The Ophthalmology Open Journal · Vol. 3, No. 1 · 2019 · pp. 10–20 · e-ISSN 2475-1278

Complications Related to Implants Used in Anterior Bleb Forming Glaucoma Surgery: A Systematic Review of the Literature

Alejandra Salinas, Fabián Rodríguez, Javier Bardavío

Department of Ophthalmology, Hospital Universitario de Torrevieja, SpainSpain
Department of Ophthalmology, Hospital Verge de la Cinta,Tortosa, SpainSpain
St Mary’s Treatment Centre, Portsmouth, UKUnited Kingdom

DOI: https://doi.org/10.17140/ooj-3-118 · Published 2019-10-11 · Licence: https://creativecommons.org/licenses/by/4.0 · © 2019 The Author(s)

Abstract

Purpose To make an account of published implant-related complications (IRC) by a systematic review of the literature. Methods A systematic search of Pubmed and Scopus databases and Google Scholar engine was performed with selection criteria to detect papers on IRC. We excluded unrelated papers and reviewed selected ones. We considered papers that did not explicitly state about occurrence or not occurrence of IRC as non-IRC reporting. Main outcome measures were the number of papers reporting on complications, IRC, and types of IRC. Results After the search, selection, and addition, we studied 109 papers. Incidence of IRC was 4.5%, half required explantation. While 26 implant studies found IRCs (23%), 13 case reports on surgical complications, 8 (61.5%) of them reported IRC. Frequent complications were conjunctival erosion, blockage of the tube, migration to anterior chamber or damage to surrounding tissues. Conclusion Most papers did not report on IRC. Length or nature of studies may skew finding IRC. The incidence of IRC was 4.5%. Hard and sharp implants carry a greater risk of IRC and explantation.

Keywords: Implant-related complications (IRC); Glaucoma surgery; Anterior bleb forming

Methods

A systematic search of Pubmed and Scopus databases and Google Scholar engine was performed with selection criteria to detect

papers on IRC. We excluded unrelated papers and reviewed selected ones. We considered papers that did not explicitly state about

occurrence or not occurrence of IRC as non-IRC reporting. Main outcome measures were the number of papers reporting on

complications, IRC, and types of IRC.

Results

After the search, selection, and addition, we studied 109 papers.

26 implant studies found IRCs (23%), 13 case reports on surgical complications, 8 (61.5%) of them reported IRC. Frequent

complications were conjunctival erosion, blockage of the tube, migration to anterior chamber or damage to surrounding tissues.

Conclusion

Most papers did not report on IRC. Length or nature of studies

and sharp implants carry a greater risk of IRC and explantation.

Keywords

Implant-related complications (IRC); Glaucoma surgery; Anterior bleb forming.

Introduction

laucoma is a chronic progressive optic neuropathy that may

cause severe visual loss and affect the quality of life. The G front line of treatment consists of reducing intraocular pressure.

Most patients achieve good medical control: some require one or

more surgical procedures. Although conventional filtering surgery

is effective has only moderate success and are subject to risks.

Some devices have been designed and introduced to improve suc-

cess and risks. Some of these implants are associated with classical

conventional fistulizing surgery or to newer filtering procedures

cc Copyright 2019 by SalinasA. This is an open-access article distributed under Creative Commons Attribution 4.0 International License (CC BY 4.0), which allows Incidence of IRC was 4.5%, half required explantation. While

may skew finding IRC. The incidence of IRC was 4.5%. Hard

that intend the construction of a paralimbal bleb (subconjuncti-

val, Subtenon’s or intra-scleral). Regulatory bodies have studied

the security and effectiveness of these implants and authorized

for human application. Some systematic reviews or meta-analysis 1-4 show that support for the use of these devices may be limitated.

This study aimed to make an account of published implant-related

complications (IRC), this is complications directly caused by the

implant, by a systematic review of the literature up to April 2015.

Only implants that contribute to creating a paralimbal bleb were

considered since those are the ones that are used as an alternative

to trabeculectomy or non-penetrating deep sclerectomy and are

potentially avoidable. We excluded equator bleb creating implants,

angular devices or anterior chamber to supraciliary space shunts

from the scope of this study. We were also interested to know

how many of the papers referred to complications related to the

implants.

Methods

Search Strategy

Three different search engines were used: Pubmed, Google Schol-

ar and Scopus in April 2015. The search was performed over all

fields of bibliographic references. We applied three levels of search

restriction and combined controlled Mesh and natural languages.

Language filters were applied. Inclusion criteria: any series of pa-

tients, whether prospective or retrospective and case reports in

English, Spanish or French (filtered) found introducing the se-

lected keywords. The word “complication” was extended to the

abstract search also. Exclusion criteria: reports on other glaucoma

Table 1. Search Strategy Details

# Searches on Google Scholar Results

((implant* OR collagen implant OR collagen implants OR “Hemacrylic Metha bar”) AND complication) AND (glaucoma)

(implant* OR esnoper OR skgel OR aquaflow OR “ex-press” OR “collagen implant” OR “collagen implants” OR “Hemacrylic” OR “reticulated 2 hyaluronic acid” OR “T flux” OR “T bar”) AND complication* AND Glaucoma-“Molteno Implants”-Molteno-Ahmed-“Krupin valve”-“krupin 2300 valves”-Baerveldt)

(“Filtering Surgery” OR “Trabeculectomy” OR “Filtering Surgeries” OR OR “deep sclerectomy” OR “Sclerostomy” OR “deep sclerectomies” OR viscocanalostomy) AND (implant* OR esnoper OR skgel OR aquaflow OR “ex-press” OR “collagen implant” OR “collagen implants” OR “Hemacrylic” OR “reticulated hyaluronic acid” OR “T flux” OR “T bar”) AND complication* AND Glaucoma-“Molteno Implants”-Molteno- Ahmed-“Krupin valve”-“krupin valves”-Baerveldt)

# Searches on Scopus

(TITLE(implant* OR esnoper OR sk gel OR aqua flow OR “express implant*” OR “collagen implant*” OR “Hemacrylic Metha Stealth” OR “reticulated hyaluronic acid” OR “T flux” OR “T bar”) OR ABS(implant* OR esnoper OR sk gel OR aqua flow OR “express implant*” OR 4 “collagen implant*” OR “Hemacrylic Metha Stealth” OR “reticulated hyaluronic acid” OR “T flux” OR “T bar”)) AND (TITLE(complication*) 1533 ORABS(complication*)) AND (TITLE(glaucoma) OR ABS(glaucoma)) AND (LIMIT-TO(LANGUAGE, “English”) OR LIMIT-TO(LANGUAGE, “French”) OR LIMIT-TO(LANGUAGE, “Spanish”))

(TITLE(implant* OR esnoper OR skgel OR aquaflow OR “express implant*” OR “collagen implant*” OR “Hemacrylic Metha Stealth” OR “reticulated hyaluronic acid” OR“T flux” OR “T bar”) OR ABS(implant* OR esnoper OR skgel OR aquaflow OR “express implant*” OR “col- 5 lagen implant*” OR “Hemacrylic Metha Stealth” OR “reticulated hyaluronic acid” OR“T flux” OR “T bar”)) AND (TITLE(complication*) OR 1106 ABS(complication*)) AND (TITLE(glaucoma) OR ABS(glaucoma)) AND NOT baerveldt) AND (LIMIT-TO(LANGUAGE,“English”) OR LIMIT-TO(LANGUAGE,“French”) OR LIMIT-TO(LANGUAGE,“Spanish”))

