A Comprehensive Analysis of the Literature on Implant-Related Complications in Anterior Bleb Forming Glaucoma Surgery
Keywords:
Implant-related complications (IRC); Glaucoma surgery; Anterior bleb forming.Abstract
Purpose
To make an account of published implant-related complications (IRC) by a systematic review of the literature.
Methods
A systematic search of Pubmed and Scopus databases and Google Scholar engine was performed with selection criteria to detect
papers on IRC. We excluded unrelated papers and reviewed selected ones. We considered papers that did not explicitly state about
occurrence or not occurrence of IRC as non-IRC reporting. Main outcome measures were the number of papers reporting on
complications, IRC, and types of IRC.
Results
After the search, selection, and addition, we studied 109 papers. Incidence of IRC was 4.5%, half required explantation. While
26 implant studies found IRCs (23%), 13 case reports on surgical complications, 8 (61.5%) of them reported IRC. Frequent
complications were conjunctival erosion, blockage of the tube, migration to anterior chamber or damage to surrounding tissues.
Conclusion
Most papers did not report on IRC. Length or nature of studies may skew finding IRC. The incidence of IRC was 4.5%. Hard
and sharp implants carry a greater risk of IRC and explantation.
References
Cheng JW, Cheng SW, Cai JP, Li Y, Wei RL. Systematic overview of the efficacy of nonpenetrating glaucoma surgery in the treat-ment of open angle glaucoma. Med Sci Monit. 2011; 17(7): RA155- RA163. doi: 10.12659/MSM.881840
Wang W, Zhang X. Meta-analysis of randomized controlled tri-als comparing EX-PRESS implantation with trabeculectomy for open-angle glaucoma. PLoS One. 2014; 9(6): e100578. doi: 10.1371/ journal.pone.0100578
Rulli E, Biagioli E, Riva I, et al. Efficacy and safety of trabeculec-tomy vs nonpenetrating surgical procedures: A systematic review and meta-analysis. JAMA Ophthalmol. 2013; 131(12):1573-1582. doi: 10.1001/jamaophthalmol.2013.5059
Cheng JW, Wei RL, Cai JP, Li Y. Efficacy and tolerability of non-penetrating filtering surgery with and without implant in treatment of open angle glaucoma: A quantitative evaluation of the evidence. J Glaucoma. 2009; 18(3): 233-237. doi: 10.1097/ IJG.0b013e3181812827
Mendeley. Web site. www.mendeley.com. Accessed September 16, 2019.
Chen G, Li W, Jiang F, Mao S, Tong Y. Ex-PRESS implantation versus trabeculectomy in open-angle glaucoma: A meta-analysis of randomized controlled clinical trials. PLoS One. 2014; 9(1): e86045. doi: 10.1371/journal.pone.0086045
Buys YM. Trabeculectomy versus Ex-PRESS shunt surgery in residency training. Surv Ophthalmol. 2012; 57(4): 375-378. doi: 10.1016/j.survophthal.2012.02.006
Demeter S, Hailey D. Non-penetrating glaucoma surgery using AquaFlow(TM)collagen implants. Issues Emerg Health Technol. 2001; (22): 1-4.
Filippopoulos T, Rhee DJ. Novel surgical procedures in glau-coma: Advances in penetrating glaucoma surgery. Curr Opin Oph-thalmol. 2008; 19(2): 149-154. doi: 10.1097/ICU.0b013e3282f4f49e
Francis BA, Singh K, Lin SC, et al. Novel glaucoma procedures: A report by the American Academy of Ophthalmology. Ophthal-mology. 2011; 118(7): 1466-1480. doi: 10.1016/j.ophtha.2011.03.028
Hendrick AM, Kahook MY. Ex-PRESS mini glaucoma shunt: Surgical technique and review of clinical experience. Expert Rev Med Devices. 2008; 5(6): 673-677. doi: 10.1586/17434440.5.6.673
Salim S. The role of the Ex-PRESS glaucoma filtration device in glaucoma surgery. Semin Ophthalmol. 2013; 28(3): 180-184. doi: 10.3109/08820538.2013.771197
Buys YM. Trabeculectomy with Ex-PRESS: Weighing the benefits and cost. Curr Opin Ophthalmol. 2013; 24(2): 111-118. doi: 10.1097/ICU.0b013e32835907a6
