The Ophthalmology Open Journal · Vol. 3, No. 1 · 2019 · pp. 10–20 · e-ISSN 2475-1278
Complications Related to Implants Used in Anterior Bleb Forming Glaucoma Surgery: A Systematic Review of the Literature
Alejandra Salinas, Fabián Rodríguez, Javier Bardavío
Department of Ophthalmology, Hospital Universitario de Torrevieja, SpainSpain
Department of Ophthalmology, Hospital Verge de la Cinta,Tortosa, SpainSpain
St Mary’s Treatment Centre, Portsmouth, UKUnited Kingdom
DOI: https://doi.org/10.17140/ooj-3-118 · Published 2019-10-11 · Licence: https://creativecommons.org/licenses/by/4.0 · © 2019 The Author(s)
Also available as PDF (figures and tables appear in the PDF).
Abstract
Purpose To make an account of published implant-related complications (IRC) by a systematic review of the literature. Methods A systematic search of Pubmed and Scopus databases and Google Scholar engine was performed with selection criteria to detect papers on IRC. We excluded unrelated papers and reviewed selected ones. We considered papers that did not explicitly state about occurrence or not occurrence of IRC as non-IRC reporting. Main outcome measures were the number of papers reporting on complications, IRC, and types of IRC. Results After the search, selection, and addition, we studied 109 papers. Incidence of IRC was 4.5%, half required explantation. While 26 implant studies found IRCs (23%), 13 case reports on surgical complications, 8 (61.5%) of them reported IRC. Frequent complications were conjunctival erosion, blockage of the tube, migration to anterior chamber or damage to surrounding tissues. Conclusion Most papers did not report on IRC. Length or nature of studies may skew finding IRC. The incidence of IRC was 4.5%. Hard and sharp implants carry a greater risk of IRC and explantation.
Keywords: Implant-related complications (IRC); Glaucoma surgery; Anterior bleb forming
Methods
A systematic search of Pubmed and Scopus databases and Google Scholar engine was performed with selection criteria to detect
papers on IRC. We excluded unrelated papers and reviewed selected ones. We considered papers that did not explicitly state about
occurrence or not occurrence of IRC as non-IRC reporting. Main outcome measures were the number of papers reporting on
complications, IRC, and types of IRC.
Results
After the search, selection, and addition, we studied 109 papers.
26 implant studies found IRCs (23%), 13 case reports on surgical complications, 8 (61.5%) of them reported IRC. Frequent
complications were conjunctival erosion, blockage of the tube, migration to anterior chamber or damage to surrounding tissues.
Conclusion
Most papers did not report on IRC. Length or nature of studies
and sharp implants carry a greater risk of IRC and explantation.
Keywords
Implant-related complications (IRC); Glaucoma surgery; Anterior bleb forming.
Introduction
laucoma is a chronic progressive optic neuropathy that may
cause severe visual loss and affect the quality of life. The G front line of treatment consists of reducing intraocular pressure.
Most patients achieve good medical control: some require one or
more surgical procedures. Although conventional filtering surgery
is effective has only moderate success and are subject to risks.
Some devices have been designed and introduced to improve suc-
cess and risks. Some of these implants are associated with classical
conventional fistulizing surgery or to newer filtering procedures
cc Copyright 2019 by SalinasA. This is an open-access article distributed under Creative Commons Attribution 4.0 International License (CC BY 4.0), which allows Incidence of IRC was 4.5%, half required explantation. While
may skew finding IRC. The incidence of IRC was 4.5%. Hard
that intend the construction of a paralimbal bleb (subconjuncti-
val, Subtenon’s or intra-scleral). Regulatory bodies have studied
the security and effectiveness of these implants and authorized
for human application. Some systematic reviews or meta-analysis 1-4 show that support for the use of these devices may be limitated.
This study aimed to make an account of published implant-related
complications (IRC), this is complications directly caused by the
implant, by a systematic review of the literature up to April 2015.
Only implants that contribute to creating a paralimbal bleb were
considered since those are the ones that are used as an alternative
to trabeculectomy or non-penetrating deep sclerectomy and are
potentially avoidable. We excluded equator bleb creating implants,
angular devices or anterior chamber to supraciliary space shunts
from the scope of this study. We were also interested to know
how many of the papers referred to complications related to the
implants.
Methods
Search Strategy
Three different search engines were used: Pubmed, Google Schol-
ar and Scopus in April 2015. The search was performed over all
fields of bibliographic references. We applied three levels of search
restriction and combined controlled Mesh and natural languages.
Language filters were applied. Inclusion criteria: any series of pa-
tients, whether prospective or retrospective and case reports in
English, Spanish or French (filtered) found introducing the se-
lected keywords. The word “complication” was extended to the
abstract search also. Exclusion criteria: reports on other glaucoma
Table 1. Search Strategy Details
# Searches on Google Scholar Results
((implant* OR collagen implant OR collagen implants OR “Hemacrylic Metha bar”) AND complication) AND (glaucoma)
(implant* OR esnoper OR skgel OR aquaflow OR “ex-press” OR “collagen implant” OR “collagen implants” OR “Hemacrylic” OR “reticulated 2 hyaluronic acid” OR “T flux” OR “T bar”) AND complication* AND Glaucoma-“Molteno Implants”-Molteno-Ahmed-“Krupin valve”-“krupin 2300 valves”-Baerveldt)
(“Filtering Surgery” OR “Trabeculectomy” OR “Filtering Surgeries” OR OR “deep sclerectomy” OR “Sclerostomy” OR “deep sclerectomies” OR viscocanalostomy) AND (implant* OR esnoper OR skgel OR aquaflow OR “ex-press” OR “collagen implant” OR “collagen implants” OR “Hemacrylic” OR “reticulated hyaluronic acid” OR “T flux” OR “T bar”) AND complication* AND Glaucoma-“Molteno Implants”-Molteno- Ahmed-“Krupin valve”-“krupin valves”-Baerveldt)
# Searches on Scopus
(TITLE(implant* OR esnoper OR sk gel OR aqua flow OR “express implant*” OR “collagen implant*” OR “Hemacrylic Metha Stealth” OR “reticulated hyaluronic acid” OR “T flux” OR “T bar”) OR ABS(implant* OR esnoper OR sk gel OR aqua flow OR “express implant*” OR 4 “collagen implant*” OR “Hemacrylic Metha Stealth” OR “reticulated hyaluronic acid” OR “T flux” OR “T bar”)) AND (TITLE(complication*) 1533 ORABS(complication*)) AND (TITLE(glaucoma) OR ABS(glaucoma)) AND (LIMIT-TO(LANGUAGE, “English”) OR LIMIT-TO(LANGUAGE, “French”) OR LIMIT-TO(LANGUAGE, “Spanish”))
(TITLE(implant* OR esnoper OR skgel OR aquaflow OR “express implant*” OR “collagen implant*” OR “Hemacrylic Metha Stealth” OR “reticulated hyaluronic acid” OR“T flux” OR “T bar”) OR ABS(implant* OR esnoper OR skgel OR aquaflow OR “express implant*” OR “col- 5 lagen implant*” OR “Hemacrylic Metha Stealth” OR “reticulated hyaluronic acid” OR“T flux” OR “T bar”)) AND (TITLE(complication*) OR 1106 ABS(complication*)) AND (TITLE(glaucoma) OR ABS(glaucoma)) AND NOT baerveldt) AND (LIMIT-TO(LANGUAGE,“English”) OR LIMIT-TO(LANGUAGE,“French”) OR LIMIT-TO(LANGUAGE,“Spanish”))