(TITLE(“Trabeculectom*” OR “Filtering Surger*” OR “deep sclerectom*” OR “Sclerostomy” OR viscocanalostom*) OR ABS(“Trabeculectom*” OR “Filtering Surger*” OR “deep sclerectom*” OR “Sclerostomy” OR viscocanalostom*)) AND (TITLE(implant* OR esnoper OR skgel OR aquaflow OR “express implant*” OR “collagen implant*” OR “Hemacrylic Metha Stealth” OR “reticulated hyaluronic 6 acid” OR“T flux” OR “T bar”) OR ABS(implant* OR esnoper OR skgel OR aquaflow OR “express implant*” OR “collagen implant*” OR 273 “Hemacrylic Metha Stealth” OR “reticulated hyaluronic acid” OR“T flux” OR “T bar”)) AND (TITLE(complication*) OR ABS(complication*)) AND (TITLE(glaucoma) OR ABS(glaucoma)) AND NOT TITLE-ABS-KEY(Molteno* OR Ahmed OR“Krupin valve*” OR baerveldt) AND(LIMIT- TO(LANGUAGE, “English”)ORLIMIT-TO(LANGUAGE, “French”) OR LIMIT-TO(LANGUAGE, “Spanish”))

# Searches on Pubmed/Medline

(“Prostheses and Implants/complications”[Mesh] OR ((implant*[tiab] OR esnoper[tiab] OR skgel[tiab] OR aquaflow[tiab] OR “express implant”[tiab] OR “express implants”[tiab] OR “collagen implant”[tiab] OR “collagen implants”[tiab] OR “HemacrylicMetha Stealth”[tiab] OR “reticulated hyaluronic acid”[tiab] OR“T flux”[tiab] OR “T bar”[tiab]) AND complication*[tiab]))) AND (“Glaucoma”[Mesh] OR glaucoma*[tiab])

(TITLE(implant* OR esnoper OR skgel OR aquaflow OR “express implant*” OR “collagen implant*” OR “Hemacrylic Metha Stealth” OR “reticulated hyaluronic acid” OR“T flux” OR “T bar”) OR ABS(implant* OR esnoper OR skgel OR aquaflow OR “express implant*” OR “col- 8 lagen implant*” OR “Hemacrylic Metha Stealth” OR “reticulated hyaluronic acid” OR“T flux" OR “T bar”)) AND (TITLE(complication*) OR 967 ABS(complication*)) AND (TITLE(glaucoma) OR ABS(glaucoma)) AND NOT baerveldt) AND(LIMIT-TO(LANGUAGE, “English”) OR LIMIT-TO(LANGUAGE, “French”) OR LIMIT-TO(LANGUAGE, “Spanish”))

(TITLE(“Trabeculectom*” OR “Filtering Surger*” OR “deep sclerectom*” OR “Sclerostomy” OR viscocanalostom*) OR ABS(“Trabeculectom*” OR “Filtering Surger*” OR “deep sclerectom*” OR “Sclerostomy” OR viscocanalostom*)) AND (TITLE(implant* OR esnoper OR skgel OR aquaflow OR “express implant*” OR “collagen implant*” OR “Hemacrylic Metha Stealth” OR “reticulated hyaluronic acid” OR“T flux” OR “T 9 bar”) OR ABS(implant* OR esnoper OR skgel OR aquaflow OR “express implant*” OR “collagen implant*” OR “Hemacrylic Metha Stealth” 314 OR “reticulated hyaluronic acid” OR“T flux” OR “T bar”)) AND (TITLE(complication*) OR ABS(complication*)) AND (TITLE(glaucoma) OR ABS(glaucoma)) AND NOT TITLE-ABS-KEY(Molteno* OR Ahmed OR“Krupin valve*” OR baerveldt) AND (LIMIT-TO(LANGUAGE, “English”) ORLIMIT-TO(LANGUAGE, “French”)ORLIMIT-TO(LANGUAGE, “Spanish”)) drainage devices like Ahmed or Krupin Valve, Baerveltd or Mol-

teno drainage devices. Any anterior chamber to supraciliary or su-

prachoroidal shunts like Gold Micro Shunt or Esnoper V-2000. Pa-

pers in any other languages. Keywords: “prostheses and implants/

complications”, “glaucoma drainage implants”, implant, ex-press,

esnoper, SK gel, aqua flow, collagen implant, Hem Acrylic, reticu-

lated hyaluronic acid, T flux, thiobarbituric acid-reactive (T BAR),

filtering surgery, filtering surgeries,deep sclerectomy, non-penetrat-

ing deep sclerectomy, viscocalanostomy, trabeculectomy, scleros-

tomy. Mendeley reference manager (www.mendeley.com,5 Elsevier)

detected duplications. Refer to Table 1 for search strategy.

Initial Revision Strategy:The Selection of Papers for Revision

All papers’ titles were reviewed by a team of trained or in training

ophthalmologist in search of papers that reported on paralimbal

bleb forming surgical procedures with the implantation of a de-

vice. When in doubt, we reviewed the abstract. We excluded papers

reporting on surgical procedures without an implant, different pro-

Stealth” OR “reticulated hyaluronic acid” OR “T flux” OR “T 11600

1572

TITLE-ABS-KEY(Molteno* OR Ahmed OR“Krupin valve*” OR

1249

TITLE-ABS-KEY(Molteno* OR Ahmed OR“Krupin valve*” OR

cedures all together (posterior forming blebs, angular surgery, an-

terior chamber to supraciliary or suprachoroidal shunts), different

languages or animal studies. We included systematic reviews and

meta-analysis of implanting surgical procedures studies to secure

relevant studies that our search could have missed out, but care-

fully made sure we did not duplicate the study papers by doing so.

Full text of all included papers was requested for revi-

sion. New references were found during revision and added under

the same criteria.

Ethical Committee approval was not requested as no pa-

tient information was searched at any time.

Outcome Measures

The following outcomes were measured: total number of papers

after exclusions, number of papers that report complications, num-

ber of papers that report IRC, type of implant, type of IRC found

as extrusion, migration, damage to surrounding tissue, dysfunc-

tional implants, explanted cases, relocated cases, other (Table 2).

Table 2. Definition of Outcomes to be Measured

Outcomes to be Measured

• Number of papers found once the non-related are excluded • Number of papers that report complications • Number of papers that report implant-related complications • Type of implant-related complications found with every implant o Extrusion o Migration o Damage of surrounding tissue o Obstructed o Explanted cases o Relocated cases o Other

Papers’revision was directed to find a statement on com-

plications within the method and the results sections of the ar-

ticle. We did not look for complications reports within the dis-

Figure 1. Final Number of Papers Considered for Study Purposes cussion section nor conclusions of the papers. Since the presence or absence of complications report is not interpretable, only one reviewer read through every paper. A paper was considered to re- port on IRC if the actual complication was reported or it explicitly stated that no IRC was found during the study. Papers that did not

state one thing or the other were considered not to report on IRC.

In case of doubt, the paper was consulted with the senior ophthal-

mologist in the study group.

A register of complications was established to include

both studies with and without complications’ reports.

Results

The result from the search of the three engines was 253 papers

from Scopus, 314 from Pubmed and 1572 from Google Scholar

that were reduced to 1102 after removing redundant papers. All ti-

tles of those papers were reviewed to discard unrelated ones when

in doubt the abstract was used. Finally, 120 full-text articles were

requested for review. From systematic revisions and meta-analysis,

we found 28 new unknown references that were also requested.

From a total of 148 papers, 23 were excluded due to belonging to

animal or experimental implants studies, reporting on the different

procedure, German language, or unfound full text. Of the 125 reviewed papers, 9 were systematic reviews, and 7 were meta-analysis 1-16 that included studies already reviewed by this research team.

Therefore, the final number of papers considered for study pur- 17-126 poses was 109 (Figure 1).

All these papers correspond to studies or case reports that

report on glaucoma implant surgery and also on surgical complica-

tions (100% of the 109 papers reviewed). Out of the 109 papers reviewed, 75 (68.8%) did not report on IRC, 34 (31.2%) reported

on IRC, 6 (17.6%) of the IRC reporting papers) explicitly report-

ing an absence of IRC (4 prospective and 2 retrospective studies).