Salim S. Ex-PRESS glaucoma filtration device: Surgical tech-nique and outcomes. Int Ophthalmol Clin. 2011; 51(3): 83-94. doi: 10.1097/IIO.0b013e31821e5b82
Shaarawy T, Goldberg I, Fechtner R. EX-PRESS glaucoma fil-tration device: Review of clinical experience and comparison with trabeculectomy. Surv Ophthalmol. 2015; 60: 327-345. doi: 10.1016/j. survophthal.2015.01.001
Ji Q, Qi B, Liu L, Guo X, Zhong J. Efficacy and safety of ologen implant versus Mitomycin C in primary trabeculectomy: A meta-analysis of randomized clinical trials. J Glaucoma. 2015; 24: e88-e94. doi: 10.1097/IJG.0000000000000147 15 Salinas A et al
Al Obeidan SA. Migration of reticulated hyaluronic acid im-plant [SK-GEL] following deep sclerectomy. Int Ophthalmol. 2010; 30(3): 329-331. doi: 10.1007/s10792-009-9325-8
Al Obeidan SA, Osman EA, Al-Muammar AM, Abu El-Asrar AM. Efficacy and safety of deep sclerectomy in uveitic glaucoma. Int Ophthalmol. 2009; 29(5): 367-372. doi: 10.1007/s10792-008- 9247-x
Al-Obeidan SA, Mousa A, Naseem A, Abu-Amero KK, Os-man EA. Efficacy and safety of non-penetrating deep sclerectomy surgery in Saudi patients with uncontrolled open angle glaucoma. Saudi Med J. 2013; 34(1): 54-61.
Arruabarrena C, Muñoz-Negrete FJ, Márquez C, Rebolleda G. Results of nonpenetrating deep sclerectomy in inflammatory glaucoma: One year follow up. Arch Soc Esp Oftalmol. 2007; 82(8): 483-487.
Ates H, Andac K, Uretmen O. Non-penetrating deep sclerec-tomy and collagen implant surgery in glaucoma patients with ad-vanced field loss. Int Ophthalmol. 1999; 23(3): 123-128.
Auer C, Mermoud A, Herbort CP. Deep sclerectomy for the management of uncontrolled uveitic glaucoma: Preliminary data. Klin Monbl Augenheilkd. 2004; 221(5): 339-342. doi: 10.1055/s-2004-812820
Cillino S, Zeppa L, Di Pace F, et al. E-PTFE (Gore-Tex) implant with or without low-dosage mitomycin-C as an adju-vant in penetrating glaucoma surgery: 2 year randomized clinical trial. Acta Ophthalmol. 2008; 86(3): 314-321. doi: 10.1111/j.1600- 0420.2007.01036.x
Collignon-Brach J, Robe-Collignon N. Non perforating trabec-ular surgery with reticulated hyaluronic acid implant. Bull Soc Belge Ophtalmol. 2000; 276: 61-68.
Coupin A, Li Q, Riss I. Ex-PRESSTM miniature glaucoma im-plant inserted under a scleral flap in open-angle glaucoma surgery: A retrospective study. J Fr Ophtalmol. 2007; 30(1): 18-23.
Dahan E, Carmichael TR. Implantation of a miniature glau-coma device under a scleral flap. J Glaucoma. 2005; 14(2): 98-102. doi: 10.1097/01.ijg.0000151688.34904.b7
Demailly P, Jeanteur-Lunel MN, Berkani M, et al. Non-pene-trating deep sclerectomy combined with a collagen implant in pri-mary open-angle glaucoma. Medium-term retrospective results. J Fr Ophtalmol. 1996; 19(11): 659-666.
Detry-Morel M. Non penetrating deep sclerectomy (NPDS) with SKGEL implant and/or 5-fluorouracile (5-FU). Bull Soc Belge Ophtalmol. 2001; (280): 23-32.
Detry-Morel M, Detry MB. Five-year experience with non pen-etrating deep sclerectomy. Bull Soc Belge Ophtalmol. 2006; (299): 83- 94.
Detry-Morel M, Pourjavan S, Detry MB. Comparative safety profile between “modern” trabeculectomy and non-penetrationg deep sclerectomy. Bull Soc Belge Ophtalmol. 2006; (300): 43-54.
Drolsum L. Conversion from trabeculectomy to deep sclerec-tomy. Prospective study of the first 44 cases. J Cataract Refract Surg. 2003; 29(7): 1378-1384. doi: 10.1016/S0886-3350(03)00013-0
Dupas B, Fardeau C, Cassoux N, Bodaghi B, LeHoang P. Deep sclerectomy and trabeculectomy in uveitic glaucoma. Eye (Lond). 2010; 24(2): 310-314. doi: 10.1038/eye.2009.82
Elgin U, Simsek T, Batman A. Use of the ex-press miniature glaucoma implant in a child with Sturge-Weber syndrome. J Pediatr Ophthalmol Strabismus. 2007; 44(4): 248-250.