(TITLE(“Trabeculectom*” OR “Filtering Surger*” OR “deep sclerectom*” OR “Sclerostomy” OR viscocanalostom*) OR ABS(“Trabeculectom*” OR “Filtering Surger*” OR “deep sclerectom*” OR “Sclerostomy” OR viscocanalostom*)) AND (TITLE(implant* OR esnoper OR skgel OR aquaflow OR “express implant*” OR “collagen implant*” OR “Hemacrylic Metha Stealth” OR “reticulated hyaluronic 6 acid” OR“T flux” OR “T bar”) OR ABS(implant* OR esnoper OR skgel OR aquaflow OR “express implant*” OR “collagen implant*” OR 273 “Hemacrylic Metha Stealth” OR “reticulated hyaluronic acid” OR“T flux” OR “T bar”)) AND (TITLE(complication*) OR ABS(complication*)) AND (TITLE(glaucoma) OR ABS(glaucoma)) AND NOT TITLE-ABS-KEY(Molteno* OR Ahmed OR“Krupin valve*” OR baerveldt) AND(LIMIT- TO(LANGUAGE, “English”)ORLIMIT-TO(LANGUAGE, “French”) OR LIMIT-TO(LANGUAGE, “Spanish”))
# Searches on Pubmed/Medline
(“Prostheses and Implants/complications”[Mesh] OR ((implant*[tiab] OR esnoper[tiab] OR skgel[tiab] OR aquaflow[tiab] OR “express implant”[tiab] OR “express implants”[tiab] OR “collagen implant”[tiab] OR “collagen implants”[tiab] OR “HemacrylicMetha Stealth”[tiab] OR “reticulated hyaluronic acid”[tiab] OR“T flux”[tiab] OR “T bar”[tiab]) AND complication*[tiab]))) AND (“Glaucoma”[Mesh] OR glaucoma*[tiab])
(TITLE(implant* OR esnoper OR skgel OR aquaflow OR “express implant*” OR “collagen implant*” OR “Hemacrylic Metha Stealth” OR “reticulated hyaluronic acid” OR“T flux” OR “T bar”) OR ABS(implant* OR esnoper OR skgel OR aquaflow OR “express implant*” OR “col- 8 lagen implant*” OR “Hemacrylic Metha Stealth” OR “reticulated hyaluronic acid” OR“T flux" OR “T bar”)) AND (TITLE(complication*) OR 967 ABS(complication*)) AND (TITLE(glaucoma) OR ABS(glaucoma)) AND NOT baerveldt) AND(LIMIT-TO(LANGUAGE, “English”) OR LIMIT-TO(LANGUAGE, “French”) OR LIMIT-TO(LANGUAGE, “Spanish”))
(TITLE(“Trabeculectom*” OR “Filtering Surger*” OR “deep sclerectom*” OR “Sclerostomy” OR viscocanalostom*) OR ABS(“Trabeculectom*” OR “Filtering Surger*” OR “deep sclerectom*” OR “Sclerostomy” OR viscocanalostom*)) AND (TITLE(implant* OR esnoper OR skgel OR aquaflow OR “express implant*” OR “collagen implant*” OR “Hemacrylic Metha Stealth” OR “reticulated hyaluronic acid” OR“T flux” OR “T 9 bar”) OR ABS(implant* OR esnoper OR skgel OR aquaflow OR “express implant*” OR “collagen implant*” OR “Hemacrylic Metha Stealth” 314 OR “reticulated hyaluronic acid” OR“T flux” OR “T bar”)) AND (TITLE(complication*) OR ABS(complication*)) AND (TITLE(glaucoma) OR ABS(glaucoma)) AND NOT TITLE-ABS-KEY(Molteno* OR Ahmed OR“Krupin valve*” OR baerveldt) AND (LIMIT-TO(LANGUAGE, “English”) ORLIMIT-TO(LANGUAGE, “French”)ORLIMIT-TO(LANGUAGE, “Spanish”)) drainage devices like Ahmed or Krupin Valve, Baerveltd or Mol-
teno drainage devices. Any anterior chamber to supraciliary or su-
prachoroidal shunts like Gold Micro Shunt or Esnoper V-2000. Pa-
pers in any other languages. Keywords: “prostheses and implants/
complications”, “glaucoma drainage implants”, implant, ex-press,
esnoper, SK gel, aqua flow, collagen implant, Hem Acrylic, reticu-
lated hyaluronic acid, T flux, thiobarbituric acid-reactive (T BAR),
filtering surgery, filtering surgeries,deep sclerectomy, non-penetrat-
ing deep sclerectomy, viscocalanostomy, trabeculectomy, scleros-
tomy. Mendeley reference manager (www.mendeley.com,5 Elsevier)
detected duplications. Refer to Table 1 for search strategy.
Initial Revision Strategy:The Selection of Papers for Revision
All papers’ titles were reviewed by a team of trained or in training
ophthalmologist in search of papers that reported on paralimbal
bleb forming surgical procedures with the implantation of a de-
vice. When in doubt, we reviewed the abstract. We excluded papers
reporting on surgical procedures without an implant, different pro-
Stealth” OR “reticulated hyaluronic acid” OR “T flux” OR “T 11600
1572
TITLE-ABS-KEY(Molteno* OR Ahmed OR“Krupin valve*” OR
1249
TITLE-ABS-KEY(Molteno* OR Ahmed OR“Krupin valve*” OR
cedures all together (posterior forming blebs, angular surgery, an-
terior chamber to supraciliary or suprachoroidal shunts), different
languages or animal studies. We included systematic reviews and
meta-analysis of implanting surgical procedures studies to secure
relevant studies that our search could have missed out, but care-
fully made sure we did not duplicate the study papers by doing so.
Full text of all included papers was requested for revi-
sion. New references were found during revision and added under
the same criteria.
Ethical Committee approval was not requested as no pa-
tient information was searched at any time.
Outcome Measures
The following outcomes were measured: total number of papers
after exclusions, number of papers that report complications, num-
ber of papers that report IRC, type of implant, type of IRC found
as extrusion, migration, damage to surrounding tissue, dysfunc-
tional implants, explanted cases, relocated cases, other (Table 2).
Table 2. Definition of Outcomes to be Measured
Outcomes to be Measured
• Number of papers found once the non-related are excluded • Number of papers that report complications • Number of papers that report implant-related complications • Type of implant-related complications found with every implant o Extrusion o Migration o Damage of surrounding tissue o Obstructed o Explanted cases o Relocated cases o Other
Papers’revision was directed to find a statement on com-
plications within the method and the results sections of the ar-
ticle. We did not look for complications reports within the dis-
Figure 1. Final Number of Papers Considered for Study Purposes cussion section nor conclusions of the papers. Since the presence or absence of complications report is not interpretable, only one reviewer read through every paper. A paper was considered to re- port on IRC if the actual complication was reported or it explicitly stated that no IRC was found during the study. Papers that did not
state one thing or the other were considered not to report on IRC.
In case of doubt, the paper was consulted with the senior ophthal-
mologist in the study group.
A register of complications was established to include
both studies with and without complications’ reports.
Results
The result from the search of the three engines was 253 papers
from Scopus, 314 from Pubmed and 1572 from Google Scholar
that were reduced to 1102 after removing redundant papers. All ti-
tles of those papers were reviewed to discard unrelated ones when
in doubt the abstract was used. Finally, 120 full-text articles were
requested for review. From systematic revisions and meta-analysis,
we found 28 new unknown references that were also requested.
From a total of 148 papers, 23 were excluded due to belonging to
animal or experimental implants studies, reporting on the different
procedure, German language, or unfound full text. Of the 125 reviewed papers, 9 were systematic reviews, and 7 were meta-analysis 1-16 that included studies already reviewed by this research team.
Therefore, the final number of papers considered for study pur- 17-126 poses was 109 (Figure 1).
All these papers correspond to studies or case reports that
report on glaucoma implant surgery and also on surgical complica-
tions (100% of the 109 papers reviewed). Out of the 109 papers reviewed, 75 (68.8%) did not report on IRC, 34 (31.2%) reported
on IRC, 6 (17.6%) of the IRC reporting papers) explicitly report-
ing an absence of IRC (4 prospective and 2 retrospective studies).
