Double-Low Dose Protocol of Computed Tomography Pulmonary Angiography (CTPA) in the Diagnosis of Pulmonary Embolism: A Feasible Approach for Reduction of Both Contrast Medium and Radiation Doses
DOI:
https://doi.org/10.17140/HROJ-4-139Keywords:
Computed tomography pulmonary angiography, Contrast medium, Optimization; Image quality, Pulmonary embolism, Radiation doseAbstract
This article discusses a hot topic on double low-dose protocol of computed tomography pulmonary
angiography (CTPA) in the diagnosis of pulmonary embolism, with a special focus
on a recent paper published in the British Journal of Radiology about the feasibility of this
technique. Three aspects will be discussed in this review: First, both kVp and contrast medium
can be reduced without affecting image quality when compared to the standard CTPA protocol;
second, a low-pitch protocol is comparable to the high-pitch spiral image acquisition with
similar image quality achieved, but at lower radiation dose; and finally, the double low-dose
CTPA protocol can achieve diagnostic images in patients with body mass index up to 35 kg/m2.
References
Mayo J, Thakur Y. Pulmonary CT angiography as first-linei-maging for PE: Image quality and radiation dose considerations. AJR Am J Roentgenol. 2013; 200(3): 522-528. doi: 10.2214/ AJR.12.9928
Wittram C. How I do it: CT pulmonary angiography. AJR Am J Roentgenol. 2007; 188(5): 1255-1261. doi: 10.2214/ AJR.06.1104
Den Exter AL, van der Hulle T, Klok FA, Huisman MV. Ad-vances in the diagnosis and management of acute pulmonary embolism. Thromb Res. 2014; 133(Suppl 2): S10-S16. doi: 10.1016/S0049-3848(14)50002-3
Righini M, Le GG, Aujesky D, et al. Diagnosis of pulmo-nary embolism by multidetector CT alone or combinedwith ve-nous ultrasonography of the leg: A randomised non-inferiority trial. Lancet. 2008; 371(9621): 1343-1352. doi: 10.1016/S0140- 6736(08)60594-2
Ghanima W, Almaas V, Aballi S, et al. Management of sus-pected pulmonary embolism [PE] by D-dimer and multi-slice computed tomography in outpatients: An outcome study. J Thromb Haemost. 2005; 3(9): 1926-1932. doi: 10.1111/j.1538- 7836.2005.01544.x
Mos IC, Klok FA, Kroft LJ, Roos AD, Dekkers OM, Huisman MV. Safety ofruling out acute pulmonary embolism by normal computed tomographypulmonary angiography in patients with an indication for computed tomography: Systematic review and meta-analysis. J Thromb Haemost. 2009; 7(9): 1491-1498. doi: 10.1111/j.1538-7836.2009.03518.x
Carrier M, Righini M, Wells PS, et al. Subsegmental pulmo-nary embolism diagnosed by computed tomography: Incidence and clinical implications. A systematic review and meta-analysis of the management outcome studies. J Thromb Haemost. 2010; 8(8): 1716-1722. doi: 10.1111/j.1538-7836.2010.03938.x
Stein PD, Chenevert TL, Fowler SE, et al. Gadolinium-en-hanced magnetic resonance angiography for pulmonary embo-lism: A multicenter prospective study [PIOPED III]. Ann Intern Med. 2010; 152(7): 434-433. doi: 10.7326/0003-4819-152-7- 201004060-00008
Sun Z, Al Dosari S, Ng C, al-Muntashari A, Almaliky S. Mul-tislice CT. Virtual intravascular endoscopy for assessing pul-monary embolisms: A pictorial review. Korean J Radiol. 2010; 11(2): 222-230. doi: 10.3348/kjr.2010.11.2.222
Schoepf UJ, Costello P. CT angiography for diagnosis of pulmonary embolism: State of the art. Radiology. 2004; 230(2): 329-337. doi: 10.1148/radiol.2302021489
Sharma S, Lucas CD. Increasing use of CTPA for the in-vestigation of suspected pulmonary embolism. Postgradu Med. 2017; 129(2): 193-197. doi: 10.1080/00325481.2017.1281084
van Es J, Douma RA, Schreuder SM, et al. Clinical impact of findings supporting an alternative diagnosis on CT pulmonary angiography in patients with suspected pulmonary embolism. Chest. 2013; 144(6): 1893-1899. doi: 10.1378/chest.13-0157
Ong CW, Malipatil V, Lavercombe M, et al. Implementation of a clinicalprediction tool for pulmonary embolism diagnosis in a tertiary teaching hospital reduces the numberof computed tomography pulmonary angiograms performed. Intern Med J. 2013; 43(2): 169-174. doi: 10.1111/j.1445-5994.2012.02926.x
Chandra S, Sarkar PK, Chandra D, Ginsberg NE, Cohen RI. Finding an alternative diagnosis does notjustify increased use of CT-pulmonary angiography. BMC Pulm Med. 2013; 13: 9. doi: 10.1186/1471-2466-13-9
Mamlouk MD, vanSonnenberg E, Gosalia R, et al. Pulmo-nary embolism at CT angiography: Implications for appropriate-ness, cost, and radiation exposure in 2003 patients. Radiology. 2010; 256(2): 625-632. doi: 10.1148/radiol.10091624