Hamard P, Plaza L, Kopel J, Quesnot S, Hamard H. Deep non-penetrating sclerectomy and open angle glaucoma. Intermediate results from the first operated patients. J Fr Ophtalmol. 1999; 22(1): 25-31.
Hamel M, Shaarawy T, Mermoud A. Deep sclerectomy with collagen implant in patients with glaucoma and high myopia. J Cataract Refract Surg. 2001; 27(9): 1410-1417. doi: 10.1016/s0886- 3350(01)00959-2
Jungkim S, Gibran SK, Khan K, Murray A. External trabecu-lectomy with T-Flux implant. Eur J Ophthalmol. 2006; 16(3): 416- 421. doi: 10.1177/112067210601600309
Karlen ME, Sanchez E, Schnyder CC, Sickenberg M, Mermoud A. Deep sclerectomy with collagen implant: Medium term results. Br J Ophthalmol. 1999; 83(1): 6-11. doi: 10.1136/bjo.83.1.6
Kim CY, Chang HR, Lee JH, Hong YJ. Surgical outcomes of deep sclerectomy with collagen implant. Korean J Ophthalmol. 2001; 15(2): 107-112. doi: 10.3341/kjo.2001.15.2.107
Kim CY, Hong YJ, Seong GJ, Koh HJ, Kim SS. Iris synechia after laser goniopuncture in a patient having deep sclerectomy with a collagen implant. J Cataract Refract Surg. 2002; 28(5): 900-902. doi: 10.1016/s0886-3350(01)01307-4
Leszczyński R, Gierek-Ciaciura S, Forminśka-Kapuścik M, Mrukwa-Kominek E, Rokita-Wala I. Nonpenetrating very deep sclerectomy with reticulated hyaluronic acid implant in glaucoma treatment. Med Sci Monit. 2008; 14(2): CR86- CR89.
Li JY, Lim MC, Mannis MJ. Traumatic cataract associated with mini glaucoma shunt. J Cataract Refract Surg. 2011; 37(7): 1360- 1362. doi: 10.1016/j.jcrs.2011.04.019
Lüke C, Dietlein TS, Jacobi PC, Konen W, Krieglstein GK. A prospective randomised trial of viscocanalostomy with and with-out implantation of a reticulated hyaluronic acid implant (SKGEL) in open angle glaucoma. Br J Ophthalmol. 2003; 87(5): 599-603. doi: 10.1136/bjo.87.5.599 Salinas A et al
Mansouri K, Shaarawy T, Wedrich A, Mermoud A. Compar-ing polymethylmethacrylate implant with collagen implant in deep sclerectomy: A randomized controlled trial. J Glaucoma. 2006; 15(3): 264-270. doi: 10.1097/01.ijg.0000212211.33265.6d
Marek R, Joanna W, Lewczuk K, Siemiatkowska A, Stankiewicz A. Efficacy and safety of deep sclerectomy and phacoemulsifica-tion and deep sclerectomy in clinical material of Military Health Service Institute--yearly observations. Klin Oczna. 2006; 108(10- 12): 385-391.
Mendrinos E, Dreifuss S, Dosso A, Baglivo E, Shaarawy T. Bacterial keratitis after nonpenetrating glaucoma surgery. J Cataract Refract Surg. 2008; 34(4): 707-709. doi: 10.1016/j.jcrs.2007.11.039
Mermoud A, Schnyder CC, Sickenberg M, Chiou AG, Hédiguer SE, Faggioni R. Comparison of deep sclerectomy with collagen im-plant and trabeculectomy in open-angle glaucoma. J Cataract Refract Surg. 1999; 25(3): 323-331. doi: 10.1016/s0886-3350(99)80079-0
Moreno-Montañés J, Rodríguez-Conde R. Bleeding during go-nioscopy after deep sclerectomy. J Glaucoma. 2003; 12(5): 427-429.
Muñoz Negrete FJ, Rebolleda G, Noval S. Non-penetrating deep sclerectomy combined with phacoemulsification. Results and complications. Arch Soc Esp Oftalmol. 2003; 78(9): 499-506.