Of the 59 prospective studies (54.1%, of 109), 17 reported on IRC

(28.8%, of 59). Of the 37 retrospective studies (33.9%, of 109), 9

(24.3%, of 37) reported IRC. Thirteen of the papers (11.9%, of

Table 3. Understudy Implants, IRC Reporting Papers, Number of Study Eyes (Eyes with Implants in IRC Reporting Papers) and Number of Complicated Cases

Implant Nº of Papers* Nº IRC Papers** Nº of Study Eyes Nº Complicated Cases

Express (Steel) 38 22 (57,9%) 940 54 (5,7%)

Collagen Implant 25 5 (16,6%) 208 9 (4,3%)

SK-gel (Hyualuronic acid) 25 3 (12%) 185 2 (1,07%)

T-flux (Acrylic) 15 1 (7,14%) 25 1 (4%)

Ethicon 1/0 Chromic Catgut 1 1 (100%) 23 1 (4,3%)

PMMA 1 1

ePTFE 1 1

Ologen (Collagen) 9 1

HEALA Flow 1 0

Aquaflow (Collagen) 5 0

Total 109 35

*Twelve papers tested two or three implant devices; **One paper tested two devices PMMA and Collagen implant

Table 4. Reported Complications

Implant Complication

Migration or 1 migration to malposition subconjunctival space

1 conjunctival erosion and extrusion that did Extrusion not require explants or repositioning

Damage to 1 Corneal abrasion surrounding 5 Descemet mb.

tissues detachment

Obstructed

Explanted During hypotony

Explanted as treatment of complication

Repositioned

109) were case reports, 8 (61.5%, of 13) reported IRC.

The total population of eyes considered in this study was

5831 (population range per study: 1-345), 4576 of which had been

implanted (78.4%), the others belonging to control groups. The

population of eyes with an implant included in studies that report-

ed on IRC was 1479. Of those, the number of eyes that suffered

an IRC was 67 (4.5%). Table 3 shows the implants understudied,

the number of IRC reporting papers, the number of study eyes

and the number of complicated cases.

A summary of the complications is in Table 4. The most

common IRC was conjunctival erosion; 20 cases after Express im-

plantation (15 through a full-thickness scleral incision, four under 30 0

35 0

33 0

0 0

0 0

1479 67 (4,5%)

Ethicon 1/0 chromic SK-gel Collagen Implant T-flux Express catgut

1 misplacement 1 migration 1 into anterior 1 migration to subconjunctival into anterior chamber after blunt space chamber trauma 5 into anterior chamber

20 conjunctiva erosions (15 without scleral flap,4 with sclera flap,1 unknown)

5 iris touch 2 endothelial touch 1 Corneal abrasion 1 crystalline lens touch 1 scleral flap erosion only

2 unreported reasons 1 due to infection repair surgery

3 relocated due to other complication Other causes

a partial-thickness scleral flap and one unreported technique) that

required implant removal, and one case of SK-gel that healed un-

der observation. Obstruction of Express tube occurred in 14 cas-

es. Four Express tubes migrated postoperatively into the anterior

chamber (AC) and another one was introduced in the AC during

the procedure. Another Express tube migrated to the subconjunc-

tival space, as it occurred to an SK gel implant, and a further Ex-

press tube was found misplaced. A T flux implant migrated into

the patient’s anterior chamber. A fragment of Ethicon 1/0 chromic

catgut, placed under a deep sclerectomy flap, migrated into de AC

also. Sixteen cases of damage to surrounding tissues were found, 5

Descemet membrane detachments and one corneal abrasion after

collagen implant, five iris touch, two endothelial touches, one cor-

neal abrasion, one crystalline lens touch that induced a white cata-

ract and one scleral flap extrusion, without conjunctival erosion,

after Express implantation. Three primary explantation of Express

tubes, one due to a bleb infection and two for unreported reasons,

as well as three collagen implant removal during the hypotony re-

pair. A total of 29 Express tubes (3% of all tubes implanted) and

three collagen implants (1.4% of all implanted) were explanted.

Discussion

The market offers implants in an attempt to improve post-operative

survival, efficacy or safety for patients requiring glaucoma surgery.

Some studies, however, suggest that traditional trabeculectomy or

newer deep sclerectomy without implants might be just as effective

in reducing intraocular pressure. Devices, therefore, attempt to im-

prove prognosis, but its mere physical presence may also increase

risks. This study aimed to make an account of published IRC by a

systematic review of the literature up to April 2015. We only stud-

ied implants that contribute to creating a paralimbal bleb. Equator

bleb creating implants, angular devices or anterior chamber to su-

praciliary space shunts are not within the scope of this study.

Our search produced a total of 1102 papers. The use of

natural language and mesh terms explain such a large number of

papers found. The initial selection resulted in 120 papers of which

23 were further excluded, and another 28 were added during revi-

sion. Seven systematic reviews, nine meta-analysis, 13 case reports,

37 retrospective and 59 prospective studies were reviewed. Only

an explicit account of IRC was considered as a report. All papers

reviewed (109/109) reported surgical complications. Only 31.2%

reported on IRC: 6 (17.6%) of those (4 prospective and two ret-

rospectives), reported an absence of IRC. Only 28.8% of all pro-

spective and 24.4% of all retrospective studies reported on IRC.

It was therefore interesting to find that of 13 case reports dealing

with complications, 8 (61.5%) were about IRC. We suspect that the

discrepancy in IRC incidence reported in retrospective and pro-

spective studies compared to case reports might be related to the

length of the studies and although inconclusive, could be relevant.

Issues related to reporting could justify further research.

Sixty-seven cases of IRC of 1479 (4.5%) patients were

found, of which 31 required implant removal. Our estimation

of IRC incidence might be biased by the search or the selection

method. The incidence of Express removal was 3% before the

tube was guarded under a scleral flap: that reduced the risk but did

not eliminate it. The initial design of the technique may show that

the risk of extrusion was underestimated. Risk of misplacement

or injury to surrounding tissues like Descemet membrane, corneal

endothelium or the crystalline lens is also relevant as well as migra-

tion into the anterior chamber of mainly the express tubes.

The physical properties of the implant are relevant: hard-

er and sharper implants that remain longer, increase the risk of

complications, mainly extrusion. In our study, the steel implant Ex-

press accounts for most cases of explanted, extruded or migrated

cases. Also, non-absorbable acrylic T flux tends to migrate if left

unfixed. Soft and absorbable implants tend to be more respect-

ful. We speculated that tensile vectors generated by scaring tissue might tend to move rigid pieces and to fold soft material: this could

explain the migration of implants or simple orientation changes.

The intermittent robbing effect of the eyelid over the conjunctiva

exerts an external pressure (tissue towards implant) that injures the

tissue in the opposite direction (implant towards sclera, conjunc-

tiva or both). Both mechanisms are active in all operated eyes, so

the risk to induce complications exists.

Study limitations include missing relevant papers by the

search (no excerpta medica dataBASE (EMBASE) search) or the

selection method: the initial search failed to detect 28 relevant pa-

pers added later. It is possible that relevant data from studies was

filtered out by title or abstract selection. A further limitation is late

publication of this study that was finalized in 2016 and sent for

publication in 2019. Finally, this review relies on what authors re-

port on their papers, under or overreporting issues would modify

these results. However, the study remains useful to stress the fact

that an implant might induce complications by itself. Also, the type

and shape of the material are relevant and worth bearing in mind

at the time of assessing the purpose and usefulness of the implant.

Further, the importance of considering IRC at the time of design-

ing studies to make them sensible to complication detection and

reporting.

Conclusion

An IRC incidence of 4.5% resulted from this study, of which

about a half required removal. Harder and sharper implants carry

a higher risk. It is worth considering the implant’s function and

the risk it may carry its implantation before making the clinical

decision of using them. Studies should consider this risk and be

methodologically designed to detect them.

Acknowledgements

The authors are thankful to Ms. Noemi Aluja for her contribution

towords the bibliography.

Conflicts Of Interest

The authors declare that they have no conflicts of interest.