Of the 59 prospective studies (54.1%, of 109), 17 reported on IRC
(28.8%, of 59). Of the 37 retrospective studies (33.9%, of 109), 9
(24.3%, of 37) reported IRC. Thirteen of the papers (11.9%, of
Table 3. Understudy Implants, IRC Reporting Papers, Number of Study Eyes (Eyes with Implants in IRC Reporting Papers) and Number of Complicated Cases
Implant Nº of Papers* Nº IRC Papers** Nº of Study Eyes Nº Complicated Cases
Express (Steel) 38 22 (57,9%) 940 54 (5,7%)
Collagen Implant 25 5 (16,6%) 208 9 (4,3%)
SK-gel (Hyualuronic acid) 25 3 (12%) 185 2 (1,07%)
T-flux (Acrylic) 15 1 (7,14%) 25 1 (4%)
Ethicon 1/0 Chromic Catgut 1 1 (100%) 23 1 (4,3%)
PMMA 1 1
ePTFE 1 1
Ologen (Collagen) 9 1
HEALA Flow 1 0
Aquaflow (Collagen) 5 0
Total 109 35
*Twelve papers tested two or three implant devices; **One paper tested two devices PMMA and Collagen implant
Table 4. Reported Complications
Implant Complication
Migration or 1 migration to malposition subconjunctival space
1 conjunctival erosion and extrusion that did Extrusion not require explants or repositioning
Damage to 1 Corneal abrasion surrounding 5 Descemet mb.
tissues detachment
Obstructed
Explanted During hypotony
Explanted as treatment of complication
Repositioned
109) were case reports, 8 (61.5%, of 13) reported IRC.
The total population of eyes considered in this study was
5831 (population range per study: 1-345), 4576 of which had been
implanted (78.4%), the others belonging to control groups. The
population of eyes with an implant included in studies that report-
ed on IRC was 1479. Of those, the number of eyes that suffered
an IRC was 67 (4.5%). Table 3 shows the implants understudied,
the number of IRC reporting papers, the number of study eyes
and the number of complicated cases.
A summary of the complications is in Table 4. The most
common IRC was conjunctival erosion; 20 cases after Express im-
plantation (15 through a full-thickness scleral incision, four under 30 0
35 0
33 0
0 0
0 0
1479 67 (4,5%)
Ethicon 1/0 chromic SK-gel Collagen Implant T-flux Express catgut
1 misplacement 1 migration 1 into anterior 1 migration to subconjunctival into anterior chamber after blunt space chamber trauma 5 into anterior chamber
20 conjunctiva erosions (15 without scleral flap,4 with sclera flap,1 unknown)
5 iris touch 2 endothelial touch 1 Corneal abrasion 1 crystalline lens touch 1 scleral flap erosion only
2 unreported reasons 1 due to infection repair surgery
3 relocated due to other complication Other causes
a partial-thickness scleral flap and one unreported technique) that
required implant removal, and one case of SK-gel that healed un-
der observation. Obstruction of Express tube occurred in 14 cas-
es. Four Express tubes migrated postoperatively into the anterior
chamber (AC) and another one was introduced in the AC during
the procedure. Another Express tube migrated to the subconjunc-
tival space, as it occurred to an SK gel implant, and a further Ex-
press tube was found misplaced. A T flux implant migrated into
the patient’s anterior chamber. A fragment of Ethicon 1/0 chromic
catgut, placed under a deep sclerectomy flap, migrated into de AC
also. Sixteen cases of damage to surrounding tissues were found, 5
Descemet membrane detachments and one corneal abrasion after
collagen implant, five iris touch, two endothelial touches, one cor-
neal abrasion, one crystalline lens touch that induced a white cata-
ract and one scleral flap extrusion, without conjunctival erosion,
after Express implantation. Three primary explantation of Express
tubes, one due to a bleb infection and two for unreported reasons,
as well as three collagen implant removal during the hypotony re-
pair. A total of 29 Express tubes (3% of all tubes implanted) and
three collagen implants (1.4% of all implanted) were explanted.
Discussion
The market offers implants in an attempt to improve post-operative
survival, efficacy or safety for patients requiring glaucoma surgery.
Some studies, however, suggest that traditional trabeculectomy or
newer deep sclerectomy without implants might be just as effective
in reducing intraocular pressure. Devices, therefore, attempt to im-
prove prognosis, but its mere physical presence may also increase
risks. This study aimed to make an account of published IRC by a
systematic review of the literature up to April 2015. We only stud-
ied implants that contribute to creating a paralimbal bleb. Equator
bleb creating implants, angular devices or anterior chamber to su-
praciliary space shunts are not within the scope of this study.
Our search produced a total of 1102 papers. The use of
natural language and mesh terms explain such a large number of
papers found. The initial selection resulted in 120 papers of which
23 were further excluded, and another 28 were added during revi-
sion. Seven systematic reviews, nine meta-analysis, 13 case reports,
37 retrospective and 59 prospective studies were reviewed. Only
an explicit account of IRC was considered as a report. All papers
reviewed (109/109) reported surgical complications. Only 31.2%
reported on IRC: 6 (17.6%) of those (4 prospective and two ret-
rospectives), reported an absence of IRC. Only 28.8% of all pro-
spective and 24.4% of all retrospective studies reported on IRC.
It was therefore interesting to find that of 13 case reports dealing
with complications, 8 (61.5%) were about IRC. We suspect that the
discrepancy in IRC incidence reported in retrospective and pro-
spective studies compared to case reports might be related to the
length of the studies and although inconclusive, could be relevant.
Issues related to reporting could justify further research.
Sixty-seven cases of IRC of 1479 (4.5%) patients were
found, of which 31 required implant removal. Our estimation
of IRC incidence might be biased by the search or the selection
method. The incidence of Express removal was 3% before the
tube was guarded under a scleral flap: that reduced the risk but did
not eliminate it. The initial design of the technique may show that
the risk of extrusion was underestimated. Risk of misplacement
or injury to surrounding tissues like Descemet membrane, corneal
endothelium or the crystalline lens is also relevant as well as migra-
tion into the anterior chamber of mainly the express tubes.
The physical properties of the implant are relevant: hard-
er and sharper implants that remain longer, increase the risk of
complications, mainly extrusion. In our study, the steel implant Ex-
press accounts for most cases of explanted, extruded or migrated
cases. Also, non-absorbable acrylic T flux tends to migrate if left
unfixed. Soft and absorbable implants tend to be more respect-
ful. We speculated that tensile vectors generated by scaring tissue might tend to move rigid pieces and to fold soft material: this could
explain the migration of implants or simple orientation changes.
The intermittent robbing effect of the eyelid over the conjunctiva
exerts an external pressure (tissue towards implant) that injures the
tissue in the opposite direction (implant towards sclera, conjunc-
tiva or both). Both mechanisms are active in all operated eyes, so
the risk to induce complications exists.
Study limitations include missing relevant papers by the
search (no excerpta medica dataBASE (EMBASE) search) or the
selection method: the initial search failed to detect 28 relevant pa-
pers added later. It is possible that relevant data from studies was
filtered out by title or abstract selection. A further limitation is late
publication of this study that was finalized in 2016 and sent for
publication in 2019. Finally, this review relies on what authors re-
port on their papers, under or overreporting issues would modify
these results. However, the study remains useful to stress the fact
that an implant might induce complications by itself. Also, the type
and shape of the material are relevant and worth bearing in mind
at the time of assessing the purpose and usefulness of the implant.
Further, the importance of considering IRC at the time of design-
ing studies to make them sensible to complication detection and
reporting.
Conclusion
An IRC incidence of 4.5% resulted from this study, of which
about a half required removal. Harder and sharper implants carry
a higher risk. It is worth considering the implant’s function and
the risk it may carry its implantation before making the clinical
decision of using them. Studies should consider this risk and be
methodologically designed to detect them.
Acknowledgements
The authors are thankful to Ms. Noemi Aluja for her contribution
towords the bibliography.
Conflicts Of Interest
The authors declare that they have no conflicts of interest.