Neudorfer M, Sadetzki S, Anisimova S, Geyer O. Nonpenetrat-ing deep sclerectomy with the use of adjunctive mitomycin C. Oph-thalmic Surg Lasers Imaging. 2004; 35(1): 6-12. doi: 10.3928/1542- 8877-20040101-03
Ollikainen ML, Puustjärvi TJ, Rekonen PK, Uusitalo HM, Teräsvirta ME. Mitomycin C-augmented deep sclerectomy in primary open-angle glaucoma and exfoliation glaucoma: A three-year prospective study. Acta Ophthalmol. 2011; 89(6): 548-555. doi: 10.1111/j.1755-3768.2009.01772.x
Ollikainen ML, Puustjärvi TJ, Rekonen PK, Uusitalo HM, Teräsvirta ME. Mitomycin-C-augmented deep sclerectomy in pa-tients with primary open-angle glaucoma and exfoliation glauco-ma: A 1-year prospective study. Acta Ophthalmol. 2010; 88(1): 20-26. doi: 10.1111/j.1755-3768.2009.01509.x
Ravinet E, Tritten JJ, Roy S, et al. Descemet membrane detach-ment after nonpenetrating filtering surgery. J Glaucoma. 2002; 11(3): 244-252. doi: 10.1097/00061198-200206000-00014
Rebolleda G, Muñoz-Negrete FJ. Deep sclerectomy with mi-tomycin C in failed trabeculectomy. Eye (Lond). 2007; 21(1): 23-28. doi: 10.1038/sj.eye.6702183
Rekonen P, Kannisto T, Puustjärvi T, Teräsvirta M, Uusitalo H. Deep sclerectomy for the treatment of exfoliation and primary open-angle glaucoma. Acta Ophthalmol Scand. 2006; 84(4): 507-511. doi: 10.1111/j.1600-0420.2006.00654.x
Rivier D, Roy S, Mermoud A. Ex-PRESS R-50 miniature glau-coma implant insertion under the conjunctiva combined with cata-ract extraction. J Cataract Refract Surg. 2007; 33(11): 1946-1952. doi: 10.1016/j.jcrs.2007.06.069
Robe-Collignon NJ, Collignon-Brach JD. Non perforating tra-becular surgery with reticulated hyaluronic acid implant. Bull Soc Belge Ophtalmol. 2000; Suppl: 45-54.
Shaarawy T, Karlen M, Schnyder C, Achache F, Sanchez E, Mermoud A. Five-year results of deep sclerectomy with colla-gen implant. J Cataract Refract Surg. 2001; 27(11): 1770-1778. doi: 10.1016/s0886-3350(01)01032-x
Sourdille P, Santiago PY, Bresson-Dumont H. Transition to nonpenetrating trabecular surgery: Sclerotrabeculectomy. J Cata-ract Refract Surg. 2002; 28(12): 2080-2083. doi: 10.1016/s0886- 3350(02)01363-9
Stewart RM, Diamond JG, Ashmore ED, Ayyala RS. Complica-tions following ex-press glaucoma shunt implantation. Am J Oph-thalmol. 2005; 140(2): 340-341. doi: 10.1016/j.ajo.2005.02.033
Suominen S, Harju M, Ihanamäki T, Vesti E. The effect of deep sclerectomy on intraocular pressure of normal-tension glau-coma patients: 1-year results. Acta Ophthalmol. 2010; 88(1): 27-32. doi: 10.1111/j.1755-3768.2009.01623.x
Traverso CE, De Feo F, Messas-Kaplan A, et al. Long term effect on IOP of a stainless steel glaucoma drainage implant (Ex-PRESS) in combined surgery with phacoemulsification. Br J Oph-thalmol. 2005; 89(4): 425-429. doi: 10.1136/bjo.2004.058032
Wagschal LD, Trope GE, Jinapriya D, Jin Y-P, Buys YM. Pro-spective randomized study comparing Ex-PRESS to trabeculecto-my: 1-year results. J Glaucoma. 2015; 24(8): 624-629. doi: 10.1097/ IJG.0000000000000029
Detry-Morel M, De Temmerman S. Assessment of nonpen-etrating deep sclerectomy with reticulated hyaluronic acid implant SKGEL and/or preoperative application of 5-luorouracil: Results of 2 and a half years. Bull Soc Belge Ophtalmol. 2003; (287): 53-62.
Gallego-Pinazo R, López-Sánchez E, Marín-Montiel J. Postop-erative outcomes after combined glaucoma surgery. Comparison of EX-PRESS miniature implant with standard trabeculectomy. Arch Soc Esp Oftalmol. 2009; 84(6): 293-297.