References

  1. ChengJWChengSWCaiJPLiYWeiRL. Systematic overview of the efficacy of nonpenetrating glaucoma surgery in the treatment of open angle glaucoma. Med Sci Monit. 2011; 17(7): RA155-RA163. doi: 10.12659/MSM.88184010.12659/msm.881840 https://doi.org/10.12659/msm.881840
  2. WangWZhangX. Meta-analysis of randomized controlled trials comparing EX-PRESS implantation with trabeculectomy for open-angle glaucoma. PLoS One. 2014; 9(6): e100578. doi: 10.1371/ journal.pone.0100578
  3. RulliEBiagioliERivaI. Efficacy and safety of trabeculectomy vs nonpenetrating surgical procedures: A systematic review and meta-analysis. JAMA Ophthalmol. 2013; 131(12):1573-1582. doi: 10.1001/jamaophthalmol.2013.505910.1001/jamaophthalmol.2013.5059 https://doi.org/10.1001/jamaophthalmol.2013.5059
  4. ChengJWWeiRLCaiJPLiY. Efficacy and tolerability of non-penetrating filtering surgery with and without implant in treatment of open angle glaucoma: A quantitative evaluation of the evidence. J Glaucoma. 2009; 18(3): 233-237. doi: 10.1097/ IJG.0b013e3181812827
  5. Mendeley. Web site. www.mendeley.com. Accessed September 16, 2019
  6. ChenGLiWJiangFMaoSTongY. Ex-PRESS implantation versus trabeculectomy in open-angle glaucoma: A meta-analysis of randomized controlled clinical trials. PLoS One. 2014; 9(1): e86045. doi: 10.1371/journal.pone.008604510.1371/journal.pone.0086045 https://doi.org/10.1371/journal.pone.0086045
  7. BuysYM. Trabeculectomy versus Ex-PRESS shunt surgery in residency training. Surv Ophthalmol. 2012; 57(4): 375-378. doi: 10.1016/j.survophthal.2012.02.00610.1016/j.survophthal.2012.02.006 https://doi.org/10.1016/j.survophthal.2012.02.006
  8. DemeterSHaileyD. Non-penetrating glaucoma surgery using AquaFlow(TM)collagen implants. Issues Emerg Health Technol. 2001; (22): 1-4
  9. FilippopoulosTRheeDJ. Novel surgical procedures in glaucoma: Advances in penetrating glaucoma surgery. Curr Opin Ophthalmol. 2008; 19(2): 149-154. doi: 10.1097/ICU.0b013e3282f4f49e10.1097/icu.0b013e3282f4f49e https://doi.org/10.1097/icu.0b013e3282f4f49e
  10. FrancisBASinghKLinSC. Novel glaucoma procedures: A report by the American Academy of Ophthalmology. Ophthalmology. 2011; 118(7): 1466-1480. doi: 10.1016/j.ophtha.2011.03.02810.1016/j.ophtha.2011.03.028 https://doi.org/10.1016/j.ophtha.2011.03.028
  11. HendrickAMKahookMY. Ex-PRESS mini glaucoma shunt: Surgical technique and review of clinical experience. Expert Rev Med Devices. 2008; 5(6): 673-677. doi: 10.1586/17434440.5.6.67310.1586/17434440.5.6.673 https://doi.org/10.1586/17434440.5.6.673
  12. SalimS. The role of the Ex-PRESS glaucoma filtration device in glaucoma surgery. Semin Ophthalmol. 2013; 28(3): 180-184. doi: 10.3109/08820538.2013.77119710.3109/08820538.2013.771197 https://doi.org/10.3109/08820538.2013.771197
  13. BuysYM. Trabeculectomy with Ex-PRESS: Weighing the benefits and cost. Curr Opin Ophthalmol. 2013; 24(2): 111-118. doi: 10.1097/ICU.0b013e32835907a610.1097/icu.0b013e32835907a6 https://doi.org/10.1097/icu.0b013e32835907a6
  14. SalimS. Ex-PRESS glaucoma filtration device: Surgical technique and outcomes. Int Ophthalmol Clin. 2011; 51(3): 83-94. doi: 10.1097/IIO.0b013e31821e5b8210.1097/iio.0b013e31821e5b82 https://doi.org/10.1097/iio.0b013e31821e5b82
  15. ShaarawyTGoldbergIFechtnerR. EX-PRESS glaucoma filtration device: Review of clinical experience and comparison with trabeculectomy. Surv Ophthalmol. 2015; 60: 327-345. doi: 10.1016/j. survophthal.2015.01.00110.1016/j.survophthal.2015.01.001 https://doi.org/10.1016/j.survophthal.2015.01.001
  16. JiQQiBLiuLGuoXZhongJ. Efficacy and safety of ologen implant versus Mitomycin C in primary trabeculectomy: A meta-analysis of randomized clinical trials. J Glaucoma. 2015; 24: e88-e94. doi: 10.1097/IJG.000000000000014710.1097/ijg.0000000000000147 https://doi.org/10.1097/ijg.0000000000000147
  17. Al Obeidan SA. Migration of reticulated hyaluronic acid implant [SK-GEL] following deep sclerectomy. Int Ophthalmol. 2010; 30(3): 329-331. doi: 10.1007/s10792-009-9325-8 https://doi.org/10.1007/s10792-009-9325-8
  18. Al Obeidan SA, Osman EA, Al-Muammar AM, Abu El-Asrar AM. Efficacy and safety of deep sclerectomy in uveitic glaucoma. Int Ophthalmol. 2009; 29(5): 367-372. doi: 10.1007/s10792-008- 9247-x https://doi.org/10.1007/s10792-008-9247-x
  19. Al-ObeidanSAMousaANaseemAAbu-AmeroKKOsmanEA. Efficacy and safety of non-penetrating deep sclerectomy surgery in Saudi patients with uncontrolled open angle glaucoma. Saudi Med J. 2013; 34(1): 54-61
  20. ArruabarrenaCMuñoz-NegreteFJMárquezCRebolledaG. Results of nonpenetrating deep sclerectomy in inflammatory glaucoma: One year follow up. Arch Soc Esp Oftalmol. 2007; 82(8): 483-487
  21. AtesHAndacKUretmenO. Non-penetrating deep sclerectomy and collagen implant surgery in glaucoma patients with advanced field loss. Int Ophthalmol. 1999; 23(3): 123-128
  22. AuerCMermoudAHerbortCP. Deep sclerectomy for the management of uncontrolled uveitic glaucoma: Preliminary data. Klin Monbl Augenheilkd. 2004; 221(5): 339-342. doi: 10.1055/s- 2004-81282010.1055/s-2004-812820 https://doi.org/10.1055/s-2004-812820
  23. CillinoSZeppaLDi PaceF. E-PTFE (Gore-Tex) implant with or without low-dosage mitomycin-C as an adjuvant in penetrating glaucoma surgery: 2 year randomized clinical trial. Acta Ophthalmol. 2008; 86(3): 314-321. doi: 10.1111/j.1600- 0420.2007.01036.x10.1111/j.1600-0420.2007.01036.x https://doi.org/10.1111/j.1600-0420.2007.01036.x
  24. Collignon-BrachJRobe-CollignonN. Non perforating trabecular surgery with reticulated hyaluronic acid implant. Bull Soc Belge Ophtalmol. 2000; 276: 61-68
  25. CoupinALiQRissI. Ex-PRESSTM miniature glaucoma implant inserted under a scleral flap in open-angle glaucoma surgery: A retrospective study. J Fr Ophtalmol. 2007; 30(1): 18-23
  26. DahanECarmichaelTR. Implantation of a miniature glaucoma device under a scleral flap. J Glaucoma. 2005; 14(2): 98-102. doi: 10.1097/01.ijg.0000151688.34904.b710.1097/01.ijg.0000151688.34904.b7 https://doi.org/10.1097/01.ijg.0000151688.34904.b7
  27. DemaillyPJeanteur-LunelMNBerkaniM. Non-penetrating deep sclerectomy combined with a collagen implant in primary open-angle glaucoma. Medium-term retrospective results. J Fr Ophtalmol. 1996; 19(11): 659-666