Mitchell AM, Jones AE, Tumlin JA, Kline JA. Prospective study of the incidence of contrast-induced nephropathy among patients evaluated for pulmonary embolism by contrast-en-hanced computed tomography. Acad Emerg Med. 2012; 19(6): 618-625. doi: 10.1111/j.1553-2712.2012.01374.x
Mountain D, Keijzers G, Chu K, et al. RESPECT-ED: Rate of pulmonary embolism (PE) and sub-segmental PE with mod-ern computed tomographic pulmonary angiograms in emer-gency departments: A multi-center observational study finds significant yield variation, uncorrelated with use or small PE rates. Plos One. 2016; 11(12): e0166483. doi: 10.1371/journal. pone.0166483
Costantino MM, Randall G, Gosselin M, et al. CT angiog-raphy in the evaluation of acute pulmonary embolus. AJR Am J Roentgenol. 2008; 191(2): 471-474. doi: 10.2214/AJR.07.2552
Costa AF, Basseri H, Sheikh A, Stiell I, Dennie C. The yield of CT pulmonary angiograms to exclude acute pulmonary em-bolism. Emerg Radiol. 2014; 21(2): 133-141. doi: 10.1007/ s10140-013-1169-x
Newman DH, Schriger DL. Rethinking testing for pulmo-nary embolism: Less is more. Ann Emerg Med. 2011; 57(6): 622-627. doi: 10.1016/j.annemergmed.2011.04.014
Shen Y, Sun Z, Xu L, et al. High-pitch, low-voltage and low-iodine concentration CT angiography of aorta: Assessment of image quality and radiation dose with iterative reconstruc-tion. Plos One. 2015; 10(2): 0117469. doi: 10.1371/journal. pone.0117469
Shen Y, Fan Z, Sun Z, et al. High pitch dual-source whole aorta CT angiography in the detection of coronary arteries: A feasibility study of using Iodixanol 270 and 100 kVp with itera-tive reconstruction. J Med Imaging Health Inf. 2015; 5: 117-125. doi: 10.1166/jmihi.2015.1367
Wang H, Xu L, Zhang N, Fan Z, Zhang Z, Sun Z. Coro-nary CT angiography in coronary artery bypass grafts: Compari-son between low concentration iodixanol 270 and Iohexol 350. J Comput Assist Tomogr. 2015; 39(1): 112-118. doi: 10.1097/ RCT.0000000000000162
Wu Q, Wang Y, Kai H, et al. Application of 80-kVp tube voltage, low-concentration contrast agent and iterative recon-struction in coronary CT angiography: Evaluation of image quality and radiation dose. Int J Clin Pract. 2016; 70: B50-B55. doi: 10.1111/ijcp.12852
Sabarudin A, Mustafa Z, Nassir KM, Hamid HA, Sun Z. Radiation dose reduction in thoracic and abdomen-pelvic CT us-ing tube current modulation: A phantom study. J Appl Clin Med Phys. 2015; 16(1): 319-328. doi: 10.1120/jacmp.v16i1.5135
Hansen NJ, Kaza RK, Maturen KE, Liu PS, Platt JF. Evalu-ation of low-dose CT angiography with model-based iterative reconstruction after endovascular aneurysm repair of a thoracic or abdominal aortic aneurysm. AJR Am J Roentgenol. 2014; 202(3): 648-655. doi: 10.2214/AJR.13.11286
Kaur M, Vijayananthan A, Kumar G, Jayarani K, Ng KH, Sun Z. Use of 100 kV versus 120 kV in computed tomography pulmonary angiography in the detection of pulmonary embo-lism: Effect on radiation dose and image quality. Quant Imag-ing Med Surg. 2015; 5(4): 524-533. doi: 10.3978/j.issn.2223- 4292.2015.04.04
Almutairi A, Sun Z, Poovathumkadavi A, Assar T. Dual-en-ergy CT angiography of peripheral arterial disease: Feasibility of using lower contrast medium volume. Plos One. 2015; 10(9): e0139275. doi: 10.1371/journal.pone.0139275
Faggioni L, Neri E, Sbragia P, et al. 80-kV pulmonary CT angiography with 40 ml of iodinated contrast material in lean patients: Comparison of vascular enhancement with iodixanol (320mg I/ml) and iomeprol (400 mg I/ml). AJR Am J Roent-genol. 2012; 199(6): 1220-1225. doi: 10.2214/AJR.11.8122
Lu G, Luo S, Meinel FG, et al. High-pitch computed tomog-raphy pulmonary angiography with iterative reconstruction at 80 kVp and 20 ml contrast agent volume. Eur Radiol. 2014; 24(12): 3260-3268. doi: 10.1007/s00330-014-3365-9
Boos J, Kropil P, Lanzman RS, et al. CT pulmonary angi-ography: Simultaneous low-pitch dual-source acquisition mode with 70 kVp and 40 ml of contrast medium and comparison with high-pitch spiral dual-source acquisition with automated tube potential selection. Br J Radiol. 2016; 89(1062): 20151059. doi: 10.1259/bjr.20151059
Sun Z, Wan YL, Hsieh IC, Liu YC, Wen MS. Coronary CT angiography in the diagnosis of coronary artery disease. Curr Med Imaging Rev. 2013; 9: 184-193. Web site. http://www.in-gentaconnect.com/content/ben/cmir/2013/00000009/00000003/ art00003. Accessed March 29, 2017.