Marzette L, Herndon LW. A comparison of the Ex-PRESSTM mini glaucoma shunt with standard trabeculectomy in the surgical treatment of glaucoma. Ophthalmic Surg Lasers Imaging. 2011; 42(6): 453-459. doi: 10.3928/15428877-20111017-03
Wevill MT, Meyer D, Van Aswegen E. A pilot study of deep sclerectomy with implantation of chromic suture material as a col-lagen implant: Medium-term results. Eye (Lond). 2005; 19(5): 549- 554. doi: 10.1038/sj.eye.6701541 17 Salinas A et al
Good TJ, Kahook MY. Assessment of bleb morphologic fea-tures and postoperative outcomes after Ex-PRESS drainage device implantation versus trabeculectomy. Am J Ophthalmol. 2011; 151(3): 507-513.e1. doi: 10.1016/j.ajo.2010.09.004
Gavrić M, Gabrić N, Jagić J, Cović A. Clinical experience with EX-PRESS mini glaucoma shunt implantation. Coll Antropol. 2011; 35 Suppl 2: 39-41.
Gindroz F, Roy S, Mermoud A, Schnyder CC. Combined Ex-PRESS LR-50/IOL implantation in modified deep sclerectomy plus phacoemulsification for glaucoma associated with cataract. Eur J Ophthalmol. 2011; 21(1): 12-19. doi: 10.5301/ejo.2010.541
Rekas M, Lewczuk K, Fuksińska B, Rudowicz J, Pawlik R, Stankiewicz A. Combined surgery for cataract and glaucoma: PDS with absorbable SK-gel implant compared with PDS with non-ab-sorbable T-flux implant - medium-term results. Curr Med Res Opin. 2010; 26(5): 1131-1137. doi: 10.1185/03007991003719428
Mansouri K, Tran HV, Ravinet E, Mermoud A. Comparing deep sclerectomy with collagen implant to the new method of very deep sclerectomy with collagen implant: A single-masked random-ized controlled trial. J Glaucoma. 2010; 19(1): 24-30. doi: 10.1097/ IJG.0b013e3181a2fa46
Dahan E, Ben Simon GJ, Lafuma A. Comparison of trabecu-lectomy and Ex-PRESS implantation in fellow eyes of the same patient: A prospective, randomised study. Eye (Lond). 2012; 26(5): 703-710. doi: 10.1038/eye.2012.13
Roy S, Thi HD, Feusier M, Mermoud A. Crosslinked sodium hyaluronate implant in deep sclerectomy for the surgical treatment of glaucoma. Eur J Ophthalmol. 2012; 22(1): 70-76. doi: 10.5301/ ejo.5000054
Mesci C, Erbil HH, Karakurt Y, Akçakaya AA. Deep sclerec-tomy augmented with combination of absorbable biosynthetic sodium hyaluronate scleral implant and mitomycin C or with mitomycin C versus trabeculectomy: Long-term results. Clin Ex-periment Ophthalmol. 2007; 40(4): e197-e207. doi: 10.1111/j.1442- 9071.2011.02611.x
Suominen S, Harju M, Kurvinen L, Vesti E. Deep sclerectomy in normal-tension glaucoma with and without mitomycin-c. Acta Ophthalmol. 2014; 92(7): 701-706 doi: 10.1111/aos.12305
Ateş H, Uretmen O, Andaç K, Azarsiz SS. Deep sclerecto-my with a nonabsorbable implant (T-Flux): Preliminary results. Can J Ophthalmol. 2003; 38(6): 482-488. doi: 10.1016/S0008- 4182(03)80027-3
Ambresin A, Shaarawy T, Mermoud A. Deep sclerectomy with collagen implant in one eye compared with trabeculectomy in the other eye of the same patient. J Glaucoma. 2002; 11(3): 214-220. doi: 10.1097/00061198-200206000-00009
Jehn AB, Bohnke M, Mojon DS. Deep sclerectomy with col-lagen implant: Initial experience. Ophthalmologica. 2002; 216(4): 235- 238. doi: 10.1159/000063846
Galassi F, Giambene B. Deep sclerectomy with SkGel im-plant: 5-year results. J Glaucoma. 2008; 17(1): 52-56. doi: 10.1097/ IJG.0b013e3180d0a885
Bissig A, Feusier M, Mermoud A, Roy S. Deep sclerectomy with the Ex-PRESS X-200 implant for the surgical treatment of glaucoma. Int Ophthalmol. 2010; 30(6): 661-668. doi: 10.1007/ s10792-010-9382-z