  28. Detry-MorelM. Non penetrating deep sclerectomy (NPDS) with SKGEL implant and/or 5-fluorouracile (5-FU). Bull Soc Belge Ophtalmol. 2001; (280): 23-32
  29. Detry-MorelMDetryMB. Five-year experience with non penetrating deep sclerectomy. Bull Soc Belge Ophtalmol. 2006; (299): 83- 94
  30. Detry-MorelMPourjavanSDetryMB. Comparative safety profile between “modern” trabeculectomy and non-penetrationg deep sclerectomy. Bull Soc Belge Ophtalmol. 2006; (300): 43-54
  31. DrolsumL. Conversion from trabeculectomy to deep sclerectomy. Prospective study of the first 44 cases. J Cataract Refract Surg. 2003; 29(7): 1378-1384. doi: 10.1016/S0886-3350(03)00013-010.1016/s0886-3350(03)00013-0 https://doi.org/10.1016/s0886-3350(03)00013-0
  32. DupasBFardeauCCassouxNBodaghiBLeHoangP. Deep sclerectomy and trabeculectomy in uveitic glaucoma. Eye (Lond). 2010; 24(2): 310-314. doi: 10.1038/eye.2009.8210.1038/eye.2009.82 https://doi.org/10.1038/eye.2009.82
  33. ElginUSimsekTBatmanA. Use of the ex-press miniature glaucoma implant in a child with Sturge-Weber syndrome. J Pediatr Ophthalmol Strabismus. 2007; 44(4): 248-250
  34. HamardPPlazaLKopelJQuesnotSHamardH. Deep nonpenetrating sclerectomy and open angle glaucoma. Intermediate results from the first operated patients. J Fr Ophtalmol. 1999; 22(1): 25-31
  35. HamelMShaarawyTMermoudA. Deep sclerectomy with collagen implant in patients with glaucoma and high myopia. J Cataract Refract Surg. 2001; 27(9): 1410-1417. doi: 10.1016/s0886- 3350(01)00959-210.1016/s0886-3350(01)00959-2 https://doi.org/10.1016/s0886-3350(01)00959-2
  36. JungkimSGibranSKKhanKMurrayA. External trabeculectomy with T-Flux implant. Eur J Ophthalmol. 2006; 16(3): 416- 421. doi: 10.1177/11206721060160030910.1177/112067210601600309 https://doi.org/10.1177/112067210601600309
  37. KarlenMESanchezESchnyderCCSickenbergMMermoudA. Deep sclerectomy with collagen implant: Medium term results. Br J Ophthalmol. 1999; 83(1): 6-11. doi: 10.1136/bjo.83.1.610.1136/bjo.83.1.6 https://doi.org/10.1136/bjo.83.1.6
  38. KimCYChangHRLeeJHHongYJ. Surgical outcomes of deep sclerectomy with collagen implant. Korean J Ophthalmol. 2001; 15(2): 107-112. doi: 10.3341/kjo.2001.15.2.10710.3341/kjo.2001.15.2.107 https://doi.org/10.3341/kjo.2001.15.2.107
  39. KimCYHongYJSeongGJKohHJKimSS. Iris synechia after laser goniopuncture in a patient having deep sclerectomy with a collagen implant. J Cataract Refract Surg. 2002; 28(5): 900-902. doi: 10.1016/s0886-3350(01)01307-410.1016/s0886-3350(01)01307-4 https://doi.org/10.1016/s0886-3350(01)01307-4
  40. LeszczyńskiRGierek-CiaciuraSForminśka-KapuścikMMrukwa-KominekERokita-WalaI. Nonpenetrating very deep sclerectomy with reticulated hyaluronic acid implant in glaucoma treatment. Med Sci Monit. 2008; 14(2): CR86-CR89
  41. LiJYLimMCMannisMJ. Traumatic cataract associated with mini glaucoma shunt. J Cataract Refract Surg. 2011; 37(7): 1360- 1362. doi: 10.1016/j.jcrs.2011.04.01910.1016/j.jcrs.2011.04.019 https://doi.org/10.1016/j.jcrs.2011.04.019
  42. LükeCDietleinTSJacobiPCKonenWKrieglsteinGK. A prospective randomised trial of viscocanalostomy with and without implantation of a reticulated hyaluronic acid implant (SKGEL) in open angle glaucoma. Br J Ophthalmol. 2003; 87(5): 599-603. doi: 10.1136/bjo.87.5.59910.1136/bjo.87.5.599 https://doi.org/10.1136/bjo.87.5.599
  43. MansouriKShaarawyTWedrichAMermoudA. Comparing polymethylmethacrylate implant with collagen implant in deep sclerectomy: A randomized controlled trial. J Glaucoma. 2006; 15(3): 264-270. doi: 10.1097/01.ijg.0000212211.33265.6d10.1097/01.ijg.0000212211.33265.6d https://doi.org/10.1097/01.ijg.0000212211.33265.6d
  44. MarekRJoannaWLewczukKSiemiatkowskaAStankiewiczA. Efficacy and safety of deep sclerectomy and phacoemulsification and deep sclerectomy in clinical material of Military Health Service Institute--yearly observations. Klin Oczna. 2006; 108(10- 12): 385-391
  45. MendrinosEDreifussSDossoABaglivoEShaarawyT. Bacterial keratitis after nonpenetrating glaucoma surgery. J Cataract Refract Surg. 2008; 34(4): 707-709. doi: 10.1016/j.jcrs.2007.11.03910.1016/j.jcrs.2007.11.039 https://doi.org/10.1016/j.jcrs.2007.11.039
  46. MermoudASchnyderCCSickenbergMChiouAGHédiguerSEFaggioniR. Comparison of deep sclerectomy with collagen implant and trabeculectomy in open-angle glaucoma. J Cataract Refract Surg. 1999; 25(3): 323-331. doi: 10.1016/s0886-3350(99)80079-010.1016/s0886-3350(99)80079-0 https://doi.org/10.1016/s0886-3350(99)80079-0
  47. Moreno-MontañésJRodríguez-CondeR. Bleeding during gonioscopy after deep sclerectomy. J Glaucoma. 2003; 12(5): 427-429
  48. Muñoz Negrete FJ, Rebolleda G, Noval S. Non-penetrating deep sclerectomy combined with phacoemulsification. Results and complications. Arch Soc Esp Oftalmol. 2003; 78(9): 499-506
  49. NeudorferMSadetzkiSAnisimovaSGeyerO. Nonpenetrating deep sclerectomy with the use of adjunctive mitomycin C. Ophthalmic Surg Lasers Imaging. 2004; 35(1): 6-12. doi: 10.3928/1542- 8877-20040101-0310.3928/1542-8877-20040101-03 https://doi.org/10.3928/1542-8877-20040101-03
  50. OllikainenMLPuustjärviTJRekonenPKUusitaloHMTeräsvirtaME. Mitomycin C-augmented deep sclerectomy in primary open-angle glaucoma and exfoliation glaucoma: A threeyear prospective study. Acta Ophthalmol. 2011; 89(6): 548-555. doi: 10.1111/j.1755-3768.2009.01772.x10.1111/j.1755-3768.2009.01772.x https://doi.org/10.1111/j.1755-3768.2009.01772.x
  51. OllikainenMLPuustjärviTJRekonenPKUusitaloHMTeräsvirtaME. Mitomycin-C-augmented deep sclerectomy in patients with primary open-angle glaucoma and exfoliation glaucoma: A 1-year prospective study. Acta Ophthalmol. 2010; 88(1): 20-26. doi: 10.1111/j.1755-3768.2009.01509.x10.1111/j.1755-3768.2009.01509.x https://doi.org/10.1111/j.1755-3768.2009.01509.x
  52. RavinetETrittenJJRoyS. Descemet membrane detachment after nonpenetrating filtering surgery. J Glaucoma. 2002; 11(3): 244-252. doi: 10.1097/00061198-200206000-0001410.1097/00061198-200206000-00014 https://doi.org/10.1097/00061198-200206000-00014
  53. RebolledaGMuñoz-NegreteFJ. Deep sclerectomy with mitomycin C in failed trabeculectomy. Eye (Lond). 2007; 21(1): 23-28. doi: 10.1038/sj.eye.670218310.1038/sj.eye.6702183 https://doi.org/10.1038/sj.eye.6702183
