Microperimetry in Optic Neuritis

Authors

  • Indrish Bhatia
  • Shveta Lukhmana
  • Digvijay Singh
  • Vimala Menon
  • Pradeep Sharma
  • Rohit Saxena

Keywords:

Microperimetry; Optic neuritis; Visual fields; Perimetry; Microperimeter

Abstract

Aim: To evaluate microperimetry as a tool for visual field assessment in optic neuritis and compare it with standard automated perimetry.
Methods: A case-control study was conducted at a tertiary eye care centre in India. Ten cases
of unilateral optic neuritis and 10 healthy controls underwent a detailed ophthalmic evaluation
and visual field testing at presentation and 1 month and 3 months follow-up. Visual fields were
charted using both the standard automated perimeter (10-2 and 30-2 programs) and microperimeter (central 20 degree program) at each visit.
Results: The visual acuity at presentation of the affected eye, fellow eye and control eyes was
0.27±0.19, 0.93±0.14 and 0.94±0.17 respectively. The affected eye visual acuity improved significantly to 0.89±0.24 (p<0.001) at the 3 months follow-up visit. The mean sensitivity thresholds of 10-2 visual field test at presentation were 14.16±11.51, 30.40±1.98 and 31.04±1.95
respectively of which the affected eye showed a significant improvement over 3 months to
28.90±8.36 (p<0.001). The mean sensitivity thresholds of 30-2 visual field test at presentation were 12.88±10.32, 26.03±2.59 and 27.99±2.31 respectively of which the affected eye and
fellow eye showed a significant improvement over 3 months to 26.02±7.54 (p<0.001) and
27.86±1.77 (p<0.03) respectively. The mean sensitivity thresholds of microperimetry 20 degree
visual fields at presentation were 5.60±7.32, 16.54±1.46 and 17.30±1.64 respectively of which
the affected eye showed a significant improvement over 3 months to 16.41±4.87 (p<0.001). Microperimetry fields did not improve completely at 1 month unlike the 10-2 fields and correlated
strongly with visual recovery. The 30-2 fields were the most sensitive to determine subclinical
affliction of the fellow eyes in optic neuritis.
Conclusion: Microperimetry is a sensitive test to evaluate visual fields in optic neuritis and
corresponds with visual recovery. The larger 30 degree module is still indispensible for visual
field assessment in optic neuritis.

Downloads

Published

2016-08-02

Issue

Section

Articles