De Feo F, Bagnis A, Bricola G, Scotto R, Traverso CE. Efficacy and safety of a steel drainage device implanted under a scleral flap. Can J Ophthalmol. 2009; 44(4): 457-462. doi: 10.3129/i09-120
Narayanaswamy A, Perera SA, Htoon HM, Hoh ST, Seah SK, Wong TT, Aung T. Efficacy and safety of collagen matrix implants in phacotrabeculectomy and comparison with Mitomycin C aug-mented phacotrabeculectomy at 1 year. Clin Experiment Ophthalmol. 2013; 41(6): 552-560. doi: 10.1111/ceo.12058
Teng CC, Radcliffe N, Huang JE, Farris E. Ex-PRESS glauco-ma shunt dislocation into the anterior chamber. J Glaucoma. 2008; 17(8): 687-689. doi: 10.1097/IJG.0b013e3181666595
Kanner EM, Netland PA, Sarkisian SR, Du H. Ex-PRESS miniature glaucoma device implanted under a scleral flap alone or combined with phacoemulsification cataract surgery. J Glaucoma. 2009; 18(6): 488-491. doi: 10.1097/IJG.0b013e31818fb44e
Menda SA, Lowry EA, Porco TC, Stamper RL, Rubin MR, Han Y. Ex-PRESS outcomes using Mitomycin-C, Ologen alone, Ologen with 5-fluorouracil. Int Ophthalmol. 2014; 35(3): 357-363. doi: 10.1007/s10792-014-9955-3
Huerva V, Carmen Sanchez M, Jesus Muniesa M. Ex-PRESS shunt surgery for glaucoma with post-traumatic aniridia. Semin Oph-thalmol. 2015; 30(3): 218-220. doi: 10.3109/08820538.2013.835845
de Jong L, Lafuma A, Aguadé AS, Berdeaux G. Five-year exten-sion of a clinical trial comparing the EX-PRESS glaucoma filtra-tion device and trabeculectomy in primary open-angle glaucoma. Clin Ophthalmol. 2011; 5: 527-533. doi: 10.2147/OPTH.S18565
Lankaranian D, Razeghinejad MR, Prasad A, et al. Intermediate-term results of the Ex-PRESS miniature glaucoma implant under a scleral flap in previously operated eyes. Clin Experiment Ophthalmol. 2011; 39(5): 421-428. doi: 10.1111/j.1442-9071.2010.02481.x
Al Obeidan SA, Osman EA, Mousa A, Al-Muammar AM, Abu El-Asrar AM. Long-term evaluation of efficacy and safety of deep sclerectomy in uveitic glaucoma. Ocul Immunol Inflamm. 2015; 23(1): 82-89. doi: 10.3109/09273948.2013.870213
Mendrinos E, Mansouri K, Mermoud A, Shaarawy T. Long-Salinas A et al term results of deep sclerectomy with collagen implant in exfo-liative glaucoma. J Glaucoma. 2009; 18(5): 361-367. doi: 10.1097/ IJG.0b013e3181879e4e
Shaarawy T, Mansouri K, Schnyder C, Ravinet E, Achache F, Mermoud A. Long-term results of deep sclerectomy with colla-gen implant. J Cataract Refract Surg. 2004; 30(6): 1225-1231. doi: 10.1016/j.jcrs.2003.10.035
Rho S, Kang SY, Hong S, Seong GJ, Jung JJ, Kim CY. Long-term results of deep sclerectomy with small collagen implant in Korean. Korean J Ophthalmol. 2013; 27(1): 34-38. doi: 10.3341/ kjo.2013.27.1.34
Russo V, Scott IU, Stella A, Balducci F, Cosma A, Barone A, Delle Noci N. Nonpenetrating deep sclerectomy with reticulated hyaluronic acid implant versus punch trabeculectomy: A pro-spective clinical trial. Eur J Ophthalmol. 2008; 18(5): 751-757. doi: 10.1177/112067210801800515
Leszczyński R, Formińska-Kapuścik M, Bubała-Stachowicz B, Mrukwa-Kominek E, Filipek E, Pawlicki K. Nonpenetrating very deep sclerectomy with hyaluronic acid implant vs trabeculectomy--a 2-year follow-up. Graefes Arch Clin Exp Ophthalmol. 2012; 250(12): 1835-1841. doi: 10.1007/s00417-012-1985-9
Netland PA, Sarkisian SR Jr, Moster MR, et al. Randomized, prospective, comparative trial of EX-PRESS glaucoma filtration device versus trabeculectomy (XVT study). Am J Ophthalmol. 2014; 157(2): 433-440.e3. doi: 10.1016/j.ajo.2013.09.014
Seider MI, Rofagha S, Lin SC, Stamper RL. Resident-performed Ex-PRESS shunt implantation versus trabeculectomy. J Glaucoma. 2012; 21(7): 469-474. doi: 10.1097/IJG.0b013e3182182bfb