  54. RekonenPKannistoTPuustjärviTTeräsvirtaMUusitaloH. Deep sclerectomy for the treatment of exfoliation and primary open-angle glaucoma. Acta Ophthalmol Scand. 2006; 84(4): 507-511. doi: 10.1111/j.1600-0420.2006.00654.x10.1111/j.1600-0420.2006.00654.x https://doi.org/10.1111/j.1600-0420.2006.00654.x
  55. RivierDRoySMermoudA. Ex-PRESS R-50 miniature glaucoma implant insertion under the conjunctiva combined with cataract extraction. J Cataract Refract Surg. 2007; 33(11): 1946-1952. doi: 10.1016/j.jcrs.2007.06.06910.1016/j.jcrs.2007.06.069 https://doi.org/10.1016/j.jcrs.2007.06.069
  56. Robe-CollignonNJCollignon-BrachJD. Non perforating trabecular surgery with reticulated hyaluronic acid implant. Bull Soc Belge Ophtalmol. 2000; Suppl: 45-54
  57. ShaarawyTKarlenMSchnyderCAchacheFSanchezEMermoudA. Five-year results of deep sclerectomy with collagen implant. J Cataract Refract Surg. 2001; 27(11): 1770-1778. doi: 10.1016/s0886-3350(01)01032-x10.1016/s0886-3350(01)01032-x https://doi.org/10.1016/s0886-3350(01)01032-x
  58. SourdillePSantiagoPYBresson-DumontH. Transition to nonpenetrating trabecular surgery: Sclerotrabeculectomy. J Cataract Refract Surg. 2002; 28(12): 2080-2083. doi: 10.1016/s0886- 3350(02)01363-910.1016/s0886-3350(02)01363-9 https://doi.org/10.1016/s0886-3350(02)01363-9
  59. StewartRMDiamondJGAshmoreEDAyyalaRS. Complications following ex-press glaucoma shunt implantation. Am J Ophthalmol. 2005; 140(2): 340-341. doi: 10.1016/j.ajo.2005.02.03310.1016/j.ajo.2005.02.033 https://doi.org/10.1016/j.ajo.2005.02.033
  60. SuominenSHarjuMIhanamäkiTVestiE. The effect of deep sclerectomy on intraocular pressure of normal-tension glaucoma patients: 1-year results. Acta Ophthalmol. 2010; 88(1): 27-32. doi: 10.1111/j.1755-3768.2009.01623.x10.1111/j.1755-3768.2009.01623.x https://doi.org/10.1111/j.1755-3768.2009.01623.x
  61. TraversoCEDe FeoFMessas-KaplanA. Long term effect on IOP of a stainless steel glaucoma drainage implant (Ex-PRESS) in combined surgery with phacoemulsification. Br J Ophthalmol. 2005; 89(4): 425-429. doi: 10.1136/bjo.2004.05803210.1136/bjo.2004.058032 https://doi.org/10.1136/bjo.2004.058032
  62. Wagschal LD, Trope GE, Jinapriya D, Jin Y-P, Buys YM. Prospective randomized study comparing Ex-PRESS to trabeculectomy: 1-year results. J Glaucoma. 2015; 24(8): 624-629. doi: 10.1097/ IJG.0000000000000029
  63. Detry-MorelMDe TemmermanS. Assessment of nonpenetrating deep sclerectomy with reticulated hyaluronic acid implant SKGEL and/or preoperative application of 5-luorouracil: Results of 2 and a half years. Bull Soc Belge Ophtalmol. 2003; (287): 53-62
  64. Gallego-PinazoRLópez-SánchezEMarín-MontielJ. Postoperative outcomes after combined glaucoma surgery. Comparison of EX-PRESS miniature implant with standard trabeculectomy. Arch Soc Esp Oftalmol. 2009; 84(6): 293-297
  65. MarzetteLHerndonLW. A comparison of the Ex-PRESSTM mini glaucoma shunt with standard trabeculectomy in the surgical treatment of glaucoma. Ophthalmic Surg Lasers Imaging. 2011; 42(6): 453-459. doi: 10.3928/15428877-20111017-0310.3928/15428877-20111017-03 https://doi.org/10.3928/15428877-20111017-03
  66. WevillMTMeyerDVan AswegenE. A pilot study of deep sclerectomy with implantation of chromic suture material as a collagen implant: Medium-term results. Eye (Lond). 2005; 19(5): 549- 554. doi: 10.1038/sj.eye.670154110.1038/sj.eye.6701541 https://doi.org/10.1038/sj.eye.6701541
  67. GoodTJKahookMY. Assessment of bleb morphologic features and postoperative outcomes after Ex-PRESS drainage device implantation versus trabeculectomy. Am J Ophthalmol. 2011; 151(3): 507-513.e1. doi: 10.1016/j.ajo.2010.09.00410.1016/j.ajo.2010.09.004 https://doi.org/10.1016/j.ajo.2010.09.004
  68. GavrićMGabrićNJagićJCovićA. Clinical experience with EX-PRESS mini glaucoma shunt implantation. Coll Antropol. 2011; 35 Suppl 2: 39-41
  69. GindrozFRoySMermoudASchnyderCC. Combined Ex-PRESS LR-50/IOL implantation in modified deep sclerectomy plus phacoemulsification for glaucoma associated with cataract. Eur J Ophthalmol. 2011; 21(1): 12-19. doi: 10.5301/ejo.2010.54110.5301/ejo.2010.541 https://doi.org/10.5301/ejo.2010.541
  70. RekasMLewczukKFuksińskaBRudowiczJPawlikRStankiewiczA. Combined surgery for cataract and glaucoma: PDS with absorbable SK-gel implant compared with PDS with non-absorbable T-flux implant-medium-term results. Curr Med Res Opin. 2010; 26(5): 1131-1137. doi: 10.1185/0300799100371942810.1185/03007991003719428 https://doi.org/10.1185/03007991003719428
  71. MansouriKTranHVRavinetEMermoudA. Comparing deep sclerectomy with collagen implant to the new method of very deep sclerectomy with collagen implant: A single-masked randomized controlled trial. J Glaucoma. 2010; 19(1): 24-30. doi: 10.1097/ IJG.0b013e3181a2fa46
  72. Dahan E, Ben Simon GJ, Lafuma A. Comparison of trabeculectomy and Ex-PRESS implantation in fellow eyes of the same patient: A prospective, randomised study. Eye (Lond). 2012; 26(5): 703-710. doi: 10.1038/eye.2012.13 https://doi.org/10.1038/eye.2012.13
  73. RoySThiHDFeusierMMermoudA. Crosslinked sodium hyaluronate implant in deep sclerectomy for the surgical treatment of glaucoma. Eur J Ophthalmol. 2012; 22(1): 70-76. doi: 10.5301/ ejo.5000054
  74. MesciCErbilHHKarakurtYAkçakayaAA. Deep sclerectomy augmented with combination of absorbable biosynthetic sodium hyaluronate scleral implant and mitomycin C or with mitomycin C versus trabeculectomy: Long-term results. Clin Experiment Ophthalmol. 2007; 40(4): e197-e207. doi: 10.1111/j.1442- 9071.2011.02611.x10.1111/j.1442-9071.2011.02611.x https://doi.org/10.1111/j.1442-9071.2011.02611.x
  75. SuominenSHarjuMKurvinenLVestiE. Deep sclerectomy in normal-tension glaucoma with and without mitomycin-c. Acta Ophthalmol. 2014; 92(7): 701-706 doi: 10.1111/aos.1230510.1111/aos.12305 https://doi.org/10.1111/aos.12305
  76. AteşHUretmenOAndaçKAzarsizSS. Deep sclerectomy with a nonabsorbable implant (T-Flux): Preliminary results. Can J Ophthalmol. 2003; 38(6): 482-488. doi: 10.1016/S0008- 4182(03)80027-310.1016/s0008-4182(03)80027-3 https://doi.org/10.1016/s0008-4182(03)80027-3
  77. AmbresinAShaarawyTMermoudA. Deep sclerectomy with collagen implant in one eye compared with trabeculectomy in the other eye of the same patient. J Glaucoma. 2002; 11(3): 214-220. doi: 10.1097/00061198-200206000-0000910.1097/00061198-200206000-00009 https://doi.org/10.1097/00061198-200206000-00009