References
- ChengJWChengSWCaiJPLiYWeiRL. Systematic overview of the efficacy of nonpenetrating glaucoma surgery in the treatment of open angle glaucoma. Med Sci Monit. 2011; 17(7): RA155-RA163. doi: 10.12659/MSM.88184010.12659/msm.881840 https://doi.org/10.12659/msm.881840
- WangWZhangX. Meta-analysis of randomized controlled trials comparing EX-PRESS implantation with trabeculectomy for open-angle glaucoma. PLoS One. 2014; 9(6): e100578. doi: 10.1371/ journal.pone.0100578
- RulliEBiagioliERivaI. Efficacy and safety of trabeculectomy vs nonpenetrating surgical procedures: A systematic review and meta-analysis. JAMA Ophthalmol. 2013; 131(12):1573-1582. doi: 10.1001/jamaophthalmol.2013.505910.1001/jamaophthalmol.2013.5059 https://doi.org/10.1001/jamaophthalmol.2013.5059
- ChengJWWeiRLCaiJPLiY. Efficacy and tolerability of non-penetrating filtering surgery with and without implant in treatment of open angle glaucoma: A quantitative evaluation of the evidence. J Glaucoma. 2009; 18(3): 233-237. doi: 10.1097/ IJG.0b013e3181812827
- Mendeley. Web site. www.mendeley.com. Accessed September 16, 2019
- ChenGLiWJiangFMaoSTongY. Ex-PRESS implantation versus trabeculectomy in open-angle glaucoma: A meta-analysis of randomized controlled clinical trials. PLoS One. 2014; 9(1): e86045. doi: 10.1371/journal.pone.008604510.1371/journal.pone.0086045 https://doi.org/10.1371/journal.pone.0086045
- BuysYM. Trabeculectomy versus Ex-PRESS shunt surgery in residency training. Surv Ophthalmol. 2012; 57(4): 375-378. doi: 10.1016/j.survophthal.2012.02.00610.1016/j.survophthal.2012.02.006 https://doi.org/10.1016/j.survophthal.2012.02.006
- DemeterSHaileyD. Non-penetrating glaucoma surgery using AquaFlow(TM)collagen implants. Issues Emerg Health Technol. 2001; (22): 1-4
- FilippopoulosTRheeDJ. Novel surgical procedures in glaucoma: Advances in penetrating glaucoma surgery. Curr Opin Ophthalmol. 2008; 19(2): 149-154. doi: 10.1097/ICU.0b013e3282f4f49e10.1097/icu.0b013e3282f4f49e https://doi.org/10.1097/icu.0b013e3282f4f49e
- FrancisBASinghKLinSC. Novel glaucoma procedures: A report by the American Academy of Ophthalmology. Ophthalmology. 2011; 118(7): 1466-1480. doi: 10.1016/j.ophtha.2011.03.02810.1016/j.ophtha.2011.03.028 https://doi.org/10.1016/j.ophtha.2011.03.028
- HendrickAMKahookMY. Ex-PRESS mini glaucoma shunt: Surgical technique and review of clinical experience. Expert Rev Med Devices. 2008; 5(6): 673-677. doi: 10.1586/17434440.5.6.67310.1586/17434440.5.6.673 https://doi.org/10.1586/17434440.5.6.673
- SalimS. The role of the Ex-PRESS glaucoma filtration device in glaucoma surgery. Semin Ophthalmol. 2013; 28(3): 180-184. doi: 10.3109/08820538.2013.77119710.3109/08820538.2013.771197 https://doi.org/10.3109/08820538.2013.771197
- BuysYM. Trabeculectomy with Ex-PRESS: Weighing the benefits and cost. Curr Opin Ophthalmol. 2013; 24(2): 111-118. doi: 10.1097/ICU.0b013e32835907a610.1097/icu.0b013e32835907a6 https://doi.org/10.1097/icu.0b013e32835907a6
- SalimS. Ex-PRESS glaucoma filtration device: Surgical technique and outcomes. Int Ophthalmol Clin. 2011; 51(3): 83-94. doi: 10.1097/IIO.0b013e31821e5b8210.1097/iio.0b013e31821e5b82 https://doi.org/10.1097/iio.0b013e31821e5b82
- ShaarawyTGoldbergIFechtnerR. EX-PRESS glaucoma filtration device: Review of clinical experience and comparison with trabeculectomy. Surv Ophthalmol. 2015; 60: 327-345. doi: 10.1016/j. survophthal.2015.01.00110.1016/j.survophthal.2015.01.001 https://doi.org/10.1016/j.survophthal.2015.01.001
- JiQQiBLiuLGuoXZhongJ. Efficacy and safety of ologen implant versus Mitomycin C in primary trabeculectomy: A meta-analysis of randomized clinical trials. J Glaucoma. 2015; 24: e88-e94. doi: 10.1097/IJG.000000000000014710.1097/ijg.0000000000000147 https://doi.org/10.1097/ijg.0000000000000147
- Al Obeidan SA. Migration of reticulated hyaluronic acid implant [SK-GEL] following deep sclerectomy. Int Ophthalmol. 2010; 30(3): 329-331. doi: 10.1007/s10792-009-9325-8 https://doi.org/10.1007/s10792-009-9325-8
- Al Obeidan SA, Osman EA, Al-Muammar AM, Abu El-Asrar AM. Efficacy and safety of deep sclerectomy in uveitic glaucoma. Int Ophthalmol. 2009; 29(5): 367-372. doi: 10.1007/s10792-008- 9247-x https://doi.org/10.1007/s10792-008-9247-x
- Al-ObeidanSAMousaANaseemAAbu-AmeroKKOsmanEA. Efficacy and safety of non-penetrating deep sclerectomy surgery in Saudi patients with uncontrolled open angle glaucoma. Saudi Med J. 2013; 34(1): 54-61
- ArruabarrenaCMuñoz-NegreteFJMárquezCRebolledaG. Results of nonpenetrating deep sclerectomy in inflammatory glaucoma: One year follow up. Arch Soc Esp Oftalmol. 2007; 82(8): 483-487
- AtesHAndacKUretmenO. Non-penetrating deep sclerectomy and collagen implant surgery in glaucoma patients with advanced field loss. Int Ophthalmol. 1999; 23(3): 123-128
- AuerCMermoudAHerbortCP. Deep sclerectomy for the management of uncontrolled uveitic glaucoma: Preliminary data. Klin Monbl Augenheilkd. 2004; 221(5): 339-342. doi: 10.1055/s- 2004-81282010.1055/s-2004-812820 https://doi.org/10.1055/s-2004-812820
- CillinoSZeppaLDi PaceF. E-PTFE (Gore-Tex) implant with or without low-dosage mitomycin-C as an adjuvant in penetrating glaucoma surgery: 2 year randomized clinical trial. Acta Ophthalmol. 2008; 86(3): 314-321. doi: 10.1111/j.1600- 0420.2007.01036.x10.1111/j.1600-0420.2007.01036.x https://doi.org/10.1111/j.1600-0420.2007.01036.x
- Collignon-BrachJRobe-CollignonN. Non perforating trabecular surgery with reticulated hyaluronic acid implant. Bull Soc Belge Ophtalmol. 2000; 276: 61-68
- CoupinALiQRissI. Ex-PRESSTM miniature glaucoma implant inserted under a scleral flap in open-angle glaucoma surgery: A retrospective study. J Fr Ophtalmol. 2007; 30(1): 18-23
- DahanECarmichaelTR. Implantation of a miniature glaucoma device under a scleral flap. J Glaucoma. 2005; 14(2): 98-102. doi: 10.1097/01.ijg.0000151688.34904.b710.1097/01.ijg.0000151688.34904.b7 https://doi.org/10.1097/01.ijg.0000151688.34904.b7
- DemaillyPJeanteur-LunelMNBerkaniM. Non-penetrating deep sclerectomy combined with a collagen implant in primary open-angle glaucoma. Medium-term retrospective results. J Fr Ophtalmol. 1996; 19(11): 659-666