Sun Z, Almoudi M, Cao Y. CT angiography in the diagno-sis of cardiovascular disease: A transformation in cardiovascular CT practice. Quant Imaging Med Surg. 2014; 4: 376-396. doi: 10.3978/j.issn.2223-4292.2014.10.02
Halliburton SS, Abbara S, Chen MY, et al. SCCT guidelines on radiation dose and dose-optimization strategies in cardiovas-cular CT. J Cardiovasc Comput Tomogr. 2011; 5(4): 198-224. doi: 10.1016/j.jcct.2011.06.001
Laqmani A, Kurfurst M, Butscheidt S, et al. CT pulmonary angiography at reduced radiation exposure and contrast mate-rial volume using iterative model reconstruction and iDose tech-nique in comparison to FBP. Plos One. 2016; 11: e0162429. doi: 10.1371/journal.pone.0162429
Szucs-Farkas Z, Schibler F, Cullmann J, et al. Diagnostic ac-curacy of pulmonary CT angiography at low tube voltage: Intra individual comparison of a normal-dose protocol at 120 kVp and a low-dose protocol at 80 kVp using reduced amount of contrast medium in a simulation study. AJR Am J Roentgenol. 2011; 197: W852-W859. doi: 10.2214/AJR.11.6750
Szucs-Farkas Z, Schaller C, Bensler S, Patak MA, Vock P, Schindera ST. Detection of pulmonary emboli with CT an-giography at reduced radiation exposure and contrast mate-rial volume: Comparison of 80 kVp and 120 kVp protocols in a matched cohort. Invest Radiol. 2009; 44(12): 793-799. doi: 10.1097/RLI.0b013e3181bfe230
Szucs-Farkas Z, Kurmann L, Strautz T, Patak MA, Vock P, Schindera ST. Patient exposure and imagequality of low-dose pulmonary computed tomography angiography: Comparison of 100- and 80-kVp protocols. Invest Radiol. 2008; 43: 871-876. doi: 10.1097/RLI.0b013e3181875e86
Laqmani A, Regier M, Veldhoen S, et al. Improved image qualityand low radiation dose with hybrid iterative reconstruc-tion with 80 kV CT pulmonary angiography. Eur J Radiol. 2014; 83(10): 1962-1969. doi: 10.1016/j.ejrad.2014.06.016
Saade C, Mayat A, El-Merhi F. Exponentially decelerated contrast media injection rate combined with a novel patient-spe-cific contrast formula reduces contrast volume administration and radiation dose during computed tomography pulmonary an-giography. J Comput Assist Tomogr. 2016; 40(3): 370-374. doi: 10.1097/RCT.0000000000000371
De Zordo T, von Lutterotti K, Dejaco C, et al. Comparison of image quality and radiation dose of different pulmonary CTA protocols on a 128-slice CT: High-pitch dual source CT, dual energy CT and conventional spiral CT. Eur Radiol. 2012; 22(2): 279-286. doi: 10.1007/s00330-011-2251-y
Tacelli N, Remy-Jardin M, Flohr T, et al. Dual-source chest CT angiography with high temporal resolution and high pitch modes: Evaluation of image quality in 140 patients. Eur Radiol. 2010; 20(5): 1188-1196. doi: 10.1007/s00330-009-1638-5
Wichmann JL, Hu X, Kerl JM, et al. 70 kVp computed to-mography pulmonary angiography: Potential for reduction of iodine load andradiation dose. J Thorac Imaging. 2015; 30(1): 69-76. doi: 10.1097/RTI.0000000000000124
Downloads
Published
How to Cite
Issue
Section
License

This work is licensed under a Creative Commons Attribution 4.0 International License.
All content published in Heart Research (HROJ) is licensed under the Creative Commons Attribution 4.0 International License (CC BY 4.0). This license allows anyone to copy, distribute, transmit, and adapt the work — including for commercial purposes — provided appropriate credit is given to the original author(s) and source.