  78. JehnABBohnkeMMojonDS. Deep sclerectomy with collagen implant: Initial experience. Ophthalmologica. 2002; 216(4): 235- 238. doi: 10.1159/00006384610.1159/000063846 https://doi.org/10.1159/000063846
  79. GalassiFGiambeneB. Deep sclerectomy with SkGel implant: 5-year results. J Glaucoma. 2008; 17(1): 52-56. doi: 10.1097/ IJG.0b013e3180d0a885
  80. BissigAFeusierMMermoudARoyS. Deep sclerectomy with the Ex-PRESS X-200 implant for the surgical treatment of glaucoma. Int Ophthalmol. 2010; 30(6): 661-668. doi: 10.1007/ s10792-010-9382-z
  81. De FeoFBagnisABricolaGScottoRTraversoCE. Efficacy and safety of a steel drainage device implanted under a scleral flap. Can J Ophthalmol. 2009; 44(4): 457-462. doi: 10.3129/i09-12010.3129/i09-120 https://doi.org/10.3129/i09-120
  82. NarayanaswamyAPereraSAHtoonHMHohSTSeahSKWongTTAungT. Efficacy and safety of collagen matrix implants in phacotrabeculectomy and comparison with Mitomycin C augmented phacotrabeculectomy at 1 year. Clin Experiment Ophthalmol. 2013; 41(6): 552-560. doi: 10.1111/ceo.1205810.1111/ceo.12058 https://doi.org/10.1111/ceo.12058
  83. TengCCRadcliffeNHuangJEFarrisE. Ex-PRESS glaucoma shunt dislocation into the anterior chamber. J Glaucoma. 2008; 17(8): 687-689. doi: 10.1097/IJG.0b013e318166659510.1097/ijg.0b013e3181666595 https://doi.org/10.1097/ijg.0b013e3181666595
  84. KannerEMNetlandPASarkisianSRDuH. Ex-PRESS miniature glaucoma device implanted under a scleral flap alone or combined with phacoemulsification cataract surgery. J Glaucoma. 2009; 18(6): 488-491. doi: 10.1097/IJG.0b013e31818fb44e10.1097/ijg.0b013e31818fb44e https://doi.org/10.1097/ijg.0b013e31818fb44e
  85. MendaSALowryEAPorcoTCStamperRLRubinMRHanY. Ex-PRESS outcomes using Mitomycin-C, Ologen alone, Ologen with 5-fluorouracil. Int Ophthalmol. 2014; 35(3): 357-363. doi: 10.1007/s10792-014-9955-310.1007/s10792-014-9955-3 https://doi.org/10.1007/s10792-014-9955-3
  86. Huerva V, Carmen Sanchez M, Jesus Muniesa M. Ex-PRESS shunt surgery for glaucoma with post-traumatic aniridia. Semin Ophthalmol. 2015; 30(3): 218-220. doi: 10.3109/08820538.2013.835845 de Jong L, Lafuma A, Aguadé AS, Berdeaux G. Five-year extension of a clinical trial comparing the EX-PRESS glaucoma filtration device and trabeculectomy in primary open-angle glaucoma. Clin Ophthalmol. 2011; 5: 527-533. doi: 10.2147/OPTH.S18565 https://doi.org/10.3109/08820538.2013.835845
  87. LankaranianDRazeghinejadMRPrasadA. Intermediateterm results of the Ex-PRESS miniature glaucoma implant under a scleral flap in previously operated eyes. Clin Experiment Ophthalmol. 2011; 39(5): 421-428. doi: 10.1111/j.1442-9071.2010.02481.x10.1111/j.1442-9071.2010.02481.x https://doi.org/10.1111/j.1442-9071.2010.02481.x
  88. Al Obeidan SA, Osman EA, Mousa A, Al-Muammar AM, Abu El-Asrar AM. Long-term evaluation of efficacy and safety of deep sclerectomy in uveitic glaucoma. Ocul Immunol Inflamm. 2015; 23(1): 82-89. doi: 10.3109/09273948.2013.870213 https://doi.org/10.3109/09273948.2013.870213
  89. MendrinosEMansouriKMermoudAShaarawyT. Longterm results of deep sclerectomy with collagen implant in exfoliative glaucoma. J Glaucoma. 2009; 18(5): 361-367. doi: 10.1097/ IJG.0b013e3181879e4e
  90. ShaarawyTMansouriKSchnyderCRavinetEAchacheFMermoudA. Long-term results of deep sclerectomy with collagen implant. J Cataract Refract Surg. 2004; 30(6): 1225-1231. doi: 10.1016/j.jcrs.2003.10.03510.1016/j.jcrs.2003.10.035 https://doi.org/10.1016/j.jcrs.2003.10.035
  91. RhoSKangSYHongSSeongGJJungJJKimCY. Longterm results of deep sclerectomy with small collagen implant in Korean. Korean J Ophthalmol. 2013; 27(1): 34-38. doi: 10.3341/ kjo.2013.27.1.34
  92. Russo V, Scott IU, Stella A, Balducci F, Cosma A, Barone A, Delle Noci N. Nonpenetrating deep sclerectomy with reticulated hyaluronic acid implant versus punch trabeculectomy: A prospective clinical trial. Eur J Ophthalmol. 2008; 18(5): 751-757. doi: 10.1177/112067210801800515 https://doi.org/10.1177/112067210801800515
  93. LeszczyńskiRFormińska-KapuścikMBubała-StachowiczBMrukwa-KominekEFilipekEPawlickiK. Nonpenetrating very deep sclerectomy with hyaluronic acid implant vs trabeculectomy- -a 2-year follow-up. Graefes Arch Clin Exp Ophthalmol. 2012; 250(12): 1835-1841. doi: 10.1007/s00417-012-1985-910.1007/s00417-012-1985-9 https://doi.org/10.1007/s00417-012-1985-9
  94. Netland PA, Sarkisian SR Jr, Moster MR, et al. Randomized, prospective, comparative trial of EX-PRESS glaucoma filtration device versus trabeculectomy (XVT study). Am J Ophthalmol. 2014; 157(2): 433-440.e3. doi: 10.1016/j.ajo.2013.09.014 https://doi.org/10.1016/j.ajo.2013.09.014
  95. SeiderMIRofaghaSLinSCStamperRL. Resident-performed Ex-PRESS shunt implantation versus trabeculectomy. J Glaucoma. 2012; 21(7): 469-474. doi: 10.1097/IJG.0b013e3182182bfb10.1097/ijg.0b013e3182182bfb https://doi.org/10.1097/ijg.0b013e3182182bfb
  96. WamsleySMosterMRRaiSAlvimHSFontanarosaJ. Results of the use of the Ex-PRESS miniature glaucoma implant in technically challenging, advanced glaucoma cases: A clinical pilot study. Am J Ophthalmol. 2004; 138(6): 1049-1051. doi: 10.1016/j. ajo.2004.06.02410.1016/j.ajo.2004.06.024 https://doi.org/10.1016/j.ajo.2004.06.024
  97. MoisseievEZunzETzurRKurtzSShemeshG. Standard trabeculectomy and Ex-PRESS miniature glaucoma shunt: A comparative study and literature review. J Glaucoma. 2015; 24(6): 410- 416. doi: 10.1097/IJG.000000000000004710.1097/ijg.0000000000000047 https://doi.org/10.1097/ijg.0000000000000047
  98. MinJKKeeCWSohnSWLeeHJWooJMYimJH. Surgical outcome of Mitomycin C-soaked collagen matrix implant in trabeculectomy. J Glaucoma. 2013; 22(6): 456-462. doi: 10.1097/ IJG.0b013e31826ab6b1
  99. SalimSDuHBoonyaleephanSWanJ. Surgical outcomes of the Ex-PRESS glaucoma filtration device in African American and white glaucoma patients. Clin Ophthalmol. 2012; 6: 955-962. doi: 10.2147/OPTH.S3228210.2147/opth.s32282 https://doi.org/10.2147/opth.s32282
  100. BissigARivierDZaninettiMShaarawyTMermoudARoyS. Ten years follow-up after deepsclerectomy with collagen implant. J Glaucoma. 2008; 17(8): 680-686. doi: 10.1097/ IJG.0b013e318182ed9e
