after TKA can differ substantially when compared to the kinematic of the normal knee. Numerous kinematic studies using standing and mobile fluoroscopy, gait analysis technologies and simple in vitro techniques have extensively evaluated those differences. In this review article, the authors reviewed the contribution of different fluoroscopic studies in understanding the biomechanical differences between the native and the replaced knee.
Keywords: Total Knee Arthroplasty (TKA); Knee; Fluoroscopy; Kinematics
Total Knee Arthroplasty (TKA) represents a very successful procedure in the treatment of patients affected by severe end-stage tricompartimental knee osteoarthritis. Many patients return to normal daily living activities, including low-impact recreational activities like biking, golfing, swimming and trekking.1 Unfortunately, TKA patients still experience substantial functional impairment compared with their age- and gender-matched peers, especially when doing biomechanically demanding activities.2 Patient satisfaction after TKA reflects these differences: despite recent advances in surgical techniques, implant designs, and patients education, the satisfaction rate after primary TKA between the last decade of the twentieth century and the first decade of the twenty-first century increased only by 3%, leaving 15% of the patients not fully satisfied.3
The normal knee motion is quite complex since Giovanni Alfonso Borelli first described the biomechanical patterns of the knee joint.4 In recent times, it has been shown that, as the normal knee goes into flexion, the lateral femoral condyle moves progressively posterior on the tibial plateau (“postero-lateral rollback”), while the medial femoral condyle pivots with a 5,6 negligible posterior motion.5,6 The authors of the current editorial note quantified this motion in a dynamic MRI study performed in healthy subjects: at 145 degrees of knee flexion, the average lateral condyle rollback was 15 mm while the average medial condyle rollback was 3 mm (Figure 1).7 Such strongly asymmetric “rollback” of the femoral condyles during normal knee flexion is accompanied by a contemporaneous external rotation of the femur with respect to the
6,8,9 10 Numerous kinematic studies using static and dynamic fluoroscopy, gait analysis, and in vitro techniques have revealed profound differences between normal knee motion
A
joint kinematics.
Currently, TKA systems differ by many characteristics: two of them, the design of the femoral condyle on the sagittal plane (single versus multiple radii) and the level of constraint (Cruciate Retaining-CR, Postero-Stabilized-PS, Medially Congruent-MC, Medial Pivot-MP and Sagittally Stabilized-SS) have been extensively studied with static and, more recently, dynamic fluoroscopy using adapted C-arm units. This editorial note reviews the results of those fluoroscopic studies and compares them with the kinematics of normal knees.
Komistek et al6 used static fluoroscopy to study the biomechanical proprieties of normal knees: in that report, during deep flexion activities, subjects experienced an average of 12.7 mm of lateral condyle motion, whereas the medial condyle motion was only 2.9 mm. During deep flexion activities, their subjects experienced a significant axial rotation of the tibia relative to the femur (average 13°).
Cruciate-Retaining (CR)-TKA designs have been extensively studied with static fluoroscopy since early 2000s: those designs have shown a lack of posterior femoral rollback and a more anterior tibio-femoral contact point during deep 14,15 knee flexion when compared with normal knees. This last phenomenon has been described by many authors as “paradoxi- 16-19 cal motion” : this unwanted finding has been attributed to the incompetent restraining function of the posterior cruciate ligament.
Because of these biomechanical differences between CR-TKA and the native knee, Posterior-stabilized (PS)-TKA have been introduced in the 1970s as an alternative to CR design in order to reproduce normal knee mechanics. Dennis et al20 fluoroscopically analyzed 811 patients after primary TKA performed using 33 different designs during stance phase of gait or a deep knee bend maneuver: kinematic patterns of fixed versus mobile-bearing designs were similar in which femorotibial contact remained relatively stationary with minimal AP femoro-
B
respect to the medial compartment.
tibial translation (“postero-lateral rollback”) respect to the native knee; on the other side, paradoxical anterior femoral translation during deep knee flexion was most commonly observed in the CR-TKA designs. In the same study, the normal medial pivoting kinematic pattern (femoral external rotation/tibial internal rotation) was observed in only 55% of all knees during deep knee flexion.
