Linearity of the Scale for Mass and Volume within the Air Displacement Plethysmograph (BOD POD): A Methodological Investigation

Crossmark: check for updates

Authors

  • Claire Mills School of Sport and Exercise, Faculty of Applied Sciences, University of Gloucestershire, Gloucester, GL2 9HW, UK
  • Mark De Ste Croix School of Sport and Exercise, Faculty of Applied Sciences, University of Gloucestershire, Gloucester, GL2 9HW, UK
  • David James School of Sport and Exercise, Faculty of Applied Sciences, University of Gloucestershire, Gloucester, GL2 9HW, UK

DOI:

https://doi.org/10.17140/SEMOJ-3-149

Keywords:

Air displacement plethysmograph, BOD POD, Linearity, Method, Mass, Volume, Reliability

Abstract

Introduction: In order to maintain the accuracy and reliability for both volume and mass measurements of the air displacement plethysmograph (BOD POD) on a day-to-day basis, quality assurance processes are undertaken. Given the importance of accurate estimation of body mass and body volume in determining body composition, the aim of this methodological investigation was to further examine the calibration approaches and to independently determine both the linearity and reliability of mass and volume measurements throughout the potential measurement range.
Methods: Routine calibration procedures for mass (sequentially add known masses ranging from 10-30 kg) range and volume (sequentially add known volume of balloons ranging from 49.900 L to 118.40 L) were conducted using BOD POD model 2000A (Life Measurement Inc. (LMI), Concord, CA, USA). Scatter plots between actual (known) against predicted (measurement) mass and volume values and bias and 95% limits of agreement plots were produced to illustrate the agreement, and paired t-tests to determine significant differences between the volumes.
Results: Results revealed that for all mass measurements between 10-30 kg the known mass and measured mass were in agreement. With respect to all volume measurements, the predicted (measured) volume differed from the actual (known) volume by as little as 0.2 L and as much as 0.9 L. There was a difference between actual (known) (mean±SD=65.1±35.9 l) and predicted (measured) (64.7±35.8 L), t9=6.35 p<0.01.
Conclusion: One might question the relevance of only being able to calibrate mass to a maximum of 30 kg, when body mass of adult participants certainly exceed 30 kg. Results from the adapted volume calibration trial using balloons revealed underreporting of predicted (measured) volumes by 0.4 L. However, on the basis of this methodological investigation, it is possible to be broadly confident with the linearity and reliability of both mass and volume measurement outcomes from the BOD POD involving a reasonable level of rigour.

References

McArdle WD. Body composition assessment. In: McArdle WD, Katch FI, Katch VL, eds. Exercise Physiology: Energy, Nu-trition and Human Performance. 6th ed. New York, USA: Lip-pincott, Williams & Williams; 2006.

Dempster P, Aitkens S. A new air displacement method for the determination of human body composition. Med Sci Sports Exerc. 1995; 27(12): 1692-1697.

Maddalozzo GF, Cardinal BJ, Snow CM. Concurrent validity of the BOD POD and dual energy x-ray absorptiometry tech-niques for assessing body composition in young women. J Am Diet Assoc. 2002: 102(11): 1677-1679. doi: 10.1016/S0002- 8223(02)90358-5

Fields DA, Higgins PB, Hunter GR. Assessment of body composition by air-displacement plethysmography: Influence of body temperature and moisture. Dyn Med. 2004; 3(3): 1-7. doi: 10.1186/1476-5918-3-3

Fields DA, Goran MI. Body composition techniques and the four-compartment model in children. J Appl Physiol (1985). 2000; 89(2): 613-620.

Vescovi JD, Zimmerman SL, Miller WC, Fernhall BO. Ef-fects of clothing on accuracy and reliability of air displacement plethysmography. Med Sci Sports Exerc. 2002; 34(2): 282-285.

Heymsfield SB, Lohman TG, Wang Z, Going SB. Human Body Composition. 2nd ed. Champaign, IL, USA: Human Kinet-ics; 2005.

Hull HR, Fields DA. Effect of short schemes on body compo-sition measurements using air-displacement plethysmography. Dyn Med. 2005; 4(1): 1-8. doi: 10.1186/1476-5918-4-8

Life Measurement Inc. BOD POD® body composition track-ing system. Operators Manual. 2006. Web site. http://www.taq. com.mx/documentos_productos/1348613957.pdf. Accessed March 14, 2017.

Buchholz AC, Majchrzak KM, Chen KY, Shankar SM, Buchowski MS. Use of air displacement plethysomography in the determination of percentage of fat mass in African Ameri-can children. Pediatr Res. 2004; 56(1): 47-54. doi: 10.1203/01. PDR.0000130477.05324.79

Millard-Stafford ML, Collins MA, Evans EM, Snow TK, Cureton KJ, Rosskopf LB. Use of air displacement plethysmog-raphy for estimating body fat in 4 compartment model. Med Sci Sports Exerc. 2001; 33(8): 1311-1317.

Fields DA, Goran MI, McCroy MA. Body composition as-sessment via air-displacement plethysmography in adults and children: A review. Am J Clin Nutr. 2002; 75(3): 453-467.

Yee AJ, Fuerst T, Salamone L, et al. Calibration and valida-tion of an air-displacement plethysmography method for esti-mating percentage body fat in an elderly population: A compari-son among compartmental models. Am J Clin Nutr. 2001; 74(5): 637-642.

Hoffman CJ, Hildebrandt LA. Use of the air displacement plethysmograph to monitor body composition: A beneficial tool for dieticians. J Am Diet Assoc. 2001; 101(9): 986-987. doi: 10.1016/S0002-8223(01)00242-5

Wells JC, Fuller NJ. Precision of measurement and body size in whole body air displacement plethysmography. Int J Obes Relat Metab Disord. 2001; 25(8): 1161-1167.

Wagner DR, Heyward VH, Gibson AL. Validation of air displacement plethysmography for assessing body compo-sition. Med Sci Sports Exerc. 2000; 32(7): 1339-1344. doi: 10.1097/00005768-200007000-00023

Collins MA, Millard-Stafford ML, Sparling PB, et al. Eval-uation of the BOD POD for assessing body fat in collegiate football players. Med Sci Sports Exerc. 1999; 31(9): 1350- 1356. doi: 10.1097/00005768-199909000-00019

Lockner DW, Heyward VH, Baumgartner RN, Jenkins KA. Comparison of air-displacement Plethysmography, hy-drodensitometry and dual X-ray absorptiometry for assessing body composition of children 10 to 18 years of age. Ann N Y Acad Sci. 2000; 904: 72-78. doi: 10.1111/j.1749-6632.2000. tb06423.x

Downloads

Published

2017-06-22

How to Cite

Claire Mills, Mark De Ste Croix, & David James. (2017). Linearity of the Scale for Mass and Volume within the Air Displacement Plethysmograph (BOD POD): A Methodological Investigation. Sports and Exercise Medicine , 3(2), 58–62. https://doi.org/10.17140/SEMOJ-3-149