  101. RavinetEBoveyEMermoudA. T-Flux implant versus Healon GV in deep sclerectomy. J Glaucoma. 2004; 13(1): 46-50. doi: 10.1097/00061198-200402000-00009 de Jong LA. The Ex-PRESS glaucoma shunt versus trabeculectomy in open-angle glaucoma: A prospective randomized study. Adv Ther. 2009; 26(3): 336-345. doi: 10.1007/s12325-009-0017-610.1097/00061198-200402000-00009 https://doi.org/10.1097/00061198-200402000-00009
  102. SugiyamaTShibataMKojimaSUekiMIkedaT. The first report on intermediate-term outcome of Ex-PRESS glaucoma filtration device implanted under scleral flap in Japanese patients. Clin Ophthalmol. 2011; 5: 1063-1066. doi: 10.2147/OPTH.S2312910.2147/opth.s23129 https://doi.org/10.2147/opth.s23129
  103. AptelFDumasSDenisP. Ultrasound biomicroscopy and optical coherence tomography imaging of filtering blebs after deep sclerectomy with new collagen implant. Eur J Ophthalmol. 2009; 19(2): 223-230. doi: 10.1177/11206721090190020810.1177/112067210901900208 https://doi.org/10.1177/112067210901900208
  104. Johnson MS, Sarkisian SR Jr. Using a collagen matrix implant (Ologen) versus Mitomycin-C as a wound healing modulator in trabeculectomy with the Ex-PRESS mini glaucoma device: A 12-month retrospective review. J Glaucoma. 2014; 23(9): 649-652. doi: 10.1097/IJG.0000000000000018 https://doi.org/10.1097/ijg.0000000000000018
  105. GargSJKanitkarKWeichelEFischerD. Trauma-induced extrusion of an Ex-PRESS glaucoma shunt presenting as an intraocular foreign body. Arch Ophthalmol. 2005; 123(9): 1270-1272. doi: 10.1001/archopht.123.9.127010.1001/archopht.123.9.1270 https://doi.org/10.1001/archopht.123.9.1270
  106. Maris PJ Jr, Ishida K, Netland PA. Comparison of trabeculectomy with Ex-PRESS miniature glaucoma device implanted under scleral flap. J Glaucoma. 2007; 16(1): 14-19. doi: 10.1097/01. ijg.0000243479.90403.cd https://doi.org/10.1097/01.ijg.0000243479.90403.cd
  107. Stein JD, Herndon LW, Brent Bond J, Challa P. Exposure of Ex-PRESS miniature glaucoma devices: Case series and technique for tube shunt removal. J Glaucoma. 2007; 16(8): 704-706. doi: 10.1097/IJG.0b013e31806ab314 https://doi.org/10.1097/ijg.0b013e31806ab314
  108. ShaarawyTNguyenCSchnyderCMermoudA. Comparative study between deep sclerectomy with and without colagen implant: Long term follow up. Br J Ophthalmol. 2004; 88(1): 95-98. doi: 10.1136/bjo.88.1.9510.1136/bjo.88.1.95 https://doi.org/10.1136/bjo.88.1.95
  109. VetrugnoMFerreriPCardasciaNSborgiaC. Surgical reposition of a dislocated Ex-PRESS miniature device. Eur J Ophthalmol. 2011; 21: 212-214. doi: 10.5301/ejo.2010.581510.5301/ejo.2010.5815 https://doi.org/10.5301/ejo.2010.5815
  110. RosentreterASchildAMJordanJFKrieglsteinGKDietleinTS. A prospective randomised trial of trabeculectomy using Mitomycin C vs an ologen implant in open angle glaucoma. Eye (Lond). 2010; 24(9): 1449-1457. doi: 10.1038/eye.2010.10610.1038/eye.2010.106 https://doi.org/10.1038/eye.2010.106
  111. CillinoSDi PaceFCillinoGCasuccioA. Biodegradable collagen matrix implant vs mitomycin-C as an adjuvant in trabeculectomy: A 24-month, randomized clinical trial. Eye (Lond). 2011; 25(12): 1598-1606. doi: 10.1038/eye.2011.21910.1038/eye.2011.219 https://doi.org/10.1038/eye.2011.219
  112. ShaarawyTMermoudA. Deep sclerectomy in one eye vs deep sclerectomy with collagen implant in the contralateral eye of the same patient: Long-term follow-up. Eye (Lond). 2005; 19(3): 298-302. doi: 10.1038/sj.eye.670146910.1038/sj.eye.6701469 https://doi.org/10.1038/sj.eye.6701469
  113. SenthilSRaoHLBabuJGMandalAKGarudadriCS. Comparison of outcomes of trabeculectomy with mitomycin C vs. ologen implant in primary glaucoma. Indian J Ophthalmol. 2013; 61(7): 338-342. doi: 10.4103/0301-4738.10952010.4103/0301-4738.109520 https://doi.org/10.4103/0301-4738.109520
  114. DemaillyPLavatPKretzGJeanteur-LunelMN. Non-penetrating deep sclerectomy (NPDS) with or without collagen device (CD) in primary open-angle glaucoma: Middle-term retrospective study. Int Ophthalmol. 1996-1997; 20(1-3): 131-140
  115. NilforushanNYadegariMHodjatP. Comparison of the success rate of trabeculectomy with oculusgen versus trabeculectomy with Mitomycin C. Iranian Journal of Ophthalmology. 2011; 23(4): 3-12
  116. EgrilmezSAtesHNalcaciSAndacKYagciA. Surgically induced corneal refractive change following glaucoma surgery: Nonpenetrating trabecular surgeries versus trabeculectomy. J Cataract Refract Surg. 2004; 30(6): 1232-1239. doi: 10.1016/j.jcrs.2003.11.05510.1016/j.jcrs.2003.11.055 https://doi.org/10.1016/j.jcrs.2003.11.055
  117. Beltran-AgulloLTropeGEJinYWagschalLDJinapriyaDBuysYM. Comparison of visual recovery following EX-PRESS versus trabeculectomy: Results of a prospective randomized controlled trial. J Glaucoma. 2015; 24(3): 181-186. doi: 10.1097/ IJG.0b013e31829e1b68
  118. PatelHYWagschalLDTropeGEBuysYM. Economic analysis of the Ex-PRESS miniature glaucoma device versus trabeculectomy. J Glaucoma. 2014; 23(6): 385-390. doi: 10.1097/ IJG.0b013e31827a06f4
  119. MareyHMMandourSSEllakwaAF. Subscleral trabeculectomy with Mitomycin-C versus ologen for treatment of glaucoma. J Ocul Pharmacol Ther. 2013; 29(3): 330-334. doi: 10.1089/ jop.2012.0120
  120. DahanERavinetEBen-SimonGJMermoudA. Comparison of the efficacy and longevity of nonpenetrating glaucoma surgery with and without a new, nonabsorbable hydrophilic implant. Ophthalmic Surg Lasers Imaging. 2003; 34(6): 457-463
  121. D'EliseoDPastenaBLonganesiLGrisantiFNegriniV. Comparison of deep sclerectomy with implant and combined glaucoma surgery. Ophthalmologica. 2003; 217(3): 208-211. doi: 10.1159/00006897410.1159/000068974 https://doi.org/10.1159/000068974
  122. DindoDClavienPA. What is a surgical complication? World J Surg. 2008; 32: 939-941. doi: 10.1007/s00268-008-9584-y10.1007/s00268-008-9584-y https://doi.org/10.1007/s00268-008-9584-y
  123. ClavienPABarkunJde OliveiraML. The clavien-dindo classification of surgical complications: Five-year experience. Ann Surg. 2009; 250: 187-196. doi: 10.1097/SLA.0b013e3181b13ca210.1097/sla.0b013e3181b13ca2 https://doi.org/10.1097/sla.0b013e3181b13ca2
  124. HazellLShakirSA. Under-reporting of adverse drug reactions: A systematic review. Drug Saf. 2006; 29(5): 385-396. doi: 10.2165/00002018-200629050-0000310.2165/00002018-200629050-00003 https://doi.org/10.2165/00002018-200629050-00003