- Detry-MorelM. Non penetrating deep sclerectomy (NPDS) with SKGEL implant and/or 5-fluorouracile (5-FU). Bull Soc Belge Ophtalmol. 2001; (280): 23-32
- Detry-MorelMDetryMB. Five-year experience with non penetrating deep sclerectomy. Bull Soc Belge Ophtalmol. 2006; (299): 83- 94
- Detry-MorelMPourjavanSDetryMB. Comparative safety profile between “modern” trabeculectomy and non-penetrationg deep sclerectomy. Bull Soc Belge Ophtalmol. 2006; (300): 43-54
- DrolsumL. Conversion from trabeculectomy to deep sclerectomy. Prospective study of the first 44 cases. J Cataract Refract Surg. 2003; 29(7): 1378-1384. doi: 10.1016/S0886-3350(03)00013-010.1016/s0886-3350(03)00013-0 https://doi.org/10.1016/s0886-3350(03)00013-0
- DupasBFardeauCCassouxNBodaghiBLeHoangP. Deep sclerectomy and trabeculectomy in uveitic glaucoma. Eye (Lond). 2010; 24(2): 310-314. doi: 10.1038/eye.2009.8210.1038/eye.2009.82 https://doi.org/10.1038/eye.2009.82
- ElginUSimsekTBatmanA. Use of the ex-press miniature glaucoma implant in a child with Sturge-Weber syndrome. J Pediatr Ophthalmol Strabismus. 2007; 44(4): 248-250
- HamardPPlazaLKopelJQuesnotSHamardH. Deep nonpenetrating sclerectomy and open angle glaucoma. Intermediate results from the first operated patients. J Fr Ophtalmol. 1999; 22(1): 25-31
- HamelMShaarawyTMermoudA. Deep sclerectomy with collagen implant in patients with glaucoma and high myopia. J Cataract Refract Surg. 2001; 27(9): 1410-1417. doi: 10.1016/s0886- 3350(01)00959-210.1016/s0886-3350(01)00959-2 https://doi.org/10.1016/s0886-3350(01)00959-2
- JungkimSGibranSKKhanKMurrayA. External trabeculectomy with T-Flux implant. Eur J Ophthalmol. 2006; 16(3): 416- 421. doi: 10.1177/11206721060160030910.1177/112067210601600309 https://doi.org/10.1177/112067210601600309
- KarlenMESanchezESchnyderCCSickenbergMMermoudA. Deep sclerectomy with collagen implant: Medium term results. Br J Ophthalmol. 1999; 83(1): 6-11. doi: 10.1136/bjo.83.1.610.1136/bjo.83.1.6 https://doi.org/10.1136/bjo.83.1.6
- KimCYChangHRLeeJHHongYJ. Surgical outcomes of deep sclerectomy with collagen implant. Korean J Ophthalmol. 2001; 15(2): 107-112. doi: 10.3341/kjo.2001.15.2.10710.3341/kjo.2001.15.2.107 https://doi.org/10.3341/kjo.2001.15.2.107
- KimCYHongYJSeongGJKohHJKimSS. Iris synechia after laser goniopuncture in a patient having deep sclerectomy with a collagen implant. J Cataract Refract Surg. 2002; 28(5): 900-902. doi: 10.1016/s0886-3350(01)01307-410.1016/s0886-3350(01)01307-4 https://doi.org/10.1016/s0886-3350(01)01307-4
- LeszczyńskiRGierek-CiaciuraSForminśka-KapuścikMMrukwa-KominekERokita-WalaI. Nonpenetrating very deep sclerectomy with reticulated hyaluronic acid implant in glaucoma treatment. Med Sci Monit. 2008; 14(2): CR86-CR89
- LiJYLimMCMannisMJ. Traumatic cataract associated with mini glaucoma shunt. J Cataract Refract Surg. 2011; 37(7): 1360- 1362. doi: 10.1016/j.jcrs.2011.04.01910.1016/j.jcrs.2011.04.019 https://doi.org/10.1016/j.jcrs.2011.04.019
- LükeCDietleinTSJacobiPCKonenWKrieglsteinGK. A prospective randomised trial of viscocanalostomy with and without implantation of a reticulated hyaluronic acid implant (SKGEL) in open angle glaucoma. Br J Ophthalmol. 2003; 87(5): 599-603. doi: 10.1136/bjo.87.5.59910.1136/bjo.87.5.599 https://doi.org/10.1136/bjo.87.5.599
- MansouriKShaarawyTWedrichAMermoudA. Comparing polymethylmethacrylate implant with collagen implant in deep sclerectomy: A randomized controlled trial. J Glaucoma. 2006; 15(3): 264-270. doi: 10.1097/01.ijg.0000212211.33265.6d10.1097/01.ijg.0000212211.33265.6d https://doi.org/10.1097/01.ijg.0000212211.33265.6d
- MarekRJoannaWLewczukKSiemiatkowskaAStankiewiczA. Efficacy and safety of deep sclerectomy and phacoemulsification and deep sclerectomy in clinical material of Military Health Service Institute--yearly observations. Klin Oczna. 2006; 108(10- 12): 385-391
- MendrinosEDreifussSDossoABaglivoEShaarawyT. Bacterial keratitis after nonpenetrating glaucoma surgery. J Cataract Refract Surg. 2008; 34(4): 707-709. doi: 10.1016/j.jcrs.2007.11.03910.1016/j.jcrs.2007.11.039 https://doi.org/10.1016/j.jcrs.2007.11.039
- MermoudASchnyderCCSickenbergMChiouAGHédiguerSEFaggioniR. Comparison of deep sclerectomy with collagen implant and trabeculectomy in open-angle glaucoma. J Cataract Refract Surg. 1999; 25(3): 323-331. doi: 10.1016/s0886-3350(99)80079-010.1016/s0886-3350(99)80079-0 https://doi.org/10.1016/s0886-3350(99)80079-0
- Moreno-MontañésJRodríguez-CondeR. Bleeding during gonioscopy after deep sclerectomy. J Glaucoma. 2003; 12(5): 427-429
- Muñoz Negrete FJ, Rebolleda G, Noval S. Non-penetrating deep sclerectomy combined with phacoemulsification. Results and complications. Arch Soc Esp Oftalmol. 2003; 78(9): 499-506
- NeudorferMSadetzkiSAnisimovaSGeyerO. Nonpenetrating deep sclerectomy with the use of adjunctive mitomycin C. Ophthalmic Surg Lasers Imaging. 2004; 35(1): 6-12. doi: 10.3928/1542- 8877-20040101-0310.3928/1542-8877-20040101-03 https://doi.org/10.3928/1542-8877-20040101-03
- OllikainenMLPuustjärviTJRekonenPKUusitaloHMTeräsvirtaME. Mitomycin C-augmented deep sclerectomy in primary open-angle glaucoma and exfoliation glaucoma: A threeyear prospective study. Acta Ophthalmol. 2011; 89(6): 548-555. doi: 10.1111/j.1755-3768.2009.01772.x10.1111/j.1755-3768.2009.01772.x https://doi.org/10.1111/j.1755-3768.2009.01772.x
- OllikainenMLPuustjärviTJRekonenPKUusitaloHMTeräsvirtaME. Mitomycin-C-augmented deep sclerectomy in patients with primary open-angle glaucoma and exfoliation glaucoma: A 1-year prospective study. Acta Ophthalmol. 2010; 88(1): 20-26. doi: 10.1111/j.1755-3768.2009.01509.x10.1111/j.1755-3768.2009.01509.x https://doi.org/10.1111/j.1755-3768.2009.01509.x
- RavinetETrittenJJRoyS. Descemet membrane detachment after nonpenetrating filtering surgery. J Glaucoma. 2002; 11(3): 244-252. doi: 10.1097/00061198-200206000-0001410.1097/00061198-200206000-00014 https://doi.org/10.1097/00061198-200206000-00014
- RebolledaGMuñoz-NegreteFJ. Deep sclerectomy with mitomycin C in failed trabeculectomy. Eye (Lond). 2007; 21(1): 23-28. doi: 10.1038/sj.eye.670218310.1038/sj.eye.6702183 https://doi.org/10.1038/sj.eye.6702183
