Study of Analysis the Sensitivity of the Computational Environment for Radiological Research Field Size Based on Two Dimensional Dose Distribution for Water Phantom Cases

Authors

  • Yati Hardiyanti Department of Physics, Institut Teknologi Bandung, Jalan Ganesa 10 Bandung 40132, Indonesia
  • Thareq Barasabha Faculty of Medicine, Universitas Brawijaya, Jalan Veteran Malang 65145, Indonesia
  • Choirul Anam Department of Physics, Universitas Diponegoro, Jalan Prof. H. Soedarto S. H,Tembalang,Semarang 50275, Indonesia
  • Novitrian Novitrian Department of Physics, Institut Teknologi Bandung, Jalan Ganesa 10 Bandung 40132, Indonesia
  • Freddy Haryanto Department of Physics, Institut Teknologi Bandung, Jalan Ganesa 10 Bandung 40132, Indonesia
  • Abdul Waris Department of Physics, Institut Teknologi Bandung, Jalan Ganesa 10 Bandung 40132, Indonesia

DOI:

https://doi.org/10.17140/ROJ-4-130

Keywords:

2D Dose Distribution, Sensitivity, Quality control, Treatment planning system, Radiation therapy dosimetry

Abstract

Purpose
This study analysed the sensitivity of the field size from variations in the target volume dimensions, depth, and position. The variations
in the target volume analysis were used to determine the width of the field size. Thus, the quality control of the radiation
beam can be obtained.
Materials and Methods
The computed tomography (CT) image of the IBA Dose 1 type of water phantom consists of 350 slices. Variations in the dimension
of the target volume were modelled in 10×10×10 cm3, 10×12×10 cm3, 10.2×10×10.2 cm3, and 15×15×15 cm3. Beam
parameters use one beam of irradiation on the central axis 0°, 6 MV energy, 100 cm source-skin distance (SSD), beamlet delta x,
and y set to 0.1 cm. Dose distribution in the form of the XZ isodose curve and dose profile was used to observe the field size.
Results
In this study, the isodose curve was successfully displayed in the XZ isodose curve. The field size’s sensitivity has been successfully
reviewed from variations of the target volume, depth, and position. The target X and Z direction analysis is used in determining
the width and length of the field size.
Conclusion
The analysis related to the field size sensitivity study was obtained from a relatively valid calculation. The field size was evaluated
with variations in depth of 1.5 cm, 5 cm, 10 cm, and variations in positions of 10 cm, 12 cm, 14 cm, 18 cm, and 20 cm. This study
will be used as a reference to validate the distribution of computational environment for radiotherapy research (CERR) dose in
the future. Thus, the accuracy of the dose calculation can be obtained.

References

Korhonen L. Methods for Dose Calculation and Beam Characterization in External Photon Beam Radiotherapy [dissertation]. Espoo, Finland: Helsinki University of Technology; 2009.

Birgani MJT, Chegeni N, Tahmasbi M, Fayazi R, Asl JF, Beh-rouzMA. Calculation of equivalen square fields in radiation therapy by an improved vadash correction factor for collimator exchange effects. Asian Pac J Cancer Prev. 2018; 19: 3053-3057. doi: 10.31557/ APJCP.2018.19.11.3053

Podgorsak E. Radiation Oncology Physics: A Handbook for Teachers andStudents. Vienna, Austria: International Atomic Energi Agency; 2005.

Craft D, Bangert M, LongT, Papp D,Unkelbach J. Shared data for intensity modulated radiation therapy (IMRT) optimiza-tion research: The CORT dataset. Gigascience. 2014; 3(1): 37. doi: 10.1186/2047-217X-3-37

Clarkson JF. A note on depth doses in fields of irregular shape. J Radiol. 1941; 14: 265-268.

Fung AYC, Enke CA, Ayyangar KM, Thompson RB, Zhen W, Raman NV, et al. Effects of field parameters on IMRT plan quality for gynecological cancer: A case study. J Appl Clin Med Phys. 2005; 6(3): 46-62. doi: 10.1120/jacmp.v6i3.2087

Jahromi AF, Bozorg O, Maleki H, Mosleh-Shirazi MA. Fluence map optimization in intensity modulated radiation therapy for fuzzy target dose. Iranian Journal of Fuzzy System. 2011; 8(4): 93-105. doi: 10.22111/IJFS.2011.310 Submit your article to this journal | https://openventio.org/submit-manuscript/ Original Research | Volume 4 | Issue 2| Hardiyanti Y et al

Ahnesjö A, Saxner M, Trepp A. A pencil beam model for photon dose calculation. Med Phys. 1992; 19(2): 263-273. doi: 10.1118/1.596856

Ahnesjö A, Aspradakis MM. Dose calculations for external photon beams in radiotherapy. Phys Med Biol.1999; 44: 99-155. doi:

1088/0031-9155/44/11/201 10. Deasy JO, BlancoAI, Clark VH. CERR: A computational envi-ronment for radiotherapy research. Med Phys. 2003; 30(5): 979-985. doi: 10.1118/1.1568978

Kalinin E, Deasy J. A method for fast 3-D IMRT dose calcula-tions: The quadrant infinite beam (QIB) algorithm. Med Phys. 2003; 50(8): 1348-1349.

Jelen U, Sohn M, Alber M. A finite size pencil beam for IMRT dose optimization. Biol Phys Med Biol. 2005; 50(8): 1747-1766. doi: 10.1088/0031-9155/50/8/009

Breedveld S, Storchi P, Keijzer M, Heijmen B. Fast, multiple optimizations of quadraticdose objective functions in IMRT. Phys Med Biol. 2006; 51(14): 35-69. doi: 10.1088/0031-9155/51/14/019

Zakarian C, Deasy J, Kalinin E. A method for fast 3-D IMRT dose calculations: The quadrant infinite beam (QIB) algorithm. 2003; 82: 5330.

Apte PA. WE-D-201-02: Walk-through of CERR capabilities. Med Phys.2015; 42 (6): 3672-3672. doi: 10.1118/1.4925955

Rahma AY. Validation of Quadrant Infinite Beam (QIB) Algorithm for Nasopharynx Cancer Intensity Modulated Radiation Therapy (IMRT) Treatment Planning, InstitutTeknologi Bandung [master’s thesis]. Band-ung, Indonesia: InstitutTeknologi Bandung; 2019.

Khan FM, Gibbons JP. The Physics of Radiation Therapy. Philadel-phia, USA: Lippincott Williams and Wilkins; 2014: 60-172.

Nath R, Biggs PJ, Bova FJ, Ling CC, Purdy JA, van de Geijn J, et al. AAPM code of practice for radiotherapy accelerators report No. 47. Med Phys. 1994; 21: 1093-1121 . doi: 10.1118/1.597398

Karzamark CJ. Advance in linear accelerator design for radio-therapy. Med Phys. 1984; 15(4): 604-610. doi: 10.1118/1.595617

Anam C. Development of 2D Isodose Curve from The PDD and Dose ProfilesUsing Matlab. 2th ISNPINSA; 2012: 48-52.

Downloads

Published

2020-10-06

How to Cite

Yati Hardiyanti, Thareq Barasabha, Choirul Anam, Novitrian Novitrian, Freddy Haryanto, & Abdul Waris. (2020). Study of Analysis the Sensitivity of the Computational Environment for Radiological Research Field Size Based on Two Dimensional Dose Distribution for Water Phantom Cases. Radiology, 4(2), 39–45. https://doi.org/10.17140/ROJ-4-130