Wamsley S, Moster MR, Rai S, Alvim HS, Fontanarosa J. Re-sults of the use of the Ex-PRESS miniature glaucoma implant in technically challenging, advanced glaucoma cases: A clinical pilot study. Am J Ophthalmol. 2004; 138(6): 1049-1051. doi: 10.1016/j. ajo.2004.06.024
Moisseiev E, Zunz E, Tzur R, Kurtz S, Shemesh G. Standard trabeculectomy and Ex-PRESS miniature glaucoma shunt: A com-parative study and literature review. J Glaucoma. 2015; 24(6): 410- 416. doi: 10.1097/IJG.0000000000000047
Min JK, Kee CW, Sohn SW, Lee HJ, Woo JM, Yim JH. Surgi-cal outcome of Mitomycin C-soaked collagen matrix implant in trabeculectomy. J Glaucoma. 2013; 22(6): 456-462. doi: 10.1097/ IJG.0b013e31826ab6b1
Salim S, Du H, Boonyaleephan S, Wan J. Surgical outcomes of the Ex-PRESS glaucoma filtration device in African American and white glaucoma patients. Clin Ophthalmol. 2012; 6: 955-962. doi: 10.2147/OPTH.S32282
Bissig A, Rivier D, Zaninetti M, Shaarawy T, Mermoud A, Roy S. Ten years follow-up after deepsclerectomy with colla-gen implant. J Glaucoma. 2008; 17(8): 680-686. doi: 10.1097/ IJG.0b013e318182ed9e
Ravinet E, Bovey E, Mermoud A. T-Flux implant versus Healon GV in deep sclerectomy. J Glaucoma. 2004; 13(1): 46-50. doi: 10.1097/00061198-200402000-00009
de Jong LA. The Ex-PRESS glaucoma shunt versus trabecu-lectomy in open-angle glaucoma: A prospective randomized study. Adv Ther. 2009; 26(3): 336-345. doi: 10.1007/s12325-009-0017-6
Sugiyama T, Shibata M, Kojima S, Ueki M, Ikeda T. The first report on intermediate-term outcome of Ex-PRESS glaucoma fil-tration device implanted under scleral flap in Japanese patients. Clin Ophthalmol. 2011; 5: 1063-1066. doi: 10.2147/OPTH.S23129
Aptel F, Dumas S, Denis P. Ultrasound biomicroscopy and optical coherence tomography imaging of filtering blebs after deep sclerectomy with new collagen implant. Eur J Ophthalmol. 2009; 19(2): 223-230. doi: 10.1177/112067210901900208
Johnson MS, Sarkisian SR Jr. Using a collagen matrix implant (Ologen) versus Mitomycin-C as a wound healing modulator in trabeculectomy with the Ex-PRESS mini glaucoma device: A 12-month retrospective review. J Glaucoma. 2014; 23(9): 649-652. doi: 10.1097/IJG.0000000000000018
Garg SJ, Kanitkar K, Weichel E, Fischer D. Trauma-induced extrusion of an Ex-PRESS glaucoma shunt presenting as an intra-ocular foreign body. Arch Ophthalmol. 2005; 123(9): 1270-1272. doi: 10.1001/archopht.123.9.1270
Maris PJ Jr, Ishida K, Netland PA. Comparison of trabecu-lectomy with Ex-PRESS miniature glaucoma device implanted un-der scleral flap. J Glaucoma. 2007; 16(1): 14-19. doi: 10.1097/01. ijg.0000243479.90403.cd
Stein JD, Herndon LW, Brent Bond J, Challa P. Exposure of Ex-PRESS miniature glaucoma devices: Case series and technique for tube shunt removal. J Glaucoma. 2007; 16(8): 704-706. doi: 10.1097/IJG.0b013e31806ab314
Shaarawy T, Nguyen C, Schnyder C, Mermoud A. Compara-tive study between deep sclerectomy with and without colagen im-plant: Long term follow up. Br J Ophthalmol. 2004; 88(1): 95-98. doi: 10.1136/bjo.88.1.95
Vetrugno M, Ferreri P, Cardascia N, Sborgia C. Surgical repo-sition of a dislocated Ex-PRESS miniature device. Eur J Ophthal-mol. 2011; 21: 212-214. doi: 10.5301/ejo.2010.5815
Rosentreter A, Schild AM, Jordan JF, Krieglstein GK, Dietlein TS. A prospective randomised trial of trabeculectomy using Mito-mycin C vs an ologen implant in open angle glaucoma. Eye (Lond). 2010; 24(9): 1449-1457. doi: 10.1038/eye.2010.106 19 Salinas A et al