- RekonenPKannistoTPuustjärviTTeräsvirtaMUusitaloH. Deep sclerectomy for the treatment of exfoliation and primary open-angle glaucoma. Acta Ophthalmol Scand. 2006; 84(4): 507-511. doi: 10.1111/j.1600-0420.2006.00654.x10.1111/j.1600-0420.2006.00654.x https://doi.org/10.1111/j.1600-0420.2006.00654.x
- RivierDRoySMermoudA. Ex-PRESS R-50 miniature glaucoma implant insertion under the conjunctiva combined with cataract extraction. J Cataract Refract Surg. 2007; 33(11): 1946-1952. doi: 10.1016/j.jcrs.2007.06.06910.1016/j.jcrs.2007.06.069 https://doi.org/10.1016/j.jcrs.2007.06.069
- Robe-CollignonNJCollignon-BrachJD. Non perforating trabecular surgery with reticulated hyaluronic acid implant. Bull Soc Belge Ophtalmol. 2000; Suppl: 45-54
- ShaarawyTKarlenMSchnyderCAchacheFSanchezEMermoudA. Five-year results of deep sclerectomy with collagen implant. J Cataract Refract Surg. 2001; 27(11): 1770-1778. doi: 10.1016/s0886-3350(01)01032-x10.1016/s0886-3350(01)01032-x https://doi.org/10.1016/s0886-3350(01)01032-x
- SourdillePSantiagoPYBresson-DumontH. Transition to nonpenetrating trabecular surgery: Sclerotrabeculectomy. J Cataract Refract Surg. 2002; 28(12): 2080-2083. doi: 10.1016/s0886- 3350(02)01363-910.1016/s0886-3350(02)01363-9 https://doi.org/10.1016/s0886-3350(02)01363-9
- StewartRMDiamondJGAshmoreEDAyyalaRS. Complications following ex-press glaucoma shunt implantation. Am J Ophthalmol. 2005; 140(2): 340-341. doi: 10.1016/j.ajo.2005.02.03310.1016/j.ajo.2005.02.033 https://doi.org/10.1016/j.ajo.2005.02.033
- SuominenSHarjuMIhanamäkiTVestiE. The effect of deep sclerectomy on intraocular pressure of normal-tension glaucoma patients: 1-year results. Acta Ophthalmol. 2010; 88(1): 27-32. doi: 10.1111/j.1755-3768.2009.01623.x10.1111/j.1755-3768.2009.01623.x https://doi.org/10.1111/j.1755-3768.2009.01623.x
- TraversoCEDe FeoFMessas-KaplanA. Long term effect on IOP of a stainless steel glaucoma drainage implant (Ex-PRESS) in combined surgery with phacoemulsification. Br J Ophthalmol. 2005; 89(4): 425-429. doi: 10.1136/bjo.2004.05803210.1136/bjo.2004.058032 https://doi.org/10.1136/bjo.2004.058032
- Wagschal LD, Trope GE, Jinapriya D, Jin Y-P, Buys YM. Prospective randomized study comparing Ex-PRESS to trabeculectomy: 1-year results. J Glaucoma. 2015; 24(8): 624-629. doi: 10.1097/ IJG.0000000000000029
- Detry-MorelMDe TemmermanS. Assessment of nonpenetrating deep sclerectomy with reticulated hyaluronic acid implant SKGEL and/or preoperative application of 5-luorouracil: Results of 2 and a half years. Bull Soc Belge Ophtalmol. 2003; (287): 53-62
- Gallego-PinazoRLópez-SánchezEMarín-MontielJ. Postoperative outcomes after combined glaucoma surgery. Comparison of EX-PRESS miniature implant with standard trabeculectomy. Arch Soc Esp Oftalmol. 2009; 84(6): 293-297
- MarzetteLHerndonLW. A comparison of the Ex-PRESSTM mini glaucoma shunt with standard trabeculectomy in the surgical treatment of glaucoma. Ophthalmic Surg Lasers Imaging. 2011; 42(6): 453-459. doi: 10.3928/15428877-20111017-0310.3928/15428877-20111017-03 https://doi.org/10.3928/15428877-20111017-03
- WevillMTMeyerDVan AswegenE. A pilot study of deep sclerectomy with implantation of chromic suture material as a collagen implant: Medium-term results. Eye (Lond). 2005; 19(5): 549- 554. doi: 10.1038/sj.eye.670154110.1038/sj.eye.6701541 https://doi.org/10.1038/sj.eye.6701541
- GoodTJKahookMY. Assessment of bleb morphologic features and postoperative outcomes after Ex-PRESS drainage device implantation versus trabeculectomy. Am J Ophthalmol. 2011; 151(3): 507-513.e1. doi: 10.1016/j.ajo.2010.09.00410.1016/j.ajo.2010.09.004 https://doi.org/10.1016/j.ajo.2010.09.004
- GavrićMGabrićNJagićJCovićA. Clinical experience with EX-PRESS mini glaucoma shunt implantation. Coll Antropol. 2011; 35 Suppl 2: 39-41
- GindrozFRoySMermoudASchnyderCC. Combined Ex-PRESS LR-50/IOL implantation in modified deep sclerectomy plus phacoemulsification for glaucoma associated with cataract. Eur J Ophthalmol. 2011; 21(1): 12-19. doi: 10.5301/ejo.2010.54110.5301/ejo.2010.541 https://doi.org/10.5301/ejo.2010.541
- RekasMLewczukKFuksińskaBRudowiczJPawlikRStankiewiczA. Combined surgery for cataract and glaucoma: PDS with absorbable SK-gel implant compared with PDS with non-absorbable T-flux implant-medium-term results. Curr Med Res Opin. 2010; 26(5): 1131-1137. doi: 10.1185/0300799100371942810.1185/03007991003719428 https://doi.org/10.1185/03007991003719428
- MansouriKTranHVRavinetEMermoudA. Comparing deep sclerectomy with collagen implant to the new method of very deep sclerectomy with collagen implant: A single-masked randomized controlled trial. J Glaucoma. 2010; 19(1): 24-30. doi: 10.1097/ IJG.0b013e3181a2fa46
- Dahan E, Ben Simon GJ, Lafuma A. Comparison of trabeculectomy and Ex-PRESS implantation in fellow eyes of the same patient: A prospective, randomised study. Eye (Lond). 2012; 26(5): 703-710. doi: 10.1038/eye.2012.13 https://doi.org/10.1038/eye.2012.13
- RoySThiHDFeusierMMermoudA. Crosslinked sodium hyaluronate implant in deep sclerectomy for the surgical treatment of glaucoma. Eur J Ophthalmol. 2012; 22(1): 70-76. doi: 10.5301/ ejo.5000054
- MesciCErbilHHKarakurtYAkçakayaAA. Deep sclerectomy augmented with combination of absorbable biosynthetic sodium hyaluronate scleral implant and mitomycin C or with mitomycin C versus trabeculectomy: Long-term results. Clin Experiment Ophthalmol. 2007; 40(4): e197-e207. doi: 10.1111/j.1442- 9071.2011.02611.x10.1111/j.1442-9071.2011.02611.x https://doi.org/10.1111/j.1442-9071.2011.02611.x
- SuominenSHarjuMKurvinenLVestiE. Deep sclerectomy in normal-tension glaucoma with and without mitomycin-c. Acta Ophthalmol. 2014; 92(7): 701-706 doi: 10.1111/aos.1230510.1111/aos.12305 https://doi.org/10.1111/aos.12305
- AteşHUretmenOAndaçKAzarsizSS. Deep sclerectomy with a nonabsorbable implant (T-Flux): Preliminary results. Can J Ophthalmol. 2003; 38(6): 482-488. doi: 10.1016/S0008- 4182(03)80027-310.1016/s0008-4182(03)80027-3 https://doi.org/10.1016/s0008-4182(03)80027-3
- AmbresinAShaarawyTMermoudA. Deep sclerectomy with collagen implant in one eye compared with trabeculectomy in the other eye of the same patient. J Glaucoma. 2002; 11(3): 214-220. doi: 10.1097/00061198-200206000-0000910.1097/00061198-200206000-00009 https://doi.org/10.1097/00061198-200206000-00009