Shimmin et al,21 in a fluoroscopic study evaluating a medial pivot (MP) TKA design (Saiph, MatOrtho, Leatherhead, UK) characterized by full conformity of themedial articular surface (ball in a socket design) and by partial conformity of the lateral compartment (not flat polyethylene), showed that the medial femoral condyles translated an average of 2 mm posterior to the tibial sulcus while the lateral femoral condyles translated an average of 6 mm posterior to the tibial sulcus during maximum knee flexion. None of the studied knees showed anterior femoral translation (“paradoxical motion”) during flexion.
Recently, Scott et al22 fluoroscopically evaluated a sagitally-stabilized medially spherical TKA (GMK Sphere, Medacta International AG, Castel San Pietro, Switzerland): this TKA design is characterized by a completely spherical femoral condyles and a tibialpolyethylene which is spherical medially and completely flat laterally. In this study, the deep flexion activity in implanted knees showed a mean of 8° tibial internal rotation, 2 mm medial posterior translation and 8 mm posterior translation on the lateral condyle.
The anatomy of the posterior femoral condyles has been extensively studied in the last thirty years: standard ra- 23 24 diologic, magnetic resonance assisted and computer-assisted studies25 have shown that the posterior femoral condyles have single-radius geometry. Historically, TKA manufactures have released two different femoral designs: a single-radius (SR) and a multiple-radii (MR). The SR TKA consists of symmetrical femoral condyles that incorporate a SR in the sagittal plane. The MR-TKA consists of a multiradii femoral design in which the lateral femoral condyle is larger than the medial femoral condyle in the sagittal plane. Recently, Grieco et al9 compared in vivo the kinematic differences in patients implanted with a SR vs. a MR-cruciate retaining (CR) design during deep knee flexion using a novel mobile fluoroscopy system. These authors reported that SR and MR groups experienced statistically similar average
Author Year TKA Design Medial Condyle AP (mm)* Lateral Condyle AP (mm)* Axial Rotation (degrees)* Komistek et al 2003 Normal knee - 2.9 mm Dennis et al 2003 CR + 1 mm (anterior translation) -1.6 mm NA Dennis et al 2003 PS - 1 mm Shimminet al 2015 MP -2 mm Scott et al 2016 SS - 2 mm
Table 1. Influence of level of constraint on fluoroscopically studied TKA kinematics during deep knee flexion. CR=Cruciate Retaining; PS=Posterior-stabilized; MP=Medial pivot; SS=Sagittally stabilized; *Fluoroscopically assisted measurements performed during deep knee flexion (average distance in mm from the tibio-femoral contact point in full knee extension).
AP motion in the lateral condyle: - 0.43 mm (SR) vs. – 1.0 mm (MR); the SR group had significant more AP motion (+ 3.51 mm) compared to the MR group (- 0.42 mm) in the medial condyle. On the other side, they showed that the SR group had a significantly larger amount of physiological axial rotation (average 5.20°) compared to the MR group (average 0.75°).
Many methods have been historically used to evaluate normal and following TKA knee kinematics. The great advantage of video fluoroscopy is to allow analysis of multiple “invivo” activities done in weight-bearing conditions. The major limitation of this technique has always been that only on-site activities (i.e. deep knee flexion) were originally tested by many authors. The introduction of mobile fluoroscopy9 allows to study TKA in vivo kinematics during normal walking and walking up and down an inclined ramp.
Classic fluoroscopic studies (Table 1) have shown that normal knee kinematics is not fully reproduced by many TKA designs. Severe kinematic abnormalities, including reduced posterior femoral rollback, paradoxical anterior femoral translation, reduced axial rotation patterns, are commonly present in our TKA patients. The “perfect” TKA should be biomechanically characterized by posterior femoral rollback during deep knee flexion and by an axial rotation in its medial compartment. Recent fluoroscopic studies by Shimmin et al21 on medial pivot (MP) and by Scott et al22 on sagittally-stable (SS) designs showed that “medially-constrained” designs reproduce closer kinematic patterns to normal knees when compared with cruciate-retaining (CR) and posterior-stabilized (PS) designs.
Further, long-term clinical investigations on “medially-constrained” TKA designs are needed to understand if their promising fluoroscopic findings correlate with a higher postoperative satisfaction rate in our TKA patients.