Cillino S, Di Pace F, Cillino G, Casuccio A. Biodegradable collagen matrix implant vs mitomycin-C as an adjuvant in trabecu-lectomy: A 24-month, randomized clinical trial. Eye (Lond). 2011; 25(12): 1598-1606. doi: 10.1038/eye.2011.219
Shaarawy T, Mermoud A. Deep sclerectomy in one eye vs deep sclerectomy with collagen implant in the contralateral eye of the same patient: Long-term follow-up. Eye (Lond). 2005; 19(3): 298-302. doi: 10.1038/sj.eye.6701469
Senthil S, Rao HL, Babu JG, Mandal AK, Garudadri CS. Comparison of outcomes of trabeculectomy with mitomycin C vs. ologen implant in primary glaucoma. Indian J Ophthalmol. 2013; 61(7): 338-342. doi: 10.4103/0301-4738.109520
Demailly P, Lavat P, Kretz G, Jeanteur-Lunel MN. Non-pene-trating deep sclerectomy (NPDS) with or without collagen device (CD) in primary open-angle glaucoma: Middle-term retrospective study. Int Ophthalmol. 1996-1997; 20(1-3): 131-140.
Nilforushan N, Yadegari M, Hodjat P. Comparison of the success rate of trabeculectomy with oculusgen versus trabecu-lectomy with Mitomycin C. Iranian Journal of Ophthalmology. 2011; 23(4): 3-12.
Egrilmez S, Ates H, Nalcaci S, Andac K, Yagci A. Surgically in-duced corneal refractive change following glaucoma surgery: Non-penetrating trabecular surgeries versus trabeculectomy. J Cataract Refract Surg. 2004; 30(6): 1232-1239. doi: 10.1016/j.jcrs.2003.11.055
Beltran-Agullo L, Trope GE, Jin Y, Wagschal LD, Jinapriya D, Buys YM. Comparison of visual recovery following EX-PRESS versus trabeculectomy: Results of a prospective randomized controlled trial. J Glaucoma. 2015; 24(3): 181-186. doi: 10.1097/ IJG.0b013e31829e1b68
Patel HY, Wagschal LD, Trope GE, Buys YM. Economic analysis of the Ex-PRESS miniature glaucoma device versus tra-beculectomy. J Glaucoma. 2014; 23(6): 385-390. doi: 10.1097/ IJG.0b013e31827a06f4
Marey HM, Mandour SS, Ellakwa AF. Subscleral trabecu-lectomy with Mitomycin-C versus ologen for treatment of glau-coma. J Ocul Pharmacol Ther. 2013; 29(3): 330-334. doi: 10.1089/ jop.2012.0120
Dahan E, Ravinet E, Ben-Simon GJ, Mermoud A. Compari-son of the efficacy and longevity of nonpenetrating glaucoma sur-gery with and without a new, nonabsorbable hydrophilic implant. Ophthalmic Surg Lasers Imaging. 2003; 34(6): 457-463.
D'Eliseo D, Pastena B, Longanesi L, Grisanti F, Negrini V. Comparison of deep sclerectomy with implant and combined glaucoma surgery. Ophthalmologica. 2003; 217(3): 208-211. doi: 10.1159/000068974
Dindo D, Clavien PA. What is a surgical complication? World J Surg. 2008; 32: 939-941. doi: 10.1007/s00268-008-9584-y
Clavien PA, Barkun J, de Oliveira ML, et al. The clavien-dindo classification of surgical complications: Five-year experience. Ann Surg. 2009; 250: 187-196. doi: 10.1097/SLA.0b013e3181b13ca2 Submit your article to this journal | https://openventio.org/submit-manuscript/ Salinas A et al
Hazell L, Shakir SA. Under-reporting of adverse drug reac-tions: A systematic review. Drug Saf. 2006; 29(5): 385-396. doi: 10.2165/00002018-200629050-00003
Downloads
Published
How to Cite
Issue
Section
License

This work is licensed under a Creative Commons Attribution 4.0 International License.
All content published in The Ophthalmology (TOOJ) is licensed under the Creative Commons Attribution 4.0 International License (CC BY 4.0). This license allows anyone to copy, distribute, transmit, and adapt the work — including for commercial purposes — provided appropriate credit is given to the original author(s) and source.