- JehnABBohnkeMMojonDS. Deep sclerectomy with collagen implant: Initial experience. Ophthalmologica. 2002; 216(4): 235- 238. doi: 10.1159/00006384610.1159/000063846 https://doi.org/10.1159/000063846
- GalassiFGiambeneB. Deep sclerectomy with SkGel implant: 5-year results. J Glaucoma. 2008; 17(1): 52-56. doi: 10.1097/ IJG.0b013e3180d0a885
- BissigAFeusierMMermoudARoyS. Deep sclerectomy with the Ex-PRESS X-200 implant for the surgical treatment of glaucoma. Int Ophthalmol. 2010; 30(6): 661-668. doi: 10.1007/ s10792-010-9382-z
- De FeoFBagnisABricolaGScottoRTraversoCE. Efficacy and safety of a steel drainage device implanted under a scleral flap. Can J Ophthalmol. 2009; 44(4): 457-462. doi: 10.3129/i09-12010.3129/i09-120 https://doi.org/10.3129/i09-120
- NarayanaswamyAPereraSAHtoonHMHohSTSeahSKWongTTAungT. Efficacy and safety of collagen matrix implants in phacotrabeculectomy and comparison with Mitomycin C augmented phacotrabeculectomy at 1 year. Clin Experiment Ophthalmol. 2013; 41(6): 552-560. doi: 10.1111/ceo.1205810.1111/ceo.12058 https://doi.org/10.1111/ceo.12058
- TengCCRadcliffeNHuangJEFarrisE. Ex-PRESS glaucoma shunt dislocation into the anterior chamber. J Glaucoma. 2008; 17(8): 687-689. doi: 10.1097/IJG.0b013e318166659510.1097/ijg.0b013e3181666595 https://doi.org/10.1097/ijg.0b013e3181666595
- KannerEMNetlandPASarkisianSRDuH. Ex-PRESS miniature glaucoma device implanted under a scleral flap alone or combined with phacoemulsification cataract surgery. J Glaucoma. 2009; 18(6): 488-491. doi: 10.1097/IJG.0b013e31818fb44e10.1097/ijg.0b013e31818fb44e https://doi.org/10.1097/ijg.0b013e31818fb44e
- MendaSALowryEAPorcoTCStamperRLRubinMRHanY. Ex-PRESS outcomes using Mitomycin-C, Ologen alone, Ologen with 5-fluorouracil. Int Ophthalmol. 2014; 35(3): 357-363. doi: 10.1007/s10792-014-9955-310.1007/s10792-014-9955-3 https://doi.org/10.1007/s10792-014-9955-3
- Huerva V, Carmen Sanchez M, Jesus Muniesa M. Ex-PRESS shunt surgery for glaucoma with post-traumatic aniridia. Semin Ophthalmol. 2015; 30(3): 218-220. doi: 10.3109/08820538.2013.835845 de Jong L, Lafuma A, Aguadé AS, Berdeaux G. Five-year extension of a clinical trial comparing the EX-PRESS glaucoma filtration device and trabeculectomy in primary open-angle glaucoma. Clin Ophthalmol. 2011; 5: 527-533. doi: 10.2147/OPTH.S18565 https://doi.org/10.3109/08820538.2013.835845
- LankaranianDRazeghinejadMRPrasadA. Intermediateterm results of the Ex-PRESS miniature glaucoma implant under a scleral flap in previously operated eyes. Clin Experiment Ophthalmol. 2011; 39(5): 421-428. doi: 10.1111/j.1442-9071.2010.02481.x10.1111/j.1442-9071.2010.02481.x https://doi.org/10.1111/j.1442-9071.2010.02481.x
- Al Obeidan SA, Osman EA, Mousa A, Al-Muammar AM, Abu El-Asrar AM. Long-term evaluation of efficacy and safety of deep sclerectomy in uveitic glaucoma. Ocul Immunol Inflamm. 2015; 23(1): 82-89. doi: 10.3109/09273948.2013.870213 https://doi.org/10.3109/09273948.2013.870213
- MendrinosEMansouriKMermoudAShaarawyT. Longterm results of deep sclerectomy with collagen implant in exfoliative glaucoma. J Glaucoma. 2009; 18(5): 361-367. doi: 10.1097/ IJG.0b013e3181879e4e
- ShaarawyTMansouriKSchnyderCRavinetEAchacheFMermoudA. Long-term results of deep sclerectomy with collagen implant. J Cataract Refract Surg. 2004; 30(6): 1225-1231. doi: 10.1016/j.jcrs.2003.10.03510.1016/j.jcrs.2003.10.035 https://doi.org/10.1016/j.jcrs.2003.10.035
- RhoSKangSYHongSSeongGJJungJJKimCY. Longterm results of deep sclerectomy with small collagen implant in Korean. Korean J Ophthalmol. 2013; 27(1): 34-38. doi: 10.3341/ kjo.2013.27.1.34
- Russo V, Scott IU, Stella A, Balducci F, Cosma A, Barone A, Delle Noci N. Nonpenetrating deep sclerectomy with reticulated hyaluronic acid implant versus punch trabeculectomy: A prospective clinical trial. Eur J Ophthalmol. 2008; 18(5): 751-757. doi: 10.1177/112067210801800515 https://doi.org/10.1177/112067210801800515
- LeszczyńskiRFormińska-KapuścikMBubała-StachowiczBMrukwa-KominekEFilipekEPawlickiK. Nonpenetrating very deep sclerectomy with hyaluronic acid implant vs trabeculectomy- -a 2-year follow-up. Graefes Arch Clin Exp Ophthalmol. 2012; 250(12): 1835-1841. doi: 10.1007/s00417-012-1985-910.1007/s00417-012-1985-9 https://doi.org/10.1007/s00417-012-1985-9
- Netland PA, Sarkisian SR Jr, Moster MR, et al. Randomized, prospective, comparative trial of EX-PRESS glaucoma filtration device versus trabeculectomy (XVT study). Am J Ophthalmol. 2014; 157(2): 433-440.e3. doi: 10.1016/j.ajo.2013.09.014 https://doi.org/10.1016/j.ajo.2013.09.014
- SeiderMIRofaghaSLinSCStamperRL. Resident-performed Ex-PRESS shunt implantation versus trabeculectomy. J Glaucoma. 2012; 21(7): 469-474. doi: 10.1097/IJG.0b013e3182182bfb10.1097/ijg.0b013e3182182bfb https://doi.org/10.1097/ijg.0b013e3182182bfb
- WamsleySMosterMRRaiSAlvimHSFontanarosaJ. Results of the use of the Ex-PRESS miniature glaucoma implant in technically challenging, advanced glaucoma cases: A clinical pilot study. Am J Ophthalmol. 2004; 138(6): 1049-1051. doi: 10.1016/j. ajo.2004.06.02410.1016/j.ajo.2004.06.024 https://doi.org/10.1016/j.ajo.2004.06.024
- MoisseievEZunzETzurRKurtzSShemeshG. Standard trabeculectomy and Ex-PRESS miniature glaucoma shunt: A comparative study and literature review. J Glaucoma. 2015; 24(6): 410- 416. doi: 10.1097/IJG.000000000000004710.1097/ijg.0000000000000047 https://doi.org/10.1097/ijg.0000000000000047
- MinJKKeeCWSohnSWLeeHJWooJMYimJH. Surgical outcome of Mitomycin C-soaked collagen matrix implant in trabeculectomy. J Glaucoma. 2013; 22(6): 456-462. doi: 10.1097/ IJG.0b013e31826ab6b1
- SalimSDuHBoonyaleephanSWanJ. Surgical outcomes of the Ex-PRESS glaucoma filtration device in African American and white glaucoma patients. Clin Ophthalmol. 2012; 6: 955-962. doi: 10.2147/OPTH.S3228210.2147/opth.s32282 https://doi.org/10.2147/opth.s32282
- BissigARivierDZaninettiMShaarawyTMermoudARoyS. Ten years follow-up after deepsclerectomy with collagen implant. J Glaucoma. 2008; 17(8): 680-686. doi: 10.1097/ IJG.0b013e318182ed9e
- RavinetEBoveyEMermoudA. T-Flux implant versus Healon GV in deep sclerectomy. J Glaucoma. 2004; 13(1): 46-50. doi: 10.1097/00061198-200402000-00009 de Jong LA. The Ex-PRESS glaucoma shunt versus trabeculectomy in open-angle glaucoma: A prospective randomized study. Adv Ther. 2009; 26(3): 336-345. doi: 10.1007/s12325-009-0017-610.1097/00061198-200402000-00009 https://doi.org/10.1097/00061198-200402000-00009
- SugiyamaTShibataMKojimaSUekiMIkedaT. The first report on intermediate-term outcome of Ex-PRESS glaucoma filtration device implanted under scleral flap in Japanese patients. Clin Ophthalmol. 2011; 5: 1063-1066. doi: 10.2147/OPTH.S2312910.2147/opth.s23129 https://doi.org/10.2147/opth.s23129
- AptelFDumasSDenisP. Ultrasound biomicroscopy and optical coherence tomography imaging of filtering blebs after deep sclerectomy with new collagen implant. Eur J Ophthalmol. 2009; 19(2): 223-230. doi: 10.1177/11206721090190020810.1177/112067210901900208 https://doi.org/10.1177/112067210901900208
- Johnson MS, Sarkisian SR Jr. Using a collagen matrix implant (Ologen) versus Mitomycin-C as a wound healing modulator in trabeculectomy with the Ex-PRESS mini glaucoma device: A 12-month retrospective review. J Glaucoma. 2014; 23(9): 649-652. doi: 10.1097/IJG.0000000000000018 https://doi.org/10.1097/ijg.0000000000000018
- GargSJKanitkarKWeichelEFischerD. Trauma-induced extrusion of an Ex-PRESS glaucoma shunt presenting as an intraocular foreign body. Arch Ophthalmol. 2005; 123(9): 1270-1272. doi: 10.1001/archopht.123.9.127010.1001/archopht.123.9.1270 https://doi.org/10.1001/archopht.123.9.1270
- Maris PJ Jr, Ishida K, Netland PA. Comparison of trabeculectomy with Ex-PRESS miniature glaucoma device implanted under scleral flap. J Glaucoma. 2007; 16(1): 14-19. doi: 10.1097/01. ijg.0000243479.90403.cd https://doi.org/10.1097/01.ijg.0000243479.90403.cd
- Stein JD, Herndon LW, Brent Bond J, Challa P. Exposure of Ex-PRESS miniature glaucoma devices: Case series and technique for tube shunt removal. J Glaucoma. 2007; 16(8): 704-706. doi: 10.1097/IJG.0b013e31806ab314 https://doi.org/10.1097/ijg.0b013e31806ab314
- ShaarawyTNguyenCSchnyderCMermoudA. Comparative study between deep sclerectomy with and without colagen implant: Long term follow up. Br J Ophthalmol. 2004; 88(1): 95-98. doi: 10.1136/bjo.88.1.9510.1136/bjo.88.1.95 https://doi.org/10.1136/bjo.88.1.95
- VetrugnoMFerreriPCardasciaNSborgiaC. Surgical reposition of a dislocated Ex-PRESS miniature device. Eur J Ophthalmol. 2011; 21: 212-214. doi: 10.5301/ejo.2010.581510.5301/ejo.2010.5815 https://doi.org/10.5301/ejo.2010.5815
- RosentreterASchildAMJordanJFKrieglsteinGKDietleinTS. A prospective randomised trial of trabeculectomy using Mitomycin C vs an ologen implant in open angle glaucoma. Eye (Lond). 2010; 24(9): 1449-1457. doi: 10.1038/eye.2010.10610.1038/eye.2010.106 https://doi.org/10.1038/eye.2010.106
- CillinoSDi PaceFCillinoGCasuccioA. Biodegradable collagen matrix implant vs mitomycin-C as an adjuvant in trabeculectomy: A 24-month, randomized clinical trial. Eye (Lond). 2011; 25(12): 1598-1606. doi: 10.1038/eye.2011.21910.1038/eye.2011.219 https://doi.org/10.1038/eye.2011.219
- ShaarawyTMermoudA. Deep sclerectomy in one eye vs deep sclerectomy with collagen implant in the contralateral eye of the same patient: Long-term follow-up. Eye (Lond). 2005; 19(3): 298-302. doi: 10.1038/sj.eye.670146910.1038/sj.eye.6701469 https://doi.org/10.1038/sj.eye.6701469
- SenthilSRaoHLBabuJGMandalAKGarudadriCS. Comparison of outcomes of trabeculectomy with mitomycin C vs. ologen implant in primary glaucoma. Indian J Ophthalmol. 2013; 61(7): 338-342. doi: 10.4103/0301-4738.10952010.4103/0301-4738.109520 https://doi.org/10.4103/0301-4738.109520
- DemaillyPLavatPKretzGJeanteur-LunelMN. Non-penetrating deep sclerectomy (NPDS) with or without collagen device (CD) in primary open-angle glaucoma: Middle-term retrospective study. Int Ophthalmol. 1996-1997; 20(1-3): 131-140
- NilforushanNYadegariMHodjatP. Comparison of the success rate of trabeculectomy with oculusgen versus trabeculectomy with Mitomycin C. Iranian Journal of Ophthalmology. 2011; 23(4): 3-12
- EgrilmezSAtesHNalcaciSAndacKYagciA. Surgically induced corneal refractive change following glaucoma surgery: Nonpenetrating trabecular surgeries versus trabeculectomy. J Cataract Refract Surg. 2004; 30(6): 1232-1239. doi: 10.1016/j.jcrs.2003.11.05510.1016/j.jcrs.2003.11.055 https://doi.org/10.1016/j.jcrs.2003.11.055
- Beltran-AgulloLTropeGEJinYWagschalLDJinapriyaDBuysYM. Comparison of visual recovery following EX-PRESS versus trabeculectomy: Results of a prospective randomized controlled trial. J Glaucoma. 2015; 24(3): 181-186. doi: 10.1097/ IJG.0b013e31829e1b68
- PatelHYWagschalLDTropeGEBuysYM. Economic analysis of the Ex-PRESS miniature glaucoma device versus trabeculectomy. J Glaucoma. 2014; 23(6): 385-390. doi: 10.1097/ IJG.0b013e31827a06f4
- MareyHMMandourSSEllakwaAF. Subscleral trabeculectomy with Mitomycin-C versus ologen for treatment of glaucoma. J Ocul Pharmacol Ther. 2013; 29(3): 330-334. doi: 10.1089/ jop.2012.0120
- DahanERavinetEBen-SimonGJMermoudA. Comparison of the efficacy and longevity of nonpenetrating glaucoma surgery with and without a new, nonabsorbable hydrophilic implant. Ophthalmic Surg Lasers Imaging. 2003; 34(6): 457-463
- D'EliseoDPastenaBLonganesiLGrisantiFNegriniV. Comparison of deep sclerectomy with implant and combined glaucoma surgery. Ophthalmologica. 2003; 217(3): 208-211. doi: 10.1159/00006897410.1159/000068974 https://doi.org/10.1159/000068974
- DindoDClavienPA. What is a surgical complication? World J Surg. 2008; 32: 939-941. doi: 10.1007/s00268-008-9584-y10.1007/s00268-008-9584-y https://doi.org/10.1007/s00268-008-9584-y
- ClavienPABarkunJde OliveiraML. The clavien-dindo classification of surgical complications: Five-year experience. Ann Surg. 2009; 250: 187-196. doi: 10.1097/SLA.0b013e3181b13ca210.1097/sla.0b013e3181b13ca2 https://doi.org/10.1097/sla.0b013e3181b13ca2
- HazellLShakirSA. Under-reporting of adverse drug reactions: A systematic review. Drug Saf. 2006; 29(5): 385-396. doi: 10.2165/00002018-200629050-0000310.2165/00002018-200629050-00003 https://doi.org/10.2165/00002018